How to do a temperature sensor comparison calibration
Hereโs how to do a simple comparison calibration of a probe-styleย RTDย at one temperature point usingย temperature sensor calibration equipment, aย dry-block calibratorย in this case.ย Fully insert the probe RTD into one of the dry-block insert holes and connect the cable to an RTD temperature readout.ย Ramp the dry block to 100 ยบC (212 ยบF).ย When the dry-block display shows a stable 100 ยบC, record the value of the RTD temperature readout.ย You now have one data point to compare the RTD temperature on the readout to the dry-block calibrator which is being used as a stable measurement reference standard.ย A typical sensor calibration would be done at several temperature points that span the working range of the sensor when used in a production environment
Comparison calibration using a dry-block calibrator
For a higher accuracy calibration, a referenceย PRTย connected to a thermometer readout could be inserted into another of the dry-block insert holes.ย In this case, the PRT which has a higher accuracy than the internal display of the dry-block calibrator is used as the reference standard and compared to the reading of the RTD temperature readout.ย 
Comparison calibration using a reference PRT with dry-block
Please readย Selecting a dry-block calibratorย for help in choosing the right dry-block calibrator. ย