PG7000 Ver 1.05f Embedded Software

PG7000 Ver 1.05f Embedded Software (.bin)

TITLEVERUSE WITH
PG7000™1.05fPG7102, PG7302 or PG7601
Instructions:
TITLEVERUSE WITH
PG7000™1.05fPG7102, PG7302 or PG7601
Instructions:
Download the file. Run Winflash and select the file downloaded as the file to write to the product.

Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center. If the update isn't performed properly, your product will be inoperable until it is.
Updates Include:
1.05f
ADDSupport for auto-float targets < 3 psia VAC.
FIXAllow recovery from autofloat overshoots that could stop the autofloat execution.
1.05d
CHGCorrected calculation of piston fall rate.
1.05c
ADDIncrease # of PCs from 3 to 18.
ADDCal & temp coefficient fields & remote command.
ADDApplication of the temp coefficient.
ADDCorrected calculation of piston fall rate.
CHGChange max # of masses in a mass group from 10 to 99.
CHGGauge PG label from 7202 to 7102.
1.05b
ADDNew motor control power speed tables for new style board. Eliminate the DIP switch setting for motor type. Use this switch to ID new style boards.
ADDA check for the serial # and query the user if it has been set yet.
ADDFactory reset entry screen & menu function for discrete facory resets (DIAG only).
ADDTEST selection to the DIAG menu to execute the COMx & IEEE production tests.
CHGThe opening credits to show correct model (PG7601, PG7102 or PG7302).
CHGThe platter rotation ready criteria from 45 to 30 rpm for the PGG & PGO.
FIXAIR medium z calculation error. Reset the AutoFloat factory settings when the PG is initialized.