Fluke 729 Pro Automatic Pressure Calibrator
主要功能
- Automatic pressure generation and regulation to 1000psi
- Replaceable pressure control modules
- Built-in test templates, one-key calibration
- HART communication
- Pressure leakage testing
- Automatic fine-pressure adjustment, automatic pressure stabilization
- Powerful, replaceable Li-ion battery
- Quick pressure release valve with a muffler
- Extend measurement range with 750P Series pressure modules
- Measure temperature with 720RTD probe
- Bright dual-three-channel /color graphic display
- Weight <5kg
- Rugged and portable design with standard 3-year warranty
- Absolute and gage pressure modules up to 7MPa.
- Use the modules to change the Fluke 729Pro to a different output pressure range
產品概述: Fluke 729 Pro Automatic Pressure Calibrator
Portable automatic pressure calibrator simplifies pressure calibration
The Fluke 729Pro Automatic Pressure Calibrator has been designed specifically with process technicians in mind to simplify the pressure calibration process and provide faster, more accurate test results. Technicians know that calibrating pressure can be a time-consuming task, but the 729Pro makes it easier than ever with an internal electric pump that provides automatic pressure generation and regulation in an in an easy-to-use, rugged, portable package.
Automatic pressure generation and control
The 729Pro allows you to simply type in a target pressure, and the calibrator will automatically pump to the desired set-point. With plug-in replaceable pressure control modules and pressure generation and regulation up to 1000 psi (70bar, 7000 kPa). Then, internal fine adjustment control automatically stabilizes the pressure at the requested value. You can also fill in a test template and the 729Pro will automatically pump to and document a multiple-point pressure calibration test at the touch of a button.
Easy calibration documentation
The Fluke 729Pro can automatically test multiple pressure test points and automatically document the results. The 729Pro does the work for you by documenting and flagging out-of-tolerance test results in red on the graphical display for:
- Applied pressure
- Measured mA, mV, or digital PV
- % error for each test point
Upload and manage documented calibration results with DPCTrack2™ Calibration Management Software, making it easy to manage your instrumentation, create scheduled tests and reports, and manage calibration data.
HART communication
Built-in HART communication capabilities enable HART transmitter mA adjustments, HART configuration, and the ability to adjust to applied 0% and 100% values. You can also configure various tasks such as changing a transmitter tag, measurement units, and ranging. Other supported HART commands include setting fixed mA outputs for troubleshooting, read device configuration, and variables and read device diagnostics.
Rugged, portable design
The 729Pro is tested and warranted to withstand a 1-meter drop test so it is ready for field instrumentation calibration work. With two models 4M up to 600 psi (40 bar, 4000 kPa), and 7M up to 1000 psi (70 bar, 7000 kPa) to choose from, the Fluke 729Pro Automatic Pressure Calibrators are designed to perform when and where you need them. A standard semi-rigid carrying case designed for field work, with room in the case to store test hoses, fittings, test leads, and small hand tools means it’s easy to carry everything needed for field pressure calibrations.
For greater accuracy and resolution on low pressure calibrations, the Fluke 729Pro allows you to select and insert one of 6 Fluke Pressure Modules to expand the pressure range between 0-600 psi (40 bar, 4000 kPa).
