Fluke Set of 3 5A/50A Current Clamps

  • Fluke Set Of 3 5A/50A Current Clamps | Fluke
Hãy mua ngay

Tổng quan sản phẩm: Fluke Set of 3 5A/50A Current Clamps

Expand the capabilities of the Fluke 1735 and Fluke 1740 Series Loggers with the dual-range 5 A/50 A current clamps. The accessory includes three precision clamps that are automatically detected and scaled by the instrument when attached.

Thông số kỹ thuật: Fluke Set of 3 5A/50A Current Clamps

Specifications
Electrical characteristics
Nominal current in: 50 A AC rms
Overload: up to 100 A rms
Conductor position influence: < 0.5% of range at 50/60 Hz
Error due to adjacent conductor: <= 15 mA/A at 50 Hz
Phase shift at 2 kHz at 200 A: < ± 1 degree
Frequency range: 40 Hz to 10 kHz (-3 dB)
Temperature coefficient: 0.015% of range / °C
Working voltage (see section safety): 600 V AC rms or DC
Order reference: 1-phase: EP0442A
 3-phase: EP0452A
 3-phase+N: EP0453A
Accuracy
Primary current 0.125 A to 2.5 A Operating error in % of reading: ± 2.5%
 Phase shift: not specified
Primary current 2.5 A to 12.5 A Operating error in % of reading: ± 1.5%
 Phase shift: < 2°
Primary current 12.5 A to 25 A Operating error in % of reading: ± 0.5%
 Phase shift: < 0.5°
Primary current 25 A to 50 A Operating error in % of reading: ± 0.5%
 Phase shift: < 1°
General characteristics
Maximum conductor size: 15 mm diameter, bus bar 15 x 17 mm
Cable length: 2 m
Operating temperature range: -10 °C to +55 °C
Storage temperature range: -20 °C to +70 °C
Operating humidity: 15% to 85% r.h. (non condensing)
Weight (per clamp): 175 g
Reference conditions
Environment temperature range: +18 °C to +26 °C
Humidity: 20% to 75% r.h.
Current: sinusoidal waveform, 48 to 65 Hz, distortion factor: < 1%, no DC component
Load impedance for voltage output: => 1 MΩ, <100 pF
Safety standards
 IEC/EN61010-1: 2001
 IEC/EN61010-2-032
 IEC/EN61010-2-031
Safety
 600 V CAT III, pollution degree 2
 The application of the clamps on uninsulated conductors is limited to 600V r.m.s. or DC to ground and frequencies below 1kHz.
EMC standards
 EN61326-1: 1997/A1: 1998
Dimensions
 102 x 42.5 x 23.5 mm

Model: Fluke Set of 3 5A/50A Current Clamps

Fluke MBX Clamp 5A/50A

Set of 3 5A/50A Current Clamps

Hãy mua ngay