Fluke Calibration 9102 Handheld Dry-Well

  • Fluke Calibration 9102 Handheld Dry-Well - 1

Principales fonctions

World’s smallest, lightest and most portable dry-wells

  • Smallest dry-wells in the world
  • Ranges from –10°C to 122°C
  • Accuracy to ±0.25°C, stability of ±0.05°C
  • RS-232 interface
  • Fluke Calibration’s line of portable dry-wells is incredible. They’re the smallest, lightest, and most portable dry-wells in the world. And now they're better than ever!

Présentation du produit: Fluke Calibration 9102 Handheld Dry-Well

9102S Dry-Well

For work in the temperature range of –10°C to 122°C, 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.

This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Model 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.

The 9102S has two wells so you can use one for a reference thermometer to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.

Spécifications: Fluke Calibration 9102 Handheld Dry-Well

Specifications

9100S9102S
Range
 
35 °C to 375 °C (95 °F to 707 °F)
 
–10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient
Accuracy
 
± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C± 0.25 °C
 
Stability
 
± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C± 0.05 °C
 
Well-to-Well Uniformity± 0.2 °C with sensors of similar size at equal depths within wells
Heating Times35 °C to 375 °C: 9.5 minutesambient to 100 °C: 10 minutes
Stabilization5 minutes7 minutes
Cooling Times375 °C to 100 °C: 14 minutesambient to 0 °C: 10 minutes
Well Depth
 
102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep102 mm (4 in)
 
Removable Inserts
 
 
 
N/A
 
 
 
Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power
 
115 V ac (± 10 %), 55–65 Hz, 1.5 A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz, 175 W94-234 V ac (± 10 %), 50/60 Hz, 60 W;or 12 VDC
 
Size (HxWxD)
 
57 x 125 x 150 mm(2.25 x 4.9 x 6.1 in)99 x 140 x 175 mm(3.9 x 5.5 x 6.9 in)
Weight1 kg (2 lb 3 oz)1.8 kg (4 lb)
Computer InterfaceRS-232 included
NIST-TraceableCalibration
 
 
Data at 50 °C, 100 °C, 150 °C, 200 °C,250 °C, 300 °C, and 375 °CData at –10 °C, 0 °C, 25 °C, 50 °C, 75 °C,100 °C, and 122 °C
 

Modèles: Fluke Calibration 9102 Handheld Dry-Well

9102S

HDRC Handheld Dry-Well

Acheter