Fluke ii Series Software and Firmware Upgrade

To ensure your ii-Series Acoustic Imager includes all the available features, make sure you have the latest version of operating firmware installed.

In order to get notifications of future upgrades and updates, please register your Fluke ii-Series Acoustic Imager here: Fluke Product Registration. Use Fluke Connect Desktop to automate firmware updates: download Fluke Connect Desktop

Fluke ii900/ii910 (FW5.2 or older):

In order to upgrade your ii900 or ii910 from firmware V5.2 or older to firmware V6.1, a 2-step upgrade path is required by first upgrading to V5.3 and then to V6.1.

Step 1:
1. First download firmware V5.3 here (.swu file)
2. The battery must be charged at least 25%.
3. Use the USB-C cable provided with the unit.
4. Connect the unit to a Windows computer and turn it on. Look for device named FLK-iiXXX on Windows file browser and copy the .SWU file in the root folder ('user data') of the imager.
5. Once the file is copied, your imager will detect it and ask to begin the firmware update. Select 'Start Update'.
6. Once the firmware update is completed, your imager will ask to reboot the unit. Select 'Restart'.

Step 2:
1. Now download firmware V6.1 here (.swu file)
2. Repeat step 2-6 as per the above outlined procedure. After completion, your imager will be updated with the latest firmware.
 

Fluke ii500/ii905/ii915 (FW6.0):
In order to upgrade your ii500, ii905 or ii915 from firmware V6.0 to firmware V6.1:

1. First download firmware V6.1 here (.swu file)
2. The battery must be charged at least 25%.
3. Use the USB-C cable provided with the unit.
4. Connect the unit to a Windows computer and turn it on. Look for device named FLK-iiXXX on Windows file browser and copy the .SWU file in the root folder ('user data') of the imager.
5. Once the file is copied, your imager will detect it and ask to begin the firmware update. Select 'Start Update'.
6. Once the firmware update is completed, your imager will ask to reboot the unit. Select 'Restart'.

Fluke ii-Series Acoustic Imager Firmware version V6.1 update:

 

ii900

ii910

ii500

ii905

ii915

Beacon Mode

 

 

X

X

X

Wi-Fi Captive Portal Support

 

 

 

X

X

Time Zone Settings

X

X

X

X

X

Translation Updates

X

X

X

X

X

Minor bug fixes

X

X

X

X

X

 

ii900

ii910

ii500

ii905

ii915

LeakQ™:
Leak size estimate and cost.

X

X

 

X

X

1-10 LeakQ™ severity index.

X

X

X

X

X

Web based reporting providing CO2 Emission estimation 

X

X

 

X

X

PDQ Mode™:
Provides a relative severity of a discharge, that provides information to the user on whether action may or may not be needed after PD has been detected

 

X

 

 

X

MecQ™:          
Optimized capture mode to locate potential mechanical issues, for example in conveyor systems, so they can be addressed on your maintenance schedule.

 

X

 

 

X

Beacon Mode™:
Optimized capture mode for Fluke SB-140 Sound Beacon

 

 

X

X

X

Please find previous firmware version here:

Version 2.1:

  - 2019 version, no leak-quantification functions

Version 3.0:

  • 2020 version
  • Estimation of distance to the leak.
  • LeakQ™ mode
  • Leak-tags
  • Thumbnail function

Version 4.0:

  • PDQ Mode™ for ii910 Precision Acoustic Imager
  • Grey scale: Show and store image in grayscale
  • Multisource mode to view markers for multiple sources on screen
  • Improved and extended tag annotation
  • Fixes an issue where the camera image could become too dark
  • Improved microphone health diagnosis
  • Notification if battery power is below 25% and does not allow firmware upgrade.
  • Improved text editor

Version 4.2:

  • Digital zoom (1-3x)
  • Swipe in memory
  • Select transparency level of heatmap
  • USB name in Windows File Explorer: FLK-ii900 Series
  • Default selectable profile at startup enabling always same settings at startup

Version 4.3:

  • Issue fixed that could cause USB communication not working properly when many files are stored on the device
  • Added icon to show memory is almost full
  • Added asset tagging using QR code

Version 4.4:

  • Issue fixed that could cause processing power performance
  • Translation updates
  • Issue fixed that could cause a LeakQ reporting error

Version 4.5:

  • New features:
    • Asset inspection status classification: ‘as-found’ or ‘as-left’
    • CO2 emission estimation reporting in Online LeakQ Reporting tool*
  • Issue fixed that could cause:
    • Visual and/or acoustic visualization failure
    • Processing power performance
    • LeakQ reporting error
  • Updates:
    • Various translation updates

Version 4.6:

  • Issue fixed that could cause visual and/or acoustic visualization failure

Version 4.7:

  • On device LeakQ™ quantification (Fluke ii900 & Fluke ii910);
  • Upgrade Visual Light Camera to 5MP (Fluke ii910);
  • Various feature and translation updates.

Version 5.0:

  • MecQ™
    • Easily, efficiently, and contactlessly identify/localize areas of potential mechanical deterioration for further inspection.
  • On-device PDQ Mode™:
    • Partial Discharge Classification.
    • Real time PRPD Plot.
    • PRPD Plot (image).
  • Annotation:
    • Action Annotation features to help our customer document and communicate the urgency of the required follow-up actions.
  • Various:
    • Automatic saving of leak type annotation for LeakQ captures.
    • Issue Fixes and minor feature enhancements.
    • Various translation improvements / updates.
    • Issue fixed that could impact visualization performance in low ambient light conditions. (ii910 only).

Version 5.0.1:

  • Issue fixed that caused incorrect display of flow rate and corresponding cost for leaks when selecting Imperial (CFM) as Unit System.
  • Issue fixed with estimated percentages for Partial Discharge Types.

Version 5.1:

  • Support for changed hardware components in new hardware versions. No functional change compared to V5.0.1.

Version 5.2:

  • Added Auto mode in MecQ™.
  • Added Source dB value to represent the dB level at the source in LeakQ™, PDQ™ and MecQ™ mode. All other dB values are measured dB values at the sensor.
  • Improved manual scaling using sliders on the dB scale.
  • Changed CenterPoint Marker to be ‘ON’ by default.
  • Changed minimum scaling for PRPD plot to prevent noise.
  • Optimized PRPD-plot Y-axis scaling.
  • Fixed issue with incorrect non-metric pressure used for LeakQ.
  • Translations updates.
  • Removed decimal point from min/max dB setting in Acoustics-menu.
  • Added dB-Scale in MecQ mode.
  • Fixed PRPD Plot sometimes not visible.