The 9144 Field Metrology Well has replaced the 9141 Dry-Well
The Fluke Calibration 9141 Dry-Well is now obsolete and has been replaced with the 9144 Field Metrology Well.
Top reasons to consider the 9144 Field Metrology Well over the 9141 Dry-Well
- Better accuracy, stability, and radial uniformity
 - Dual heater zones improve vertical temperature gradient
 - Greater immersion depth
 - Modern user interface and industrial design
 - Process Electronics option to read external reference probes and UUTs
 
Comparing the 9141 Dry-Well with the 9144 Field Metrology Well (with and without Process Electronics)
9141 Dry-Well  | 9144 Field Metrology Well  | 9144-P Field Metrology Well with Process Electronics  | |
Built-in thermometer readout  | No  | No  | Yes  | 
Temperature range  | 50 °C to 650 °C  | 50 °C to 660 °C  | 50 °C to 660 °C  | 
Accuracy (at max)  | ± 1.0 °C  | ± 0.5 °C  | ± 0.07 °C  | 
Stability (at max)  | ± 0.12 °C  | ± 0.05 °C  | ± 0.05 °C  | 
Radial uniformity (at max)  | ± 0.12 °C  | ± 0.10 °C  | ± 0.10 °C  | 
Axial uniformity (at max)  | No spec  | ± 0 5 °C  | ± 0.5 °C  | 
Heating speed (C/min)  | 52  | 41  | 41  | 
Cooling speed (C/min)  | 22  | 17  | 17  | 
Immersion depth (mm)  | 124  | 150  | 150  | 