Fluke 787 ProcessMeter Digital Multimeter
Compare products: digital multimeters
- Fluke 87V Industrial Multimeter
- Fluke 117 Electrician's Multimeter with Non-Contact Voltage
- Fluke 179 True-RMS Digital Multimeter
- Fluke 115 Field Technicians Digital Multimeter
- Fluke 177 True-RMS Digital Multimeter
- Fluke 289 True-RMS Data Logging Multimeter
- Fluke 28 II Rugged Digital Multimeter
- Fluke 116 Digital HVAC Multimeter
- Fluke 287 True-RMS Electronics Logging Multimeter
- Fluke 114 Electrical Multimeter
- Fluke 113 Digital Multimeter
- Fluke 233 Remote Display Digital Multimeter
- Fluke 83V Average Responding Industrial Multimeter
- Fluke 28 II Ex Intrinsically Safe True RMS Digital Multimeter
- Fluke 175 True-RMS Digital Multimeter
- Fluke 27 II Rugged Digital Multimeter
- Fluke 88V Deluxe Automotive Multimeter
- Fluke 77 IV Digital Multimeter
- Fluke 107 Palm-sized, CAT III Digital Multimeter
Key features
- DMM and a loop calibrator in one rugged, handheld tool
- Provides a precision 1000 V, 440 mA True-RMS digital multimeter designed to meet 1000 volt IEC 1010 CAT III standards
- Features a clear LCD with backlight; 4,000 counts (30,000 counts for DC current)
- Performs frequency measurements to 20 kHz
Product overview: Fluke 787 ProcessMeter Digital Multimeter
Fluke 787 ProcessMeter doubles your capabilities but not your cost
The Fluke 787 ProcessMeter combines a DMM and a Loop Calibrator in one rugged, handheld tool, for about what you would expect to pay for a loop calibrator alone. It's a maintenance and calibration tool that will be at home in every instrumentation technician's tool box. Based on the trusted Fluke 87 DMM, the 787 adds the ability to measure, source, and simulate dc loop current with 1 microamp resolution and 0.05% accuracy. It puts a complete solution for troubleshooting and calibrating current loop applications in the palm of your hand so you can get more information, faster and easier.
So if you're tired of having to track down a calibrator, lugging around two tools, and switching back and forth, get the convenient solution for doubling your power. The 787 ProcessMeter from Fluke.
Other useful features:
- Provides Min/Max/Average/Hold/Relative modes, and diode test and continuity beeper
- Displays simultaneous mA and % of scale readout
- Provides 20 mA DC current source/loop calibrator/simulator
- Manual step (100%, 25%, Coarse, Fine) plus auto step and auto ramp
- Externally accessible battery for easy battery changes
- 1000 V overload protection on V, ohms, frequency
- 150 V overload protection on mA, backed up by 440 mA 1000V fuse
Want a built in 24-Volt loop supply and 250 Ω HART resistor? Consider the Fluke 789.
What's in the box:
- Fluke 787 ProcessMeter Digital Multimeter
- (1 pair) TL75 Safety-Designed Test lead Set
- (1 pair) AC70A Alligator Clips
- C81Y Protective Yellow Holster with Flex-Stand™
- Installed 9V Alkaline Battery
- Quick reference guide
Specifications: Fluke 787 ProcessMeter Digital Multimeter
Specifications | ||
Current measurement | Range | 0-1 A |
Resolution | 1 mA | |
Accuracy | 0.2% + 2 LSD | |
Range | 0-30 mA | |
Resolution | 0.001 mA | |
Accuracy | 0.05% + 2 LSD | |
Current sourcing | Range | 0-20 mA or 4-20 mA |
Resolution | 0.05% of span | |
Drive Capability | 500 Ω @ 24 mA | |
Frequency | To 19.999 kHz, 0.005% + 1 LSD | |
Continuity | Beeps for resistance < 100 ohms | |
Voltage measurement | Range | 0-1000 V AC or DC |
Resolution | 0.1 mV to 1.0 V | |
Accuracy | 0.1% Rdg + 1 LSD (VDC) | |
Resistance measurement | To 40 MΩ, 0.2% + 1 LSD | |
Diode test | 2.4 V shows diode voltage drop | |
Environmental Specifications | ||
Operating temperature | -20°C to 55°C | |
Storage temperature | -40°C to 60°C | |
Humidity (without condensation) | 95% up to 30°C | |
75% up to 40°C | ||
45% up to 50°C | ||
35% up to 55°C | ||
Temperature coefficient | 0.05 x (specified accuracy) per °C (for temperatures < 18°C or > 28°C) | |
Safety Specifications | ||
Electrical safety | Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 | |
Overvoltage category | IEC 1010-1 CAT III | |
Maximum voltage | 1000 V rms | |
Mechanical and General Specifications | ||
Size | 32 x 87 x 187 mm | |
Weight | 369 g | |
Batteries | Single 9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61) | |
Warranty | 3 years | |
Battery life | 50 hours typical (measurement), 12 hours typical (sourcing 12 mA) | |
Shock and vibration | Random, 2 g, 5-500 Hz | |
1 meter drop test | ||
Display current and % of span | Yes | |
Auto step, auto ramp | Yes |
Models: Fluke 787 ProcessMeter Digital Multimeter
Includes:
- AC70A Alligator Clips
- C81Y Protective Yellow Holster with Flex-Stand™
- One 9V Alkaline Battery
- Quick Reference Guide
- TL75 Safety-Designed Test lead Set