Fluke 729 Pro Automatic Pressure Calibrator
Nøkkelfunksjoner
- Automatic pressure generation and regulation to 1000psi
- Replaceable pressure control modules
- Built-in test templates, one-key calibration
- HART communication
- Pressure leakage testing
- Automatic fine-pressure adjustment, automatic pressure stabilization
- Powerful, replaceable Li-ion battery
- Quick pressure release valve with a muffler
- Extend measurement range with 750P Series pressure modules
- Measure temperature with 720RTD probe
- Bright dual-three-channel /color graphic display
- Weight <5kg
- Rugged and portable design with standard 3-year warranty
- Absolute and gage pressure modules up to 7MPa.
- Use the modules to change the Fluke 729Pro to a different output pressure range
Produktoversikt: Fluke 729 Pro Automatic Pressure Calibrator
Bærbar automatisk trykkalibrator forenkler trykkalibrering
Den automatiske trykkalibratoren Fluke 729 Pro er utformet spesielt for prosessteknikere for å forenkle trykkalibreringsprosessen og gi raskere og mer nøyaktige testresultater. Trykkalibrering kan være en tidkrevende oppgave. Med 729 Pro blir det enklere takket være en intern, elektrisk pumpe som gir automatisk trykkgenerering og -regulering. I tillegg er den brukervennlig, solid og bærbar.
Automatisk trykkgenerering og -regulering
Med 729 Pro trenger du bare å angi ønsket trykk, så pumper kalibratoren automatisk opp til det. Den har utskiftbare trykkmoduler samt trykkgenerering og -regulering opp til 1000 psi (70 bar, 7000 kPa). En intern finjusteringskontroll stabiliserer trykket automatisk etter den inntastede verdien. Du kan også fylle inn en testmal, så vil 729 Pro med et enkelt tastetrykk automatisk pumpe opp til, og dokumentere, en flerpunkts trykkalibreringstest.
Enkel kalibreringsdokumentasjon
Fluke 729 Pro kan teste flere trykkpunkter og dokumentere resultatene automatisk. Så tar 729 Pro seg av jobben med å dokumentere og flagge testresultater som ligger utenfor toleransegrensene, med rødt på den grafiske visningen av
- anvendt trykk
- målt mA, mV eller digital PV
- %-vis avvik for hvert testpunkt
Last opp og administrer dokumenterte kalibreringsresultater med programvaren for kalibreringsadministrasjon, DPCTrack2™. Dette vil gjøre det enkelt for deg å administrere instrumenteringen, opprette planlagte tester og rapporter samt administrere kalibreringsdata.
HART-kommunikasjon
Innebygget HART-kommunikasjon muliggjør mA-justeringer på HART-transmittere, HART-konfigurering og justering i henhold til anvendte 0 %- og 100 %-verdier. Du kan også konfigurere ulike oppgaver som å endre transmitteretiketter og måleenheter og -områder. Andre støttede HART-kommandoer er innstilling av faste mA ut-verdier for feilsøking, leserkonfigurering, samt diagnostikk av variabler og lesere.
Robust bærbar utforming
729 Pro er testet for og kan garantert tåle et fall fra én meters høyde, og er egnet for kalibrering av instrumenter i felten. Enten du velger modellen 4M opp til 600 psi (40 bar, 4000 kPa) eller 7M opp til 1000 psi (70 bar, 7000 kPa) er Fluke 729 Pro automatiske trykkalibratorer konstruert for å kunne brukes når som helst og hvor som helst. En halvstiv veske følger med. Den er laget for feltarbeid og har plass til store testslanger, koblinger, testledninger og små håndverktøy, så det er enkelt å få med seg alt som trengs til trykkalibrering i felten.
For høyere nøyaktighet og oppløsning på lavtrykkalibreringer kan du med Fluke 729 Pro sette inn en av seks Fluke-trykkmoduler for å utvide trykkområdet mellom 0 og 600 psi (40 bar, 4000 kPa).
