Fluke 754 Documenting Process Calibrator-HART
Key features
- Complete pressure, temperature, and mA loop calibrator
- Troubleshoot and calibrate HART smart digital transmitters
- Create calibration procedures and automatically document results
- Connect to calibration management software
Sale ends soon.
Shop now!
Product overview: Fluke 754 Documenting Process Calibrator-HART
Fluke Premium Care eligible
When you invest in your Fluke test tools, you want your money to go as far as possible. Fluke Premium Care is a paid offering that provides coverage above and beyond the original product warranty, so you don’t need to worry about unexpected downtime caused by damaged test equipment, accessories, or tools in need of calibration or repair.
All-in-One Multi-Function Calibrator for Process Professionals
Maximize your investment with the Fluke 754 Documenting Process Calibrator-HART—a portable, rugged, and versatile calibrator trusted by professionals worldwide. Whether you’re calibrating instruments, troubleshooting, or performing routine maintenance, the Fluke 754 helps you work faster and more efficiently.
Why Choose the Fluke 754?
- Multi-function capability: Source, simulate, and measure pressure, temperature, and electrical signals—all in one handheld tool.
- Integrated HART communication: Calibrate and maintain HART smart transmitters and instrumentation without a separate communicator.
- Bright dual display: View sourced and measured parameters at the same time for faster, error-free work.
- Long battery life: Rechargeable Li-Ion battery delivers up to 10 hours of operation, plus direct AC power support.
- Seamless documentation: Automate calibration procedures, store results, and ensure compliance with ISO 9000, FDA, EPA, and OSHA standards.
- Portable and rugged: Designed for demanding industrial environments.
Key Measurement and Source Capabilities
- AC/DC voltage
- mA loop with 24V loop power
- Pressure (compatible with Fluke 750P Series Pressure Modules)
- Thermocouples & RTDs
- Frequency
- Resistance
Enhanced Process Calibration. Simplified.
The Fluke 754 is more than just a calibrator. Its built-in HART interface lets you perform nearly all digital communicator tasks. Easily handle fast-pulsed RTD transmitters and PLCs, custom auto-step and ramping, custom units, programmable measurement delay, and one- or two-point switch testing—all with one device.
Download procedures, lists, and instructions created in your calibration management software, or upload calibration results for documentation and analysis. The Fluke 754 supports a full week of procedures and is compatible with leading asset management solutions.
Go Paperless with Fluke DPCTrack2 Software
For a seamless, paperless calibration management experience, pair your Fluke 754 Documenting Process Calibrator-HART with Fluke DPCTrack2 Calibration Management Software. Simplify record-keeping, automate procedures, and streamline compliance.
Ready to Work Smarter? Upgrade to the Fluke 754 Today.
Experience the power, precision, and reliability of the Fluke 754 Documenting Process Calibrator-HART—the process industry’s most trusted multi-function calibrator.
