1594A/1595A Firmware update version 1.22

1594A/1595A Firmware update version 1.22 (.ud)

Title: 1594A & 1595A
Ver: 1.22/1.02
Use with: All 1594A & 1595A
Date Modified: 2/20/2014

Description of the change:  
It was recently determined that when a customer selects the “Restore Factory Default” button in the Config Menu, the resistor calibration values are reset to the default values without requiring a password or notification to the customer that the values are being reset. Engineering has investigated this issue and the instrument firmware has been revised to ensure that the resistor calibration values are not reset to the default values. A password is now required to access the Restore Factory Default capability of the instrument.
Fluke is willing to upgrade your instrument with new firmware at no cost to you. There are three options with regards to upgrading instruments depending on the instrument firmware version. We have verified that upgrading the firmware does not affect the calibration.


Download Requirements: 
A computer with USB port and USB flash drive. 

Instructions:

From your computer:
Download the file named 159x.ud above.
Save the file to your computer.
Save a copy of the file to a USB memory drive.

1594a or 1595a:
Press the menu button
Press the f5 button for the system menu
Press the f3 button for the config
Press the f5 button for the update firmware
Enter the password.  The default is 1234.  If that does not work contact tech support temperaturesupport@flukecal.com for a new password.
Plug the USB memory drive into the 1594a or 1595a.
Press the f5 button for ok
1594a or 1595a will display “please wait…”
1594a or 1595a will display “caution: update the firmware?”
Press the f4 button for “yes”

1594a or 1595a will display the following message:
“the firmware is being updated”
“do not switch off the power”
“please allow 2 min.”

Then, 1594a or 1595a will display the following message:
“operation completed”
“please switch off the power”
“then on again to restart”

Remove the USB memory drive
Turn off the 1594a or 1595a
Turn on the 1594a or 1595a
The update is now complete
The new firmware revision is 1.22
If you have problems with the download or questions about this release, please call customer service at 1.877-355-3225 or email: temperaturesupport@flukecal.com

Updates Include:
Latest firmware

 