The CNX 3000 General Maintenance System is part of the team of Fluke wireless troubleshooting tools.
Connect the CNX i3000 iFlex AC Wireless Current Module to your test point, and view the results up to 20 meters away on the CNX Wireless Multimeter. Or connect the CNX v3000 AC Wireless Voltage Module and view both voltage and current measurements simultaneously. You’ll save time, with less run-around collecting multiple measurements. Use multiple modules for measurements on three-phase systems. Or use as a stand-alone measurement tool or combine with other CNX modules as a system for multiple measurements. From short distances, you can even view readings from modules through closed electrical panels. Plus no more writing down data as the CNX remote modules capture up to 65,000 sets of time stamped min/max/avg readings, using the optional PC adapter. CNX wireless test tools also offer increased safety by letting you view readings in a separate location from the test point. Now you can take readings on moving machinery, with only the measurement module in harm’s way.
View other members of the wireless team »
|
|
The CNX 3000 General Maintenance System provides measurement of all common electrical parameters.
The CNX 3000 Wireless Multimeter offers:
- True-RMS AC and DC voltage measurements to 1000V
- True-RMS AC and DC current with 0.01 mA resolution
- Continuity, resistance, diode test, capacitance and frequency measurements
- MIN/MAX Recording
- CAT III 1000 V; CAT IV 600 V; IP54
The CNX v3000 AC Wireless Voltage Module offers:
- True-RMS AC voltage measurements to 1000V
- ±1% accuracy to 500 Hz, ± 2% accuracy to 1 kHz
- Memory recording up to 65,000 readings
- Backlit LCD display
- CAT III 1000 V, CAT IV 600 V, CAT III 1000 V; IP42
The CNX i3000 iFlex AC Wireless Current Module offers:
- True-RMS AC current measurements to 2500 A
- 3% accuracy
- Memory recording up to 65,000 readings
- iFlex current probe allows you to get into tight, awkward spaces
- Backlit LCD display
- CAT III 1000, CAT IV 600 V
- IP42
|
 |
Using the optional PC Adapter, the CNX wireless system can track up to 10 measurement modules simultaneously, with results sent to a PC for further analysis.
|
CNX 3000 Wireless Multimeter
For all specifications: accuracy is specified for one year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
| Detailed Specifications |
| AC voltage |
| Range ¹ |
Resolution |
Accuracy ² ³ |
| 45 Hz to 500 Hz |
500 Hz to 1 kHz |
| 600.0 mV |
0.1 mV |
1.0 % + 3 |
2.0 % + 3 |
| 6.000 V |
0.001 V |
| 60.00 V |
0.01 V |
| 600.0 V |
0.1 V |
| 1000 V |
1 V |
¹ All AC voltage ranges are specified from 1 % of range to 100 % of range.
² Crest factor of ≤ 3 at full scale p to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
³ For non-sinusoidal waveforms, add –(2 % of reading + 2 % full scale) typical, for crest factor up to 3. |
| DC Voltage, Continuity, Resistance, Diode Test and Capacitance |
| Function |
Range |
Resolution |
Accuracy |
| mV |
600.0 mV |
0.1 mV |
0.09 % + 2 |
| V |
6.000 V |
0.001 V |
0.09 % + 2 |
| 60.00 V |
0.01 V |
| 600.0 V |
0.1 V |
| 1000 V |
1 V |
0.15 % + 2 |
| Ω |
600 Ω |
1 Ω |
Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 μs or longer. |
| Ω |
600.0 Ω |
0.1 Ω |
0.5 % + 2 |
| 6.000 kΩ |
0.001 kΩ |
0.5 % + 1 |
| 60.00 kΩ |
0.01 kΩ |
| 60.00 kΩ |
0.1 kΩ |
| 600.0 kΩ |
0.001 MΩ |
| 50.00 MΩ |
0.01 MΩ |
1.5 % + 3 |
| Diode test |
2.000 V |
0.001 V |
1 % + 2 |
| µF |
1000 nF |
1 nF |
1.2 % + 2 |
| 10.00 μF |
0.01 μF |
| 100.0 μF |
0.1 μF |
| 9999 μF¹ |
1 μF |
10 % typical |
| ¹ In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2. |
| AC and DC Current |
| Function |
Range ¹ |
Resolution |
Accuracy |
mA AC
(45 Hz to 1 kHz) |
60.00 mA |
0.01 mA |
1.5 % + 3 |
| 400.0 mA ³ |
0.1 mA |
| mA DC ² |
60.00 mA |
0.01 mA |
0.5 % + 3 |
| 400.0 mA |
0.1 mA |
¹ All AC current ranges are specified from 5 % of range to 100 % of range.