What's in the box:
- Fluke 754 Documenting Process Calibrator-HART
- BC7240 battery charger
- Li-on BP7240 battery pack
- NIST-traceable calibration report and data
- (3 sets) TP220 test probes
- (3 sets) "Extended tooth" alligator clips
- (3 sets) Stackable test leads
- (2 sets) AC280 hook clips
- C799 Soft Field Case
- USB communication cable, Fluke 754HHC HART communications cable
- Quick Reference Guide
- 3-year warranty
Specifications: Fluke 754 Documenting Process Calibrator-HART
Measurement Accuracy | |||||
Voltage DC | Range/ Resolution | 1 Year | 2 Years | ||
100.000 mV | 0.02% + 0.005 mV | 0.03% + 0.005 mV | |||
3.00000 V | 0.02% + 0.00005 V | 0.03% + 0.00005 V | |||
30.0000 V | 0.02% + 0.0005 V | 0.03% + 0.0005 V | |||
300.00 V | 0.05% + 0.05 V | 0.07% + 0.05 V | |||
Voltage AC | 3.000 V (40 Hz to 500 Hz) / 0.001 V | 0.5% + 0.002 V | 1.0% + 0.004 V | ||
30.00 V (40 Hz to 500 Hz) / 0.01 V | 0.5% + 0.02 V | 1.0% + 0.04 V | |||
300.0 V (40 Hz to 500 Hz) / 0.1 V | 0.5% + 0.2 V | 1.0% + 0.2 V | |||
Current DC | 30.000 mA | 0.01% + 5 uA | 0.015% + 7 uA | ||
110.00 mA | 0.01% + 20 uA | 0.015% + 30 uA | |||
Resistance | 10.000 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | ||
100.00 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | |||
1.0000 kΩ | 0.05% + 500 mΩ | 0.07% + 0.5 Ω | |||
10.000 kΩ | 0.1% + 10 Ω | 0.15% + 15 Ω | |||
Frequency | 1.00 to 110.00 Hz / 0.01 Hz | 0.05 Hz | |||
110.1 to 1100.0 Hz / 0.1 Hz | 0.5 Hz | ||||
1.101 to 11.000 kHz / 0.001 kHz | 0.005 kHz | ||||
11.01 to 50.00 kHz / 0.01 kHz | 0.05 kHz | ||||
Source Accuracy | |||||
1 Year | 2 Years | ||||
Voltage DC | 100.000 mV | 0.01% + 0.005 mV | 0.015% + 0.005 mV | ||
1.00000 V | 0.01% + 0.00005 V | 0.015% + 0.0005 V | |||
15.0000 V | 0.01% + 0.0005 V | 0.015% + 0.0005 V | |||
Current DC | 22.000 mA (source) | 0.01% + 0.003 mA | 0.02% + 0.003 mA | ||
Current sink (simulate) | 0.02% + 0.007 mA | 0.04% + 0.007 mA | |||
Resistance | 10.000 Ω | 0.01% + 10 mΩ | 0.015% + 15 mΩ | ||
100.00 Ω | 0.01% + 20 mΩ | 0.015% + 30 mΩ | |||
1.0000 kΩ | 0.02% + 0.2 Ω | 0.03% + 0.3 Ω | |||
10.000 kΩ | 0.02% + 3 Ω | 0.03% + 5 Ω | |||
Frequency | 0.1 to 10.99 Hz | 0.01 Hz | |||
0.01 to 10.99 Hz | 0.01 Hz | ||||
11.00 to 109.99 Hz | 0.1 Hz | ||||
110.0 to 1099.9 Hz | 0.1 Hz | ||||
1.100 to 21.999 kHz | 0.002 kHz | ||||
22.000 to 50.000 kHz | 0.005 kHz | ||||
Technical Data | |||||
Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | |||
Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | ||||
Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | ||||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Rate | 4 steps/second | ||||
Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | ||||
Loop power function | Voltage | Selectable, 26 V | |||
Accuracy | 10%, 18 V minimum at 22 mA | ||||
Maximum current | 25 mA, short circuit protected | ||||
Maximum input voltage | 50 V DC | ||||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Manual step | Selectable step, change with arrow buttons | ||||
Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | ||||
Environmental Specifications | |||||
Operating temperature | -10°C to +50°C | ||||
Storage temperature | -20°C to +60°C | ||||
Dust/water resistance | Meets IP52, IEC 529 | ||||
Operating altitude | 3000 m above mean sea level (9842 ft) | ||||
Safety Specifications | |||||
Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 | ||||
Mechanical and General Specifications | |||||
Size | 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) | ||||
Weight | 1.2 kg (2.7 lb) | ||||
Batteries | Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh | ||||
Battery life | >8 hours typical | ||||
Battery replacement | Replace without opening calibrator; no tools required | ||||
Side port connections | Pressure module connector | ||||
USB Connector to interface to your PC | |||||
