Fluke 729 Pro Automatic Pressure Calibrator
Tärkeimmät ominaisuudet
- Automatic pressure generation and regulation to 1000psi
- Replaceable pressure control modules
- Built-in test templates, one-key calibration
- HART communication
- Pressure leakage testing
- Automatic fine-pressure adjustment, automatic pressure stabilization
- Powerful, replaceable Li-ion battery
- Quick pressure release valve with a muffler
- Extend measurement range with 750P Series pressure modules
- Measure temperature with 720RTD probe
- Bright dual-three-channel /color graphic display
- Weight <5kg
- Rugged and portable design with standard 3-year warranty
- Absolute and gage pressure modules up to 7MPa.
- Use the modules to change the Fluke 729Pro to a different output pressure range
Tuotekatsaus: Fluke 729 Pro Automatic Pressure Calibrator
Kannettava automaattipainekalibraattori tekee painekalibroinnista yksinkertaista
Fluke 729 Pro -automaattipainekalibraattori on suunniteltu erityisesti prosessiteknikoiden tarpeisiin. Sen ansiosta painekalibrointi on yksinkertaista ja testaus nopeampaa ja tarkempaa. Teknikot tietävät hyvin, että painekalibrointi vie paljon aikaa. Fluke 729 Pro helpottaa kalibrointia, sillä sen sisäinen sähköpumppu takaa automaattisen paineentuoton ja -säädön. Lisäksi se on helppokäyttöinen, kestävä ja kevyt kantaa mukana.
Automaattinen paineensyöttö ja -hallinta
729 Pro on ihanteellinen kannettava painekalibraattori, sillä käyttäjän tarvitsee vain määrittää tavoitepaine, ja kalibraattori asettaa paineen automaattisesti halutulle tasolle. Kalibraattorissa on pistoliitettävät paineensäätömoduulit ja paineensyöttö- ja säätö 1 000 psi:n (70 bar, 7 000 kPa) paineeseen asti. Lisäksi sisäinen hienosäädin vakauttaa paineen automaattisesti määritettyyn asetusarvoon. Voit myös täyttää testauspohjan, jolloin 729 Pro pumppaa automaattisesti ja dokumentoi usean pisteen painekalibrointitestin painiketta painamalla.
Helppo kalibroinnin dokumentointi
Fluke 729 Prolla voi tehdä automaattisia, useiden testipisteiden painetestauksia, joiden tulokset laite myös dokumentoi automaattisesti. 729 Pro tekee työn puolestasi dokumentoimalla ja ilmoittamalla seuraavat toleranssin ulkopuoliset testitulokset punaisella graafisessa näytössä:
- Käytetty paine
- Mitattu mA, mV tai digitaalinen PV
- Virhe-% jokaiselle testipisteelle
Käyttäjä voi ladata ja hallinnoida dokumentoituja kalibrointituloksia DPCTrack2™-kalibroinninhallintaohjelmistolla (CMS), mikä helpottaa instrumentoinnin ja kalibrointitietojen hallintaa sekä ajastettujen testausten ja raporttien laadintaa.
HART-kommunikointi
Sisäisten HART-kommunikointitoimintojen avulla voidaan tehdä HART-lähettimen mA-säätöjä, HART-määrityksiä ja säätää arvoja tasolle 0 % ja 100 %. Käyttäjä voi myös tehdä määrityksiä, kuten vaihtaa lähettimen tunnistetta, mittayksiköitä ja aluetta. Muut tuetut HART-komennot ovat kiinteiden mA-arvojen määritys vianhakua varten, laitemääritysten ja -muuttujien lukeminen sekä laitediagnostiikan lukeminen.
Kestävä kannettava rakenne
729 Pro on testattu ja varmistettu takuuta varten siten, että se kestää pudotuksen yhden metrin korkeudelta. Siksi se sopii täydellisesti kenttäkalibrointityöhön. Fluke 729 Pro -automaattipainekalibraattorit toimivat juuri silloin ja siellä, missä niitä tarvitaan. Valittavana on kaksi eri mallia: 4M 600 psi:n (40 bar, 4 000 kPa) ja 7M 1 000 psi:n (70 bar, 7 000 kPa) paineeseen asti. Vakiokantolaukku sopii kenttäkäyttöön. Siinä on tilaa testijohtojen ja -sovittimien, mittausjohtojen ja pienten käsityökalujen säilyttämiseen, joten voit pitää mukanasi kaiken tarvittavan kenttäpainekalibrointityötä varten.