規格: Fluke 729 Pro Automatic Pressure Calibrator
關鍵參數 | 729 Pro 7M | 729 Pro 4M |
壓力控制量程 | -13 至 1000 psi,-90 至 7000 kPa, -0.90 至 70 bar | -13 至 600 psi,-90 至 4000 kPa, -0.90 至 40 bar |
壓力類型 | 表壓/絕對壓 (取決於內部或外部壓力模組的類型 | |
壓力準確度和控制穩定性 | 0.02% FS,0.005% FS | |
可更換的壓力測量模組 | 有,6 個模組 | 有,4 個模組 |
mA 電流源/測量 | 有 | |
電壓測量 | 有 | |
迴路電源 | 24 V dc | |
壓力開關測試 | 有 | |
壓力洩漏測試 | 有 | |
HART 通訊 | 有,支援基本指令 | |
自動階差與自動升降 | 有 (具備可自訂的速度、模式和樣式) | |
電池 | 充電式鋰電池,附充電指示功能 (可在儀器中充電或獨立充電) | |
語言 | 英文 |
電氣規格 | |
所有規格在 110% 量程內均有效,但 24 mA 電流源與模擬功能除外,此兩者僅在 100 % 量程內有效。 | |
量程 | |
mA 測量、電流源和模擬 | 0 mA 到 24 mA |
直流伏特測量 | 0 V dc 至 30 V dc |
解析度 | |
mA dc 電流源、模擬與測量 | 1 μA |
直流電壓測量 | 1 mV |
準確度 | 0.01% FS ±2 LSD 所有量程 (於 23 °C ± 5 °C) |
穩定度 | 20 ppm FS /°C (-10 °C 至 +18 °C,以及 28 °C 至 50 °C) |
mA 模擬工作電壓 | 12 V dc 至 30 V dc |
迴路相容電壓 | 24 V dc @ 20 mA |
溫度測量,100 Ω Pt(385) RTD | -50 °C 至 150 °C |
溫度解析度 | 0.01 °C |
溫度精準度 | 使用 720 RTD 探針 (選購配件) 時,具 ±0.1 °C ±0.25 °C 的合併不確定度 |
驅動能力 (mA 電流源) | 1200 Ω (無 HART 電阻器);950 Ω (具內部 HART 電阻器) |
機械規格 | |
尺寸 (高 x 寬 x 長) | 290 x 215 x100 mm,(4.3 x 4.4 x 9.4 in) |
重量 | 4.9 kg,(10.8 lbs) |
環境規格 | |
操作溫度 | -10 °C 至 50 °C |
壓力控制 | 0 °C 至 50 °C |
電池充電 | 0 °C 至 40 °C |
存放溫度 | -20 °C 至 60 °C |
運作海拔高度 | <3000 m |
存放海拔高度 | <13000 m |
運作濕度 | 非凝結 (<10 °C) ≤90 % RH (10 °C 至 30 °C) ≤75 % RH (30 °C 至 40 °C) ≤45 % RH (40 °C 至 50 °C) |
壓力模組選擇表格 | ||||||
壓力模組 | 壓力類型 | 量程 | 解析度 | 整體不確定性 | 729Pro 4M | 729Pro 7M |
FLK-PMM-200K | 表壓 | (-15 至 30) psi | 0.0001 psi | 0.02 % FS / 年,最大值 | • | • |
(-100 至 200) kPa | 0.001 kPa | |||||
(-1 至 2) bar | 0.00001 bar | |||||
FLK-PMM-200KA | 絕對壓 | (0 至 30) psi | 0.0001 psi | 0.05 % FS / 年,最大值 | • | • |
(0 至 200) kPa | 0.001 kPa | |||||
(0 至 2) bar | 0.00001 bar | |||||
FLK-PMM-2000K | 表壓 | (-14 至 300) psi | 0.001 psi | 0.02 % FS / 年,最大值 | • | • |
(-95 至 2000) kPa | 0.01 kPa | |||||
(-0.95 至 20) bar | 0.0001 bar | |||||
FLK-PMM-4000K | 表壓 | (-14 至 600) psi | 0.01 psi | 0.02 % FS / 年,最大值 | • | • |
(-95 至 4000) kPa | 0.01 kPa | |||||
(-0.95 至 40) bar | 0.0001 bar | |||||
FLK-PMM-7000K | 表壓 | (-14 至 1000) psi | 0.01 psi | 0.02 % FS / 年,最大值 | • | |
(-95 至 7000) kPa | 0.01 kPa | |||||
(-0.95 至 70) bar | 0.0001 bar | |||||
FLK-PMM-7000KA | 絕對壓 | (0 至 1000) psi | 0.01 psi | 0.02 % FS / 年,最大值 | • | |
(0 至 7000) kPa | 0.01 kPa | |||||
(0 至 70) bar | 0.0001 bar | |||||
僅限乾燥空氣或非腐蝕性氣體 / 15 °C 至 35 °C,所有量程 |