4950 Multifunction Transfer Standard
Discontinued
This product is discontinued or reclassified
Fluke suggests:5790B AC Measurement Standard
Product overview: 4950 Multifunction Transfer Standard
Specifications: 4950 Multifunction Transfer Standard
| General Specifications | |
| Voltage DC | Range : 0V to ± 1000V Best Transfer Stability: 2.1 ppm |
| Voltage AC | Range : 1 mV to 1000V 10 Hz to 1 MHz Best Transfer Stability: 14 ppm |
| Current DC | Range : 0 to ± 10 Amps Best Transfer Stability : 9.8 ppm |
| Current AC | Range : 10 Hz to 20 kHz 100 mA to 10A Best Transfer Stability : 56 ppm |
| Resistance | Range : 1Ω to 100 MΩ Best Transfer Stability : 4.1 ppm |
| Line Supply | Power supply 90V to 145V or 187V to 292V (selectable from rear panel) 46 Hz to 66 Hz |
| Power Consumption | 37 VA maximum |
| Dimensions | (H x W x D) 3.46 ins x 16.8 ins x 18.9 ins |
| Weight | 13.5 kg (30 lbs) |
| Interface Compatibility | IEEE 488.1 for electrical interface IEEE 488.2 for syntax and protocols |
| External Current Shunt | A 10A current shunt (Model 4953) is supplied with every 4950. The characteristics of this shunt and its serial number will be entered into the 4950 during calibration. |
| Signal Input Lead | A signal input lead will be provided with the instrument. The characteristics of this and its serial number will be entered into the 4950 during calibration. |
| Safety | Designed to UL 1244 and IEC 348 |
| Warranty * | 1 year |
*All stabilities are based on 30 days tcal ± 1°C