Fluke 校正 9009 可擕式雙體乾式爐
Model#: Fluke 9009-B-156
主要功能
- Temperature ranges from –15 °C to 350 °C.
- Two wells in each block for simultaneous comparison calibrations.
- The cold block is calibrated to within ±0.2°C with stability of ±0.05°C. The hot block’s display is accurate to ±0.6°C with stability of ±0.05°C.
- Rugged, lightweight, and water-resistant enclosure.
產品概述: Fluke 校正 9009 可擕式雙體乾式爐
Two blocks. One portable calibrator. Full temperature coverage.
Dual Independent Temperature Blocks
The Fluke Calibration 9009 combines two independently controlled temperature blocks into one compact, rugged system:
- Hot block range: 50 °C to 350 °C
- Cold block range: –15 °C to 110 °C
- Each block is powered by a precision Hart Scientific temperature controller—not off-the-shelf components, but industry-leading technology from the world’s top temperature calibration experts.
Flexible Multi-Probe Calibration
Each block includes two wells with removable inserts, giving you unmatched flexibility:
- Calibrate up to four probes simultaneously
- Use two probes with an external reference thermometer (e.g., Hart 1521 LLK)
- Perform quick zero/span checks for transmitter calibration with both blocks in use
Portable and Field-Ready
The 9009 is built for portability and field durability:
- Housed in a rugged Pelican™ case—airtight, watertight, and impact-resistant
- Weighs only 10 lbs (4.5 kg), compact and easy to carry
- Internal storage holds the power cord and four extra inserts
- Inserts available for sensor sizes from 1/16” (1.6 mm) to 7/16” (11.1 mm)
This system is truly ready to go wherever your work takes you.
Reliable Accuracy and Stability
Performance you expect from a Fluke/Hart Scientific source:
Block | Display Accuracy | Stability |
Cold | ±0.2 °C | ±0.05 °C |
Hot | ±0.6 °C | ±0.05 °C |
NIST-traceable calibration certificate included for both blocks.
PC Connectivity and Automation
- RS-232 interface included for temperature monitoring and control via PC
- Fully compatible with MET/TEMP II software for automated calibration of RTDs, thermocouples, and thermistors
The Complete Field Calibration Package
With the Fluke 9009 Dual-Well Dry-Well, you get:
- Two independent blocks
- Full temperature coverage from –15 °C to 350 °C
- True field portability in a compact, durable case
- Flexibility for multiple sensors and calibration modes
- Built-in automation and PC integration
規格: Fluke 校正 9009 可擕式雙體乾式爐
General Specifications | ||
Hot block | Cold block | |
Range | 50°C to 350°C (122°F to 662°F) | –15°C to 110°C (5°F to 230°F) (–8°C [18°F] with hot block at 350°C [662°F]) |
Accuracy | ±0.6°C | ±0.2°C |
Stability | ±0.05°C | |
Well-to-well uniformity | ±0.1°C | |
Display resolution | 0.1° | |
Heating times | 30 minutes from 25°C to 350°C | 15 minutes from 25°C to 110°C |
Cooling times | 40 minutes from 350°C to 100°C | 16 minutes from 25°C to –15°C |
Stabilization times | 8 minutes | |
Well depth | 4 in (102 mm) | |
Removable inserts | Two 6.4 mm (1/4 in) and two 4.8 mm (3/16 in) inserts included; see ordering information for other available inserts | |
Computer interface | RS-232 included | |
Power | 115 V AC (±10 %), 3 A, or 230 V AC (±10 %), 2 A, specify, 50/60 Hz, 280 W | |
Size (H x W x D) | 178 x 267 x 248 mm (7 x 10.5 x 9.75 in) | |
Weight | 4.5 kg (10 lb) | |
NIST-traceable calibration | Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C | Data at –8°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 110°C |
機型: Fluke 校正 9009 可擕式雙體乾式爐
Fluke 9009-B-156
Fluke Calibration 9009 Dual-Well Dry-Well, Dual Block (Black), 110V 50/60HZ
Includes two 1/4 in (6.4 mm) and two 3/16 in (4.8 mm) inserts.
Fluke 9009-Y-156
Fluke Calibration 9009 Dual-Well Dry-Well, Dual Block (Yellow), 110V 50/60HZ
Includes two 1/4 in (6.4 mm) and two 3/16 in (4.8 mm) inserts.