RPM4-E-DWT v1.00c Embedded Software

RPM4-E-DWT v1.00c Embedded Software (.bin)

TITLEVERUSE WITH
E-DWT1.00c
2014.08.06
RPM4-E-DWT
Instructions:
Download the file.

Run Winflash and select the file downloaded as the file to write to the product.

Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

If your product has an model number ending in "-AF", it requires specific "-AF" embedded software.

The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center.

If the update isn't performed properly, your product will be inoperable until it is.
Updates Include:
Ver 1.00c (2014 August 06)
ADD (EDWT-17): Increase limit for Low RPT alert pressure when using High RPT
FIX (EDWT-9): Difference in US and SI version ranges prevent use of full range of the E-DWT
FIX (EDWT-16): Two extra points created when using AutoTest
FIX (EDWT-26): "HEAD=" command and front panel head entry resolution too low

Ver 1.00a (2010 Sept 10)
ADD (EDWT-3): RPM4-E-DWT is now compatible with alternate front panel display hardware.