Fluke 2600A-101 Sprt Probe

    Key features

    • The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinum Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard.

    Product overview: Fluke 2600A-101 Sprt Probe

    Secondary Platinum Resistance Thermometer Probe
    The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinium Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard. The 2600A-101 has stripped leads suitable for connection to any Fluke data acquisition product or other screw terminal connector. This probe comes with calibration certificates and data for -200 to 6600°C.

    Probe hook up diagram:
    The probe connects to channel 1(input) and 11(excitation) on the Hydra input connector.

    2600A-101 SPRT Probe

    2600A-101 SPRT Probe

    Warning: Probe handle surfaces are conductive. Use caution near electric circuits. Voltage levels above 30V rms, 42 V peak, or 60 V dc should not be applied to the probe.

    Specifications: Fluke 2600A-101 Sprt Probe

    General Specifications
    Use Temperature200 °C to 660 °C
    StabilityTypical draft < 0.01°C after 100 times from 25 to 660 °C
    Calibrated Accuracy
    TemperatureUncertainty
    - 183°C+/- 0.01°C
    0°C+/- 0.007°C
    200°C+/- 0.024°C
    75%°C+/- 0.033°C
    660 °C+/- 0.050 °C
    Performance Specifications (nominal)
    Ice point resistance:100 Ω ± 0.1 Ω
    Alpha value:0.00385 ± 0.00001 Ω /Ω /C
    Insulation resistance:1000 Meg Ω at 100 Vdc at 20°C
     100 Meg Ohms at 100 Vdc at 400°C
    Hysteresis:0.01°C Max. using 0, 400°C as the end points
    Self heating:0.02°C / mw in well stirred bath
    Thermoelectric voltage:When tested in an ice bath, with immersion depth from 4 to 10 inches (10 to 25 cm), thermo-electric voltage will not exceed 2 µV
    Time constant:9 seconds typical for 63.2% response to step change
    Long term stability:Stability at 0°C after exposure to upper temperature limits
    Dimensions
    Sheath diameter (D):.25 inches (6.3 mm)
    Length (L):12 inches (0.3 m)
    Cable length (Y):6 feet (1.83 m)
    Wire:22 AWG, PTFE jacketed, rated to 200 C
    Wire terminations:flying lead; stripped / tinned