5500A-SC600 | 300 MHz Oscilloscope Calibration Option
Discontinued
This product is discontinued or reclassified
Product overview: 5500A-SC600 | 300 MHz Oscilloscope Calibration Option
Specifications: 5500A-SC600 | 300 MHz Oscilloscope Calibration Option
General Specifications | |
DC Voltage | Ranges: 0 to ±6.6V (50Ω) 0 to ±33V (1 MΩ) Best 1 Year Specification: ±0.05% |
AC Voltage (Squarewave) | Ranges: ±1 mV to ±6.6V p-p (50Ω) ±1 mV to ±130V p-p (1 MΩ) Best 1 Year Specification: ±0.25% |
Fast Edge | Ranges: 4 mV to 2.5V p-p (50Ω) Best 1 Year Specification: =300 ps Rise Time |
Leveled Sinewave | Ranges: 50 kHz to 600 MHz 50 kHz to 300 MHz Best 1 Year Specification: ±1.5% Flatness |
Time Markers | Ranges: 5s to 2 ns Spike, square, 20% pulse, sine Best 1 Year Specification: ±2.5 ppm |
Wave Generator | Ranges: 10 mV to 2.5V amplitude1.8 mV to 55V p-p (1 MΩ) 1.8 mV to 2.2V p-p (50Ω) 10 Hz to 100 kHz Square, Sine, Triangle Best 1 Year Specification: ±3% |
Pulse Generator | Ranges: 10 mV to 2.5V amplitude 4 ns to 500 ns width 20 ms to 150 ns period Best 1 Year Specification: ±5% ±2.5 ppm |
TV Trigger | NTSC, SECAM, PAL, PAL-M |
Input Impedance Measurement | Range: 40Ω to 1.5 MΩ Best 1 Year Specification: ±0.1% ±0.1% Range: 5 pF to 50 pF Best 1 Year Specification: ±5% |
Overload Measurement | 5V to 9V (dc or ac squarewave), 5 to 60 seconds |
Models: 5500A-SC600 | 300 MHz Oscilloscope Calibration Option
5500A-SC600
600 MHz Oscilloscope Calibration Option