MET/TEMP II (9938) and 1594A/1595A Waiting Command Notification

MET/TEMP II and 1594A/1595A Notification (261.41 KB)

Fluke Calibration is voluntarily notifying customers of a required set up procedure in order to prevent 9938 software from waiting for a command and to prevent the 159X from not sending a response, which is causing the heat source to continue heating until the cut out temperature is reached.

Please review the document above for complete details.