Digital instrument (HART) connector | |||||
Connection for optional battery charger/eliminator | |||||
Data storage capacity | 1 week of calibration procedures results | ||||
90 day specifications | The standard specification interval for the 750 Seriesare 1 and 2 years. | ||||
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. | |||||
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant. | |||||
Temperature, Resistance Temperature Detectors | |||||
Degrees or % of reading - Type (α) | Range °C | Measure °C1 | |||
1 year | 2 year | ||||
100 Ω Pt (385) | -200 to 100 100 to 800 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | ||
200 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | ||
500 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | ||
1000 Ω Pt (385) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | ||
100 Ω Pt (3916) | -200 to 100 100 to 630 | 0.07°C 0.02% + 0.05°C | 0.14°C 0.04% + 0.10°C | ||
100 Ω Pt (3926) | -200 to 100 100 to 630 | 0.08°C 0.02% + 0.06°C | 0.16°C 0.04% + 0.12°C | ||
10 Ω Cu (427) | -100 to 260 | 0.2°C | 0.4°C | ||
120 Ω Ni (672) | -80 to 260 | 0.1°C | 0.2°C | ||
Source current | Source °C | Allowable current2 | |||
1 year | 2 year | ||||
1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | ||
500 μA | 0.06°C 0.017% + 0.05°C | 0.12°C 0.034% + 0.10°C | 0.1 mA to 1 mA | ||
250 μA | 0.06°C 0.017% + 0.05°C | 0.12°C 0.034% + 0.10°C | 0.1 mA to 1 mA | ||
150 μA | 0.06 C 0.017% + 0.05°C | 0.12 C 0.034% + 0.10°C | 0.1 mA to 1 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C | 0.10°C 0.025% + 0.08°C | 0.1 mA to 10 mA | ||
3 mA | 0.2°C | 0.4°C | 0.1 mA to 10 mA | ||
1 mA | 0.04°C | 0.08°C | 0.1 mA to 10 mA | ||
1. For two and three-wire RTD measurements, add 0.4°C to the specifications. 2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms | |||||
Temperature, Thermocouples | |||||
Type | Source °C | Measure °C | Source °C | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000 | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |
Models: Fluke 754 Documenting Process Calibrator-HART
Fluke 754 Documenting Process Calibrator-HART
UPC#: 0 95969 49175 4
Includes:
- BC7240 battery charger
- Li-on BP7240 battery pack
- DPCTrack 2™ Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- Three sets of TP220 test probes
- Three sets of "extended tooth" alligator clips
- Three sets stackable test leads
- Two sets AC280 hook clips
- C799 Soft Field Case
- USB communication cable, Fluke 754HHC HART communications cable
Premium Care 1-Year Standard Plan for Fluke 754 and 753
Includes:
- Premium Care 1-Year Standard Plan
Premium Care 3-Year Standard Plan for Fluke 754 and 753
Includes:
- Premium Care 3-Year Standard Plan
Fluke 754 Bundled with Premium Care 1-Year Standard Plan
Includes:
- Premium Care 1-Year Standard Plan
- Fluke 754
- BC7240 battery charger
- Li-on BP7240 battery pack
- DPCTrack 2™️ Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- Three sets of TP220 test probes with three sets of "extended tooth" alligator clips
- Two sets AC280 hook clips
- C799 Soft Field Case
- USB communication cable, Fluke 754HHC HART communications cable
Is it possible to use this communicator to setup up a HART device? We tried to use this for a device setup but were not able. All that could really be done is read the device outputs which is almost useless.
Manuals + resources: Fluke 754 Documenting Process Calibrator-HART
- 753/754 | Users manual
- 754 Hart | Users manual
- DPC/TRACK | Users manual
- DPC/TRACK | Users manual
- DPCTrack2 | Users manual
- DPCTrack2 | Users manual
- 753/754 | Statement of memory volatility
- 753/754 | Calibration manual
- 753/754 | Calibration manual supplement
- 753/754 | Getting started
- 753/754 | Product specification
- 75X EL | Instruction sheet
- DPC/TRACK | Installation guide
- DPCTrack2 | Getting started