Fluke Calibration P3031 and P3032 Pneumatic Deadweight Testers
Key features
- Pressure ranges to 2000 psi (140 bar)
- 0.015% of reading accuracy (0.008% optional)
- Screw press for fine pressure control
- Piston flotation indicator
- Mounted spirit level with adjustable feet
- Series 3 non-magnetic stainless steel weights
Product overview: Fluke Calibration P3031 and P3032 Pneumatic Deadweight Testers
The Models P3031 and P3032 feature a gas driven oil lubricated piston/cylinder assembly which ensures that the instrument is less susceptible to performance issues due to contamination or the operating environment. These gas driven deadweight testers eliminate the maintenance issues associated with conventional high pressure pneumatic deadweight testers. The supply gas acts directly on the oil lubricated pistons so there is no effect on system accuracy.
Specifications: Fluke Calibration P3031 and P3032 Pneumatic Deadweight Testers
Accuracy | |
Accuracy* | ±0.015% of reading (±0.008% optional) |
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Mass and PCU Materials of Construction | |
Standard weight material | Series 3 non-magnetic austenitic stainless steel |
Weight density | 7.8 g/cm³ |
Optional fractional weights | Solution heat treated aluminum |
Weight density | 2.7 g/cm³ |
Models (P) P3031, P3032 (P & V) | Piston Material: Tungsten carbide Cylinder Material: Martensitic steel Coefficient of Expansion: 16.5 ppm/°C |
General | |
Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP |
Instrument weight | 24 lbs |
Instrument size | 17.5 W x 12 D x 8.5 H |
Mass set weight (typical) | 65 lbs |
Minimum Standard Weight Increments | |
P3016, P3031, P3032 | 1 psi, (0.1bar) |
Optional Fractional Weights | |
P3014, P3015, P3025 Pressure | 0.1 psi (0.01 bar) |
Built-in Hand Pumps | |
Pressure mode | Max pressure 300 psi (20 bar) |
Vacuum mode | To 90% vacuum |
Operating Fluid for Liquid Lubricated Models | |
55-655 | mineral oil (standard) |
Krytox | for oxygen-safe applications (optional) |
Options | |
PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. |
Krytox | As operating fluid for liquid lubricated gas operated models. |
Models: Fluke Calibration P3031 and P3032 Pneumatic Deadweight Testers
P3031-KGCM2
P3031-KGCM2 Liquid Lubricated Gas Operated Model
Pressure Range 1 to 70 kgf/cm
Pressure Range 1 to 70 kgf/cm
P3031-KPA
P3031-KPA Liquid Lubricated Gas Operated Model
Pressure Range 100 to 7,000 kPa
Pressure Range 100 to 7,000 kPa
P3031-MPA
P3031-MPA Liquid Lubricated Gas Operated Model
Pressure Range 0.1 to 7 Mpa
Pressure Range 0.1 to 7 Mpa
P3032-BAR
P3032-BAR Liquid Lubricated Gas Operated Model
Pressure Range 1 to 140 bar
Pressure Range 1 to 140 bar
P3032-KGCM2
P3032-KGCM2 Liquid Lubricated Gas Operated Model
Pressure Range 1 to 140 kgf/cm
Pressure Range 1 to 140 kgf/cm
P3032-PSI
P3032-PSI Liquid Lubricated Gas Operated Model
Pressure Range 10 to 2,000 psi
Pressure Range 10 to 2,000 psi
P3032-KPA
P3032-KPA Liquid Lubricated Gas Operated Model
Pressure Range 100 to 14,000 kPa
Pressure Range 100 to 14,000 kPa
P3032-MPA
P3032-MPA Liquid Lubricated Gas Operated Model
Pressure Range 0.1 to 14 Mpa
Pressure Range 0.1 to 14 Mpa
P3031-PSI
P3031-PSI Liquid Lubricated Gas Operated Model
Pressure Range 10 to 1,000 psi
Pressure Range 10 to 1,000 psi
P3031-BAR
P3031-BAR Liquid Lubricated Gas Operated Model
Pressure Range 1 to 70 bar
Pressure Range 1 to 70 bar