Fluke Calibration 9102S Handheld Dry-Well

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 54.17ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • Fluke Calibration 9102 Handheld Dry-Well - 1

Key Features

  • Temperature ranges from –10°C to 375°C.
  • Accuracy to ±0.25°C, with stability of ±0.05°C at 0°C.
  • Equipped with an RS-232 interface for connectivity.

Product overview: Fluke Calibration 9102S Handheld Dry-Well

9102S Dry-Well


For work in the temperature range of –10°C to 122°C, 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.

This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Model 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.

The 9102S has two wells so you can use one for a reference thermometer to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.

Specifications: Fluke Calibration 9102S Handheld Dry-Well

Specifications9100S9102S
Range35°C to 375°C (95°F to 707°F)–10°C to 122°C (14°F to 252°F) at 23°C ambient
Accuracy±0.25°C at 50°C; ±0.25°C at 100°C; ±0.5°C at 375°C±0.25°C
Stability±0.07°C at 50°C; ±0.1°C at 100°C; ±0.3°C at 375°C±0.05°C
Well-to-well uniformity±0.2°C with sensors of similar size at equal depths within wells
Heating times35°C to 375°C: 9.5 minutesAmbient to 100°C: 10 minutes
Stabilization5 minutes7 minutes
Cooling times375°C to 100°C: 14 minutesAmbient to 0°C: 10 minutes
Well depth102 mm (4 in); 1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep102 mm (4 in)
Removable insertsN/AAvailable in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power115 V AC (±10%), 55–65 Hz, 1.5 A or 230 V AC (±10%), 0.8 A, 45–55 Hz, 175 W94-234 V AC (±10%), 50/60 Hz, 60 W; or 12 V DC
Size (H x W x D)57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in)99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in)
Weight1 kg (2 lb 3 oz)1.8 kg (4 lb)
Computer interfaceRS-232 included
NIST-traceable calibrationData at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C and 375°CData at –10°C, 0°C, 25 °C, 50°C, 75 °C,100°C and 122°C

Models: Fluke Calibration 9102S Handheld Dry-Well

Fluke 9102S-156

9102S-156,DRY-WELL, HH LOW-TEMP


Buy it
9102S
HDRC Handheld Dry-Well

Buy it
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 22.81ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 22.9ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.