Tarkempaa ja paremman erottelukyvyn matalapainekalibrointia varten voit asettaa Fluke 729 Pro -kalibraattoriin yhden kuudesta eri Fluke-painemoduulista, joiden avulla painealue laajenee 0–600 psi:n (40 bar, 4 000 kPa) välillä.
Tekniset tiedot: Fluke 729 Pro Automatic Pressure Calibrator
Key parameters | 729 Pro 7M | 729 Pro 4M |
Pressure control range | -13 to 1000 psi, -90 to 7000 kPa, -0.90 to 70 bar | -13 to 600 psi, -90 to 4000 kPa, -0.90 to 40 bar |
Pressure type | Gauge/absolute (Depends on the type of internal or external pressure module | |
Pressure accuracy and Control stability | 0.02% of FS, 0.005% of FS | |
Replaceable pressure measurement modules | Yes, 6 modules | Yes, 4 modules |
mA sourcing/measurement | Yes | |
Voltage measurement | Yes | |
Loop power | 24 V dc | |
Pressure switch testing | Yes | |
Pressure leakage testing | Yes | |
HART Communication | Yes, basic commands supported | |
Autostep / auto ramp | Yes (with customizable speed, mode and style) | |
Battery | Rechargeable lithium battery with charge indication (charged in the instrument or separately) | |
Language | English |
Electrical specification | |
All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range. | |
Ranges | |
mA measure, source and simulate | 0 mA to 24 mA |
Volts dc measurement | 0 V dc to 30 V dc |
Resolution | |
mA dc source, simulate and measure | 1 μA |
Voltage dc measurement | 1 mV |
Accuracy | 0.01% FS ±2 LSD all ranges (at 23 °C ±5 °C) |
Stability | 20 ppm FS /°C (-10 °C to +18 °C and 28 °C to 50 °C) |
mA simulation working voltage | 12 V dc to 30 V dc |
Loop compliance voltage | 24 V dc @ 20 mA |
Temperature measurement , 100 Ω Pt(385) RTD | -50 °C to 150 °C |
Temperature resolution | 0.01 °C |
Temperature accuracy | ±0.1 °C ±0.25 °C combined uncertainty when using 720 RTD probe (optional accessory) |
Drive capability (mA source) | 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |
Mechanical specification | |
Dimension (HxWxL) | 290 x 215 x100 mm, (4.3 x 4.4 x 9.4 in) |
Weight | 4.9 kg, (10.8 lbs) |
Environmental specification | |
Operating temperature | -10 °C to 50 °C |
Pressure control | 0 °C to 50 °C |
Battery charging | 0 °C to 40 °C |
Storage temperature | -20 °C to 60 °C |
Operating altitude | <3000 m |
Storage altitude | <13000 m |
Operating humidity | Non-condensing (<10°C) ≤90 % RH (10 °C to 30 °C) ≤75 % RH (30 °C to 40 °C) ≤45 % RH (40 °C to 50 °C) |
Pressure Module Selection Table | ||||||
Pressure module | Pressure type | Range | Resolution | Overall uncertainty | 729Pro 4M | 729Pro 7M |
FLK-PMM-200K | Gauge | (-15 to 30) psi | 0.0001 psi | 0.02 % FS / year, maximum | • | • |
(-100 to 200) kPa | 0.001 kPa | |||||
(-1 to 2) bar | 0.00001 bar | |||||
FLK-PMM-200KA | Absolute | (0 to 30) psi | 0.0001 psi | 0.05 % FS / year, maximum | • | • |
(0 to 200) kPa | 0.001 kPa | |||||
(0 to 2) bar | 0.00001 bar | |||||
FLK-PMM-2000K | Gauge | (-14 to 300) psi | 0.001 psi | 0.02 % FS / year, maximum | • | • |
(-95 to 2000) kPa | 0.01 kPa | |||||
(-0.95 to 20) bar | 0.0001 bar | |||||
FLK-PMM-4000K | Gauge | (-14 to 600) psi | 0.01 psi | 0.02 % FS / year, maximum | • | • |
(-95 to 4000) kPa | 0.01 kPa | |||||
(-0.95 to 40) bar | 0.0001 ba | |||||
FLK-PMM-7000K | Gauge | (-14 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(-95 to 7000) kPa | 0.01 kPa | |||||
(-0.95 to 70) bar | 0.0001 ba | |||||
FLK-PMM-7000KA | Absolute | (0 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(0 to 7000) kPa | 0.01 kPa | |||||
(0 to 70) bar | 0.0001 ba | |||||
Dry air or non-corrosive gases only / 15 °C to 35 °C, all ranges |