PM700 Pressure Measurement Modules
Key features
- 0.008% reading measurement uncertainty from 30% to 100% of the modules’ span
- Quartz Reference Pressure Transducer (Q-RPT) technology provides precise measurement with long term stability
- Onboard barometer included with absolute mode pressure modules
- Ability to be used in both absolute and gauge mode
Product overview: PM700 Pressure Measurement Modules
The Fluke Calibration PM700 Pressure Measurement Modules deliver laboratory-grade pressure measurement performance with the flexibility required for modern calibration and test environments. Built on Fluke's proven Quartz Reference Pressure Transducer (Q-RPT) technology, the PM700 provides exceptional accuracy, long-term stability, and broad workload coverage, making it ideal for use with Fluke 6270A, 8270A, and 8370A Pressure Controller/Calibrators. With measurement uncertainty as low as 0.008% of reading across a wide operating range, the PM700 helps calibration labs improve confidence, streamline operations, and reduce the number of modules needed to cover diverse pressure applications. Absolute pressure models include an onboard barometer, enabling both absolute and gauge measurements from a single module for added versatility.
Specifications: PM700 Pressure Measurement Modules
Performance specifications
The performance specifications describe the complete instrumental uncertainty of the Product. The specifications include all relevant error components (linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1-year drift, and temperature effects). The specifications are provided at a level of confidence of 95 %, k=2, normally distributed. Precision uncertainty includes linearity, hysteresis, repeatability, resolution, and temperature effects.
PM700 Modules measurement specifications
Specifications are valid from 15 °C to 35 °C.
| Model | Absolute Mode Range | Absolute Mode Range (Imperial Units) | Gauge Mode Range 3 (SI Units) | Gauge Mode Range (Imperial Units) | 1-Year Instrumental Uncertainty (% of reading or % FS, whichever is greater) | Precision Uncertainty (% of reading or % FS, whichever is greater) |
| PM700-BG15KP | - | - | -15 kPa to 15 kPa | -60 inH2O to 60 inH2O | 0.008 or 0.0024 | 0.005 or 0.0015 |
| PM700-G100KP | - | - | 0 kPa to 100 kPa | 0 psi to 15 psi | 0.008 or 0.0024 | 0.005 or 0.0015 |
| PM700-G200KP | - | - | 0 kPa to 200 kPa | 0 psi to 30 psi | 0.008 or 0.0024 | 0.005 or 0.0015 |
| PM700-A700KP | 18 kPa to 700 kPa | 2.6 psi to 100 psi | -82 kPa to 700 kPa | -12.1 psi to 100 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A1.4MP | 0.035 MPa to 1.4 MPa | 5 psi to 200 psi | -0.065 MPa to 1.4 MPa | -10 psi to 200 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A2MP | 0.07 MPa to 2 MPa | 10 psi to 300 psi | -0.03 MPa to 2 MPa | -5 psi to 300 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A3.5MP | 0.07 MPa to 3.5 MPa | 10 psi to 500 psi | -0.03 MPa to 3.5 MPa | -5 psi to 500 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A7MP | ATM2 to 7 MPa | ATM2 to 1000 psi | 0 MPa to 7 MPa | 0 psi to 1000 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A20MP | ATM2 to 20 MPa | ATM2 to 3000 psi | 0 MPa to 20 MPa | 0 psi to 3000 psi | 0.008 or 0.00241 | 0.005 or 0.0015 |
| PM700-A28MP | ATM3 to 28 MPa | ATM3 to 4000 psi | 0 MPa to 28 MPa | 0 psi to 4000 psi | 0.008 or 0.00242 | 0.005 or 0.0015 |
| PM700-A35MP | ATM3 to 35 MPa | ATM2 to 5000 psi | 0 MPa to 35 MPa | 0 psi to 5000 psi | 0.008 or 0.00242 | 0.005 or 0.0015 |
| PM700-A40MP | ATM3 to 40 MPa | ATM3 to 6000 psi | 0 MPa to 40 MPa | 0 psi to 6000 psi | 0.008 or 0.00242 | 0.005 or 0.0015 |
| Notes: | ||||||
1. For PM700s absolute mode modules used in absolute mode, root sum square (RSS) with 0.007 % of FS (reduced to k=1 by square root of 3). ![]() This specification may be reduced or eliminated with the use of a separate more stable module as an AutoZero reference. 2. PM700 module with full scales of 28 MPa and higher use an internal barometer in the PMM to correct for changes in barometric pressure when they are used in gauge mode and as a zeroing reference when used in absolute mode, hence there is no need to RSS 0.007 % FS. ATM is any atmospheric pressure from 70 kPa to 110 kPa (10 psi to 16 psi). 3. For absolute ranges used in gauge mode there is an additional uncertainty of ±14 Pa for dynamic barometric compensation. 4. These modules can be used as a barometric reference module on all chassis. 5. Compatible with 6270A 6. Compatible with 6270A and 8270A 7. Compatible with 6270A, 8270A and 8370A 8. Compatible with 8270A and 8370A | ||||||
Models: PM700 Pressure Measurement Modules
PRESSURE MEAS MODULE,-15 TO 15 KPA(-2.2 TO 2.2 PSI)GAUGE PREM
UPC#: 1 95112 17489 0
PRESSURE MEAS MODULE,0 TO 100 KPA (0 TO 15 PSI) GAUGE PREM
UPC#: 1 95112 17490 6
PRESSURE MEAS MODULE,0 TO 200 KPA (0 TO 30 PSI) GAUGE PREM
UPC#: 1 95112 17491 3
PRESSURE MEAS MODULE,700 KPA(100 PSI)ABSOLUTE PREM
UPC#: 1 95112 17492 0
PRESSURE MEAS MODULE,1.4 MPA(200 PSI)ABSOLUTE PREM
UPC#: 1 95112 17493 7
PRESSURE MEAS MODULE,2 MPA (300 PSI) ABSOLUTE PREM
UPC#: 1 95112 17494 4
PRESSURE MEAS MODULE,7 MPA (1000 PSI) ABSOLUTE PREM
UPC#: 1 95112 17496 8
PRESSURE MEAS MODULE,20 MPA (3000 PSI) ABSOLUTE PREM
UPC#: 1 95112 17497 5
PRESSURE MEAS MODULE,28 MPA (4000 PSI) ABSOLUTE PREM
UPC#: 1 95112 17498 2
PRESSURE MEAS MODULE,40 MPA (6000 PSI) ABSOLUTE PREM
UPC#: 1 95112 17500 2