Fluke 9118A Thermocouple Calibration Furnace

  • 9118A Thermocouple Calibration Furnace
  • 9118A Thermocouple Calibration Furnace
  • 9118A Thermocouple Calibration Furnace
  • 9118A Thermocouple Calibration Furnace

Tính năng chính

  • High-performance thermocouple calibrations up to 1200 °C.
  • Horizontal, open-ended tube furnace.
  • Temperature range of 300 °C to 1200 °C.
  • Comparison calibration of noble and base-metal thermocouples by secondary high-temperature labs and instrument shops.
  • Ideal for industries such as aerospace, automotive, energy, metals, and plastics.

Tổng quan sản phẩm: Fluke 9118A Thermocouple Calibration Furnace

Precision Calibration for High-Temperature Thermocouples

The Standard for High-Temperature Thermocouple Calibration

The Fluke 9118A Thermocouple Calibration Furnace delivers unrivaled performance and flexibility for demanding high-temperature calibration labs. Engineered for accuracy, reliability, and ease of use, the 9118A furnace helps you meet industry standards while improving productivity and lowering cost of ownership.

Key Features and Benefits of the Fluke 9118A

  • Wide Temperature Range – Calibrate thermocouples from 300 °C to 1200 °C, covering most high-temperature industrial and laboratory applications.
  • Flexible Configuration – Easily switch between tube furnace and isothermal block configurations to calibrate a wide variety of thermocouple types—accommodating both base-metal and noble-metal designs.
  • Outstanding Uniformity and Stability – Achieve industry-best temperature uniformity and stability:
    • Axial uniformity: ±0.2 °C over a 60 mm zone at 1200 °C
    • Radial uniformity: ±0.25 °C at 1200 °C (isothermal block)
    • Temperature stability: ±0.1 °C or better across the full range
  • Automated Setpoint Control – Boost lab productivity with programmable, automated temperature setpoint control. Supports up to eight setpoints, ramp rates, and dwell times—plus remote automation and data logging with the Fluke 1586A Super-DAQ.
  • Non-Metallic, Contamination-Free Block – Ceramic alumina well and isothermal block minimize thermocouple contamination and reduce ongoing calibration costs.

Thông số kỹ thuật: Fluke 9118A Thermocouple Calibration Furnace

General specifications
Operating conditions
Operating temperature5 °C to 40 °C
Storage temperature–20 °C to 70 °C
Humidity80 % maximum for temperatures <31 °C, decreasing linearly to 50 % at 40 °C
Altitude<2,000 m (6,562 feet)
Power requirements230 V ac (±10 %), 50/60 Hz, 20 A
Heater power4000 watts at 230 V ac
Over-current protection
System20 A, 250 V resettable circuit breaker
Main heater fuseF 12 A, 250 V
Zone heater fuseF 12 A, 250 V
Computer interfaceRS-232 and USB
DisplayMonochrome LCD, °C or °F user-selectable
Display resolution0.1 °C or °F
Size (H x W x L)400 mm x 337 mm x 700 mm (15.7 in x 13.3 in x 27.6 in)
Net weight29 kg (63.9 lb) without isothermal block
Isothermal block (optional)
Isothermal block constructionAlumina
Block outer diameter37 mm (1.5 in)
Block length380 mm (15.0 in)
Well diameter(four wells total) 6.7 mm (0.26 in)
Well depth365 mm (14.4 in)
Net weight0.84 kg (1.9 lb) isothermal block only
Accuracy specifications
All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 °C to 33 °C
The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation.
Furnace temperature range300 °C to 1200 °C
Set-point accuracy± 5 °C
Radial uniformity
Temperature9118A
(14 mm from geometry
central point)
9118A-ITB
(hole to hole)
300 °C± 0.5 °C± 0.1 °C
700 °C± 0.5 °C± 0.20 °C
1200 °C± 0.5 °C± 0.25 °C
Axial uniformity
Temperature9118A
(± 30 mm axial length from
geometry central point)
9118A-ITB
(at 60 mm from full immersion)
Full range± 0.25 °C± 0.2 °C
Temperature stability
Specification9118A9118A-ITB
Stability± 0.2 °C± 0.1 °C
Stabilization time2 hours, full range3 hours at or below 700 °C
2 hours above 700 °C
Note: Temperature stability measured as 2-sigma over 30 minutes
 9118A9118A-ITB
Heating time (23  °C to 1200 °C)40 minutes45 minutes
Cooling time (1200 °C to 300°C)180 minutes200 minutes

Model: Fluke 9118A Thermocouple Calibration Furnace

9118A

Thermocouple Calibration Furnace


Hãy mua ngay
9118-ITB
Thermocouple Calibration Furnace with Isothermal Block 

Hãy mua ngay
Chat với chúng tôiTrợ lý Fluke
Xóa chat