molbloc-L Laminar Flow Element for molbox2 only

Model#: 1E1-L-2Part#: 6084512
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 142.16ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • molbloc-L Laminar Flow Element
  • molbloc-L Laminar Flow Element
  • molbloc-L Laminar Flow Element

Key features

  • Compatible ONLY with molbox2
  • Flow ranging from 1 sccm to 100 slm.
  • Molbloc flow path is the gap between a piston and a close fitting cylindrical bore in the molbloc body.
  • Different flow ranges are achieved by varying the size of the gap, with larger bodies used for higher flow ranges.

Product overview: molbloc-L Laminar Flow Element for molbox2 only

Features at a glance

  • Compatible ONLY with molbox2
  • Useable with existing molbox1+ and molbox RFM mass flow terminals and COMPASS software
  • Integrated filter to protect against contamination
  • Integral gas temperature conditioning and measurement
  • Continuous, real-time reading
  • Can be fully automated
  • No moving parts

molbloc-L Ranges
molblocs sizes are designated by a number relative to their nominal flow impedance. The nominal impedance results in nominal flow ranges when flowing nitrogen gas with 250 kPa absolute (36 psia) upstream pressure and 50 kPa (7.5 psid) differential pressure across the molbloc. In practice, the molbloc flow range may be quite different from the nominal flow range. The difference comes from differences in density and viscosity between gases and from different molbloc operating pressure calibration options.

Specifications: molbloc-L Laminar Flow Element for molbox2 only

Specifications - Standard molbloc-L Calibrations - Used with molbox1+ and molbox2 terminals
Measurement Update Rate1 second
Range0 to molbloc full scale depending on gas and molbloc pressure
dependent calibration type (see Section 1.2.5.1.3)
Resolution0.0015 % FS
Linearity± 0.05 % of reading from 10 to 100 % FS,
± 0.005 % FS under 10 % FS
Repeatability± 0.05 % of reading from 10 to 100 % FS,
± 0.005 % FS under 10 % FS
Precision1± 0.07 % of reading from 10 to 100 % FS,
± 0.007 % FS under 10 % FS
Stability2
(1 year)
± 0.09 % of reading from 10 to 100 % FS,
± 0.009 % FS under 10 % FS
Measurement Uncertainty3± 0.2 % of reading from 10 to 100 % FS,
± 0.02 % FS under 10 % FS
(N2 and any molbox1+ supported gas for which the molbloc in use is calibrated)1E5 molbloc
± 0.5% of reading from 25 to 100 % FS,
± 0.125 % FS under 25 % FS
1 Precision: Combined linearity, hysteresis, repeatability.
2 Stability: Maximum change in zero and span over specified time period for typical molbox1+ and molbloc used under typical conditions. As stability can only be predicted, stability for a specific molbloc and molbox1+ should be established from experience.
3 Measurement uncertainty (accuracy): Maximum deviation of the molbox1+ flow indication from the true value of the flow through the molbloc including precision, stability and Fluke calibration standard measurement uncertainty.
Specifications - Premium molbloc-L Calibrations - Used with molbox1+ terminal
Measurement Update Rate1 second
Range0 to molbloc full scale depending on gas and molbloc pressure dependent calibration type (see Section 1.2.5.1.2)
Resolution0.0015 % FS
Linearity± 0.05 % of reading from 10 to 100 % FS,
± 0.005 % FS under 10 % FS
Repeatability± 0.05 % of reading from 10 to 100 % FS,
± 0.005 % FS under 10 % FS
Precision1± 0.07 % of reading from 10 to 100 % FS,
± 0.007 % FS under 10 % FS
Stability2
(1 year)
± 0.03 % of reading from 10 to 100 % FS,
± 0.003 % FS under 10 % FS
Measurement Uncertainty3± 0.125 % of reading from 10 to 100 % FS,
± 0.0125 % FS under 10 % FS
(N2 and any molbox1+ supported gas for which the molbloc in use is calibrated)1E5 molbloc-L Premium calibration not available
1 Precision: Combined linearity, hysteresis, repeatability.
2 Stability: Maximum change in zero and span over specified time period for typical molbox1+ and molbloc used under typical conditions. As stability can only be predicted, stability for a specific molbloc and molbox1+ should be established from experience.
3 Measurement uncertainty (accuracy): Maximum deviation of the molbox1+ flow indication from the true value of the flow through the molbloc including precision, stability and Fluke calibration standard measurement uncertainty.

Models: molbloc-L Laminar Flow Element for molbox2 only

1E1-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084512
5E1-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084596
1E2-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084583
2E2-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084577
5E2-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084565
1E3-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084633
5E3-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084622
1E4-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084614
3E4-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084605
1E5-L-2

LAMINAR MOLBLOC FLOW ELEMENT

Part#: 6084554
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 128.39ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 24.35ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.
Chat with ourFluke assistant
Clear Chat