Fluke Calibration 9100S Handheld Dry-Well
Key features
- Temperature ranges from –10°C to 375°C.
- Accuracy to ±0.25°C, with stability of ±0.05°C at 0°C.
- Equipped with an RS-232 interface for connectivity.
Sale ends soon.
Shop now!
Product overview: Fluke Calibration 9100S Handheld Dry-Well
9100S Dry-Well
Since we introduced the world's first truly handheld dry-well, many have tried to duplicate it—in vain. Despite its small size (2¼ inches high and 5 inches wide) and light weight, the 9100S outperforms every dry-well in its class in the world.
It's simple and convenient, too. Anyone can learn to use one in less than 15 minutes. It has a range to 375 °C (707 °F) and is perfect for checking RTDs, thermocouples, and small bimetal thermometers in the field.
Plug it in, switch it on, set the temperature with the front-panel buttons, and insert your probe into the properly sized well. Compare the reading of your device to the display temperature or to an external reference, and the difference is the error in your device. With a proprietary Fluke Calibration temperature controller, the 9100S has a display resolution of 0.1 degrees, an accuracy of ±0.5 °C, a stability of ±0.1 °C, and a stabilization time of just five minutes.
Specifications: Fluke Calibration 9100S Handheld Dry-Well
Specifications | 9100S | 9102S |
Range | 35°C to 375°C (95°F to 707°F) | –10°C to 122°C (14°F to 252°F) at 23°C ambient |
Accuracy | ±0.25°C at 50°C; ±0.25°C at 100°C; ±0.5°C at 375°C | ±0.25°C |
Stability | ±0.07°C at 50°C; ±0.1°C at 100°C; ±0.3°C at 375°C | ±0.05°C |
Well-to-well uniformity | ±0.2°C with sensors of similar size at equal depths within wells | |
Heating times | 35°C to 375°C: 9.5 minutes | Ambient to 100°C: 10 minutes |
Stabilization | 5 minutes | 7 minutes |
Cooling times | 375°C to 100°C: 14 minutes | Ambient to 0°C: 10 minutes |
Well depth | 102 mm (4 in); 1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep | 102 mm (4 in) |
Removable inserts | N/A | Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included] |
Power | 115 V AC (±10%), 55–65 Hz, 1.5 A or 230 V AC (±10%), 0.8 A, 45–55 Hz, 175 W | 94-234 V AC (±10%), 50/60 Hz, 60 W; or 12 V DC |
Size (H x W x D) | 57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) | 99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in) |
Weight | 1 kg (2 lb 3 oz) | 1.8 kg (4 lb) |
Computer interface | RS-232 included | |
NIST-traceable calibration | Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C and 375°C | Data at –10°C, 0°C, 25 °C, 50°C, 75 °C,100°C and 122°C |
Models: Fluke Calibration 9100S Handheld Dry-Well
Fluke Calibration 9100S-A Handheld Dry-Well
HDRC Handheld Dry-Well A
For 9100S with other inserts below, please visit request a quote.
UPC#: 0 95969 01783 1
HDRC Handheld Dry-Well B
UPC#: 0 95969 20560 3
HDRC Handheld Dry-Well C
UPC#: 0 95969 01787 9
HDRC Handheld Dry-Well D
UPC#: 0 95969 01789 3