Verifying portable thermometers

Calibration

On-the-spot temperature checks using portable thermometers are frequently performed in the field by technicians in order to ensure product quality. It is a best practice to verify the portable thermometer, either liquid-filled or electronic, against an accurate reference before beginning rounds to take those spot checks.

Using a Reference Thermometer for VerifyingVerify Portable Thermometers

Ideally, the thermometer will be checked as close as possible to the expected temperature measurement. For example, the expected temperature of the gas in a pipeline. Personal electronic thermometers (PETs) can be verified in the field with a portable dry well. The dry well is set to the expected temperature and the thermometer inserted to verify the reading.

Thermometers can also be checked in an office, laboratory or, if in a remote area, at a tool shed with a dry-block or microbath calibrator.

Dry blocks aren’t suitable for liquid-filled glass thermometers so the accuracy check is performed using a constant temperature fluid bath, such as the Fluke Calibration 6330.

If the error of a liquid filled thermometer is large, it is best to discard the thermometer. Another option if the thermometer is slightly off is to write a correction table and adjust the reading accordingly.

Some portable electronic thermometers can be recalibrated if the readings are off, though others must be replaced.

Hi, I'm your virtual assistant, I will do my very best to help you navigate all things Fluke related. You can ask me anything! I'm new and might make a few mistakes. For safe use of Fluke tools, refer to the product manuals, safety sheets, and your company's safety requirements. To speak with a person, visit our Contact Us or Talk to Sales page. For information on how we handle personal data, see our Privacy Policy.