PG7000 Ver 2.01 Embedded Software

PG7000 Ver 2.01 Embedded Software (.bin)

TITLEVERUSE WITH
PG7000™2.01PG7102, PG7302 or PG7601
Instructions:
TITLEVERUSE WITH
PG7000™2.01PG7102, PG7302 or PG7601

Instructions:
-Ver 2.00 Users
Your PG7000™ time and date will not operate with Ver 2.01 until a hardware change is implemented.

-Ver 1.0x Users
Ver 2.01 is a major revision which should be accompanied by hardware changes. Contact a Fluke Calibration Authorized Service Center for additional information.

Download the file. Run Winflash and select the file downloaded as the file to write to the product.

Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

If your product has an model number ending in "-AF", it requires specific "-AF" embedded software.

The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center.

If the update isn't performed properly, your product will be inoperable until it is.

Updates Include:
2.01
ADDUses new Ver 2.0 function keys and structure.
ADDDifferential mode support (PG7601 only).
CHGPPC1 and PPC2 auto float support improvements.
FIXAuto rotation limits when using AUTO FLOAT mode.
FIXView and edit SETUP file when COM2 has been selected as ambient pressure source.