[교정] Fluke 753-754 Process Calibrator Firmware Upgrade Software Download

Fluke-75X Version 1.11 Firmware Upgrade

Applies only to Firmware Version 1.10

To upgrade your 75X to the latest firmware, follow these simple instructions:

  1. Download Fluke−75X Version 1.11 firmware upgrade and save the firmware upgrade file.
  2. Ensure your Fluke 75x is connected to your Windows PC computer via USB cable and using the battery charger for power. Run the file and follow the instructions for doing the update. Do not turn off the PC or the 75x while the update is in progress. Once the update is complete, cycle the 75x off and on and verify that the new version
    number is version 1.11 ). It is NOT necessary to recalibrate the 75x after the firmware update has been completed.
  3. Optional information is available for download on software licensing for the open−source software used.

Version 1.11 corrects an issue in version 1.10 when the 75x is measuring mA in a >12V circuit where >35mA is expected, the 75x can open the loop displaying a near zero reading. Version 1.11 does not apply for instruments with firmware version 1.0x.

If you experience any problems during the update, launch the update application and try again. If problems still occur, contact Fluke Support for further assistance.



Fluke-75X Version 1.01 Firmware Upgrade

Applies only to Firmware Version 1.00

To upgrade your 75X to the latest firmware, follow these simple instructions:

  1. Download Fluke−75X Version 1.01 firmware upgrade and save the firmware upgrade file.
  2. Ensure your Fluke 75x is connected to your Windows PC computer via USB cable and using the battery charger for power. Run the file and follow the instructions for doing the update. Do not turn off the PC or the 75x while the update is in progress. Once the update is complete, cycle the 75x off and on and verify that the new version number is version 1.01 ). It is NOT necessary to recalibrate the 75x after the firmware update has been completed.
  3. Optional information is available for download on software licensing for the open−source software used.

This firmware update includes Fluke 754 support for the Rosemount 3051C rev 5.7 HART pressure transmitter. Also in the V1.01 upgrade is additional language translations of the user interface for both the 753 and 754. Extended languages require a special access code and are enabled in the applicable countries. Version 1.01 does not apply for instruments with firmware version 1.1x.

If you experience any problems during the update, launch the update application and try again. If problems still occur, contact Fluke Support for further assistance.

Fluke 753/754 Windows Driver

Download 753/754 Windows Driver »

This is the driver that will enable Windows programs to communicate with the 753 and 754 through the USB port. Download this self-extracting installer and run it. Once installed, the 753/754 will be assigned to a Windows COM port. If necessary for the software package you are using, launch the windows device manager and you see what COM port has been assigned. If you have already installed DPCTrack2, then it installs this driver and it should not be necessary to do it again. For other software that communicates with the 753/754, you will need to install this driver. Some software will automatically find the appropriate COM port and some will require that you manually configure it by inputting the COM port that the 753/754 is communicating on (this is shown in the Windows Device Manager). See the particular software for more information.