² Input burden voltage (typical): 400 mA input 2 mV/mA.
³ 400.0 mA accuracy specified up to 600 mA overload. |
| Frequency |
| Range |
Resolution |
Accuracy ¹ |
| 99.99 Hz |
0.01 Hz |
0.1 % + 1 |
| 999.9 Hz |
0.1 Hz |
| 9.999 kHz |
0.001 kHz |
| 99.99 kHz |
0.01 kHz |
| ¹ Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps. |
| Input Characteristics |
| Function |
Overload
Protection |
Input Impedance
(nominal) |
Common Mode Rejection Ratio (1 kΩ unbalance) |
Normal Mode Rejection |
 |
1100 V RMS |
> 10 MΩ < 100 pF |
> 120 dB at DC, 50 Hz or 60 Hz |
> 60 dB at 50 Hz or 60 Hz |
 |
1100 V RMS |
> 10 MΩ < 100 pF |
> 60 dB, DC to 60 Hz |
|
 |
1100 V RMS |
> 10 MΩ < 100 pF |
> 120 dB at DC, 50 Hz or 60 Hz |
> 60 dB at 50 Hz or 60 Hz |
| Open circuit test voltage |
Full scale |
Typical short circuit current |
| To 6 MΩ |
50 MΩ |
 |
1100 V RMS |
< 2.7 V DC |
< 0.7 V DC |
< 0.9 V DC |
< 350 mA |
 |
1100 V RMS |
< 2.7 V DC |
2.000 V DC |
< 1.1 mA |
| Function |
Overload Protection |
Overload |
| mA |
Fused, 44/100 A, 1000 V FAST Fuse |
600 mA overload for 2 minutes maximum, 10 minutes rest minimum |
| MIN/MAX Recording |
| Function |
Accuracy |
| DC functions |
The specified accuracy of the measurement function ± 12 counts for changes > 350 mS in duration. |
| AC functions |
The specified accuracy of the measurement function ± 40 counts for changes > 900 mS in duration. |
| General Specifications |
| Maximum voltage between any terminal and earth ground |
1000 V DC or AC RMS |
| Ω fuse protection from A inputs |
0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only |
| Display (LCD) |
Update rate: 4/sec
Volts, amps, ohms: 6000 counts Frequency: 10,000 counts Capacitance: 1,000 counts |
| Battery type |
Three AA Alkaline batteries, NEDA 15A IEC LR6 |
| Battery life |
300 hours minimum |
| Temperature |
Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C |
| Relative humidity |
0 % to 90 % (0 °C to 35 °C)
0 % to 75 % (35 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C) |
| Altitude |
Operating: 2,000 m
Storage: 12,000 m |
| Temperature coefficient |
0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
| Wireless frequency |
2.4 GHz ISM Band 20 meter range |
| Size (H x W x L) |
4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in) |
| Weight |
340 g (12 oz) |
| Safety standards |
US ANSI: ANSI/ISA 61010-1 / (82.02.01): 3rd edition
CSA: CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL:UL 61010-1: 3rd edition
CE European: IEC/EN 61010-1:2010 |
Electromagnetic compatibility
EMI, RFI, EMC, RF |
EN 61326-1:2006, EN 61326-2-2:2006
ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
FCCID : T68-FWCS IC:6627A-FWCS |
| Ingress Protection (IP) rating |
IP54 |
Fluke CNX v3000 AC Wireless Voltage Module
Accuracy is specified as ± ([% of reading] + [number of least significant digits]). All ranges are autoranging. Accuracy is specified from 5 % to 100 % of the range obtained by autoranging, from 18 °C to 28 °C.
| AC Voltage |
|
|
Range ¹
|
Resolution
|
Accuracy ² ³
|
|
45 Hz to 500 Hz
|
500 Hz to 1 kHz
|
|
6.000 V
|
0.001 V
|
1.0 % + 3
|
2.0 % + 3
|
|
60.00 V
|
0.01 V
|
|
600.0 V
|
0.1 V
|
|
1000 V
|
1 V
|
¹ All AC voltage ranges are specified from 1 % of range to 100 % of range.
² Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor ≤ 1.5 at 1000 V.
³ For non-sinusoidal waveforms, add –(2 % of reading + 2 % full scale) typical, for crest factor up to 3. |
| General Specifications |
|
LCD w/backlight
|
3½ digits, 6000 counts, updates 4/sec
|
|
Battery type
|
2 AA, NEDA 15 A, IEC LR6
|
|
Battery life
|
400 Hours
|
|
Memory
|
Record up to 65,000 readings
|
|
Unbound listen interval
|
5 seconds
|
|
RF communications
|
2.4 GHZ ISM Band
|
|
RF communication range
|
20 Meters
|
|
Operating temperature
|
-10 °C to +50 °C
|
|
Storage temperature
|
-40 °C to +60 °C
|
|
Operating humidity
|
90 % at 35 °C, 45 % at 40 °C, 45 % at 50 °C
|
|
EMC
|
EN 61326-1:2006
|
|
Safety compliance
|
EN/IEC 61010-1:2010 to 1000 V Measurement Category (CAT) III
600 V Measurement Category (CAT) IV
EN/IEC 61010-2-030:2010
EN/IEC 61010-031:2002+A1:2008
|
|
Pollution degree
|
2
|
|
Temperature coefficient
|
0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C)
|
|
Safety rating
|
CAT IV 600 V, CAT III 1000 V
|
|
Certifications
|
CSA, FCC T68-FWCS IC:6627A-FWCS
|
|
Ingress Protection (IP) rating
|
IP42
|
|
Size (HxWxD)
|
16.5 cm x 6.35 cm x 3.56 cm
(6.5 in x 2.5 in x 1.4 in)
|
|
Weight
|
.22 kg (8 oz)
|
Fluke CNX i3000 iFlex AC Wireless Current Module
| Specifications |
|
Range
|
2500 A AC
|
|
Resolution
|
0.1 A
|
|
Accuracy
|
3 % ± 5 digits
|
|
Crest factor (50 Hz/60 Hz)
|
3.0 at 1100 A, 2.5 at 1400 A, 1.42 at 2500 A, add 2 % for C.F. > 2
|
|
Display
|
3-1/2 digits, LCD w/backlight
|
|
Log rate/interval
|
1 sec minimum/adjustable by PC or front panel
|
|
Battery type
|
2 AA, NEDA 15 A, IEC LR6
|
|
Battery life
|
400 hours
|
|
Memory
|
Record up to 65,000 readings
|
|
RF communications
|
2.4 GHZ ISM Band
|
|
RF communication range
|
20 Meters
|
|
Operating temperature
|
-10 °C to +50 °C
|
|
Storage temperature
|
-40 °C to +60 °C
|
|
Operating humidity
|
90 % at 35 °C, 75 % at 40 °C, 45 % at 50 °C
|
|
EMC
|
EN 61326-1:2006
|
|
Safety compliance
|
EN/IEC 61010-1:2010 to 1000 V Measurement Category (CAT) III
600 V Measurement Category (CAT) IV
EN/IEC 61010-2-030:2010
EN/IEC 610101-2-031:2002
EN/IEC 61010-2-032:2002
|
|
Pollution degree
|
2
|
|
Temperature coefficient
|
Add 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
|
|
Safety rating
|
CAT IV 600 V, CAT III 1000 V
|
|
Certifications
|
CSA, FCC T68-FWCS IC:6627A-FWCS
|
|
Ingress Protection (IP) rating
|
IP42
|
|
Size (HxWxD)
|
16.5 cm x 6.35 cm x 3.56 cm
(6.5 in x 2.5 in x 1.4 in) |
|
Weight
|
.22 kg (8 oz)
|
|
Jaw opening
|
25.4 cm (10 in) coil
|
|
| Model Name | Description | Price List | Buy it | | CNX 3000 GM | Fluke CNX 3000 General Maintenance System
