Fluke 884X-RTD 100 Ohm RTD Temperature Probe
Model#: 884X-RTD
Key features
- Manufactured using a coil suspension element design for increased shock and vibration resistance, Mineral-insulated sheath with a minimum bend radius of 19 mm (3/4 in) for flexibility and durability
Product overview: Fluke 884X-RTD 100 Ohm RTD Temperature Probe
Specifications: Fluke 884X-RTD 100 Ohm RTD Temperature Probe
| Specifications | |
| Resistance | Nominal 100 Ω |
| Temperature coefficient | 0.00385 Ω/Ω ºC nominal |
| Temperature range | -200 C to 300 ºC (transition and cable temperature: 0 ºC to 25% ºC) |
| Drift rate | + 0.13 ºC at 0 ºC after 1000 hours at 300 ºC |
| Sheath Material | 316 stainless steel |
| Leads | PTFE-insulated, nickel-plated stranded copper, 22 AWG |
| Termination | 4-wire banana |
| Time constant | Four seconds maximum for 63.2% response to step change in water moving at 3 fps |
| Bending radius | 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip |
| Calibration | Includes manufacturer’s NIST-traceable calibration and table with R vs. T values in 1 ºC increments from -196 ºC to 300 ºC. Callendar-van Dusen coefficients included |
| Immersion | At least 100 mm (4 in) recommended |
| Accuracy (includes calibration uncertainty and short-term stability) | ±0.50 ºC at -196 ºC ±0.050 ºC at 0 ºC ±0.052 ºC at 200 ºC ±0.055 ºC at 300 ºC |
| Size | 9 in L x 3/16 in diameter |
| Marking | Each RTD is marked with the Fluke name and part number |
Models: Fluke 884X-RTD 100 Ohm RTD Temperature Probe
884X-RTD
100 Ohm RTD Temperature Probe