Spesifikasjoner: Fluke 729 Pro Automatic Pressure Calibrator
Key parameters | 729 Pro 7M | 729 Pro 4M |
Pressure control range | -13 to 1000 psi, -90 to 7000 kPa, -0.90 to 70 bar | -13 to 600 psi, -90 to 4000 kPa, -0.90 to 40 bar |
Pressure type | Gauge/absolute (Depends on the type of internal or external pressure module | |
Pressure accuracy and Control stability | 0.02% of FS, 0.005% of FS | |
Replaceable pressure measurement modules | Yes, 6 modules | Yes, 4 modules |
mA sourcing/measurement | Yes | |
Voltage measurement | Yes | |
Loop power | 24 V dc | |
Pressure switch testing | Yes | |
Pressure leakage testing | Yes | |
HART Communication | Yes, basic commands supported | |
Autostep / auto ramp | Yes (with customizable speed, mode and style) | |
Battery | Rechargeable lithium battery with charge indication (charged in the instrument or separately) | |
Language | English |
Electrical specification | |
All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range. | |
Ranges | |
mA measure, source and simulate | 0 mA to 24 mA |
Volts dc measurement | 0 V dc to 30 V dc |
Resolution | |
mA dc source, simulate and measure | 1 μA |
Voltage dc measurement | 1 mV |
Accuracy | 0.01% FS ±2 LSD all ranges (at 23 °C ±5 °C) |
Stability | 20 ppm FS /°C (-10 °C to +18 °C and 28 °C to 50 °C) |
mA simulation working voltage | 12 V dc to 30 V dc |
Loop compliance voltage | 24 V dc @ 20 mA |
Temperature measurement , 100 Ω Pt(385) RTD | -50 °C to 150 °C |
Temperature resolution | 0.01 °C |
Temperature accuracy | ±0.1 °C ±0.25 °C combined uncertainty when using 720 RTD probe (optional accessory) |
Drive capability (mA source) | 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |
Mechanical specification | |
Dimension (HxWxL) | 290 x 215 x100 mm, (4.3 x 4.4 x 9.4 in) |
Weight | 4.9 kg, (10.8 lbs) |
Environmental specification | |
Operating temperature | -10 °C to 50 °C |
Pressure control | 0 °C to 50 °C |
Battery charging | 0 °C to 40 °C |
Storage temperature | -20 °C to 60 °C |
Operating altitude | <3000 m |
Storage altitude | <13000 m |
Operating humidity | Non-condensing (<10°C) ≤90 % RH (10 °C to 30 °C) ≤75 % RH (30 °C to 40 °C) ≤45 % RH (40 °C to 50 °C) |
Pressure Module Selection Table | ||||||
Pressure module | Pressure type | Range | Resolution | Overall uncertainty | 729Pro 4M | 729Pro 7M |
FLK-PMM-200K | Gauge | (-15 to 30) psi | 0.0001 psi | 0.02 % FS / year, maximum | • | • |
(-100 to 200) kPa | 0.001 kPa | |||||
(-1 to 2) bar | 0.00001 bar | |||||
FLK-PMM-200KA | Absolute | (0 to 30) psi | 0.0001 psi | 0.05 % FS / year, maximum | • | • |
(0 to 200) kPa | 0.001 kPa | |||||
(0 to 2) bar | 0.00001 bar | |||||
FLK-PMM-2000K | Gauge | (-14 to 300) psi | 0.001 psi | 0.02 % FS / year, maximum | • | • |
(-95 to 2000) kPa | 0.01 kPa | |||||
(-0.95 to 20) bar | 0.0001 bar | |||||
FLK-PMM-4000K | Gauge | (-14 to 600) psi | 0.01 psi | 0.02 % FS / year, maximum | • | • |
(-95 to 4000) kPa | 0.01 kPa | |||||
(-0.95 to 40) bar | 0.0001 ba | |||||
FLK-PMM-7000K | Gauge | (-14 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(-95 to 7000) kPa | 0.01 kPa | |||||
(-0.95 to 70) bar | 0.0001 ba | |||||
FLK-PMM-7000KA | Absolute | (0 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(0 to 7000) kPa | 0.01 kPa | |||||
(0 to 70) bar | 0.0001 ba | |||||
Dry air or non-corrosive gases only / 15 °C to 35 °C, all ranges |