Fluke CNX 3000 Wireless Multimeter
Fluke CNX v3000 AC Wireless Voltage Module
Fluke CNX i3000 iFlex AC Wireless Current Module
iFlex i2500-10 Flexible Current Probe
Fluke CNX pc3000 Wireless PC Adapter and Software
TL224 Test Leads
TL175 Test Leads
AC175 Alligator Clips
AC285 Alligator Clips
Two (2) Magnetic Hanging Straps
Soft Carrying Case
|  | | | CNX 3000 IND | Fluke CNX 3000 Industrial System
Fluke CNX 3000 Wireless Multimeter
Fluke CNX v3000 AC Wireless Voltage Module
Three (3) Fluke CNX i3000 iFlex AC Wireless Current Modules
Three (3) iFlex i2500-10 Flexible Current Probes
Fluke CNX pc3000 Wireless PC Adapter and Software
TL224 Test Leads
TL175 Test Leads
AC175 Alligator Clips
AC285 Alligator Clips
Four (4) Magnetic Hanging Straps
|  | | | CNX 3000 HVAC | Fluke CNX 3000 HVAC System
Fluke CNX 3000 Wireless Multimeter
Fluke CNX a3000 AC Wireless Current Clamp Module
Fluke CNX t3000 K-Type Wireless Temperature Module
Fluke CNX pc3000 Wireless PC Adapter and Software
TL175 Test Leads
AC175 Alligator Clips
80PK-1 K-Type Bead Thermocouple
Magnetic Hanging Strap
Soft Carrying Case
|  | | | CNX 3000 | Fluke CNX 3000 Wireless Multimeter
True-RMS Wireless Multimeter
TL175 Test Leads
AC175 Alligator Clips
|  | | | CNX v3000 | Fluke CNX v3000 AC Wireless Voltage Module
True-RMS AC Voltage Module
TL224 Test Leads
AC285 Alligator Clips
Magnetic Hanging Strap
|  | | | CNX i3000 | Fluke CNX i3000 iFlex AC Wireless Current Module
True-RMS Flexible Current Meter
iFlex i2500-10 Flexible Current Probe
TL175 Test Leads
AC175 Alligator Clips
Magnetic Hanging Strap
|  | | | CNX a3000 | Fluke CNX a3000 AC Wireless Current Clamp Module
True-RMS AC Clamp Meter
|  | | | CNX t3000 | Fluke CNX t3000 K-Type Wireless Temperature Module
K-Type Thermocouple Thermometer
80PK-1 K-Type Bead Thermocouple
Magnetic Hanging Strap
|  | | | CNX i3000 Kit | Fluke CNX i3000 iFlex AC Current Measurement Kit
Fluke CNX 3000 Wireless Multimeter
Fluke CNX i3000 iFlex AC Wireless Current Module
iFlex i2500-10 Flexible Current Probe
TL175 Test Leads
AC175 Alligator Clips
Magnetic Hanging Strap
Soft Carrying Case
|  | | | CNX v3000 Kit |
Fluke CNX v3000 AC Voltage Kit
Fluke CNX 3000 Wireless Multimeter
Fluke CNX v3000 AC Wireless Voltage Module
TL224 Test Leads
TL175 Test Leads
AC175 Alligator Clips
AC285 Alligator Clips
Magnetic Hanging Strap
Soft Carrying Case
|  | | | CNX a3000 Kit | Fluke CNX a3000 AC Current Clamp Kit
Fluke CNX 3000 Wireless Multimeter
Fluke CNX a3000 AC Wireless Current Clamp Module
TL175 Test Leads
AC175 Alligator Clips
Soft Carrying Case
|  | | | CNX t3000 Kit | Fluke CNX t3000 Temperature Measurement Kit
Fluke CNX 3000 Wireless Multimeter
Fluke CNX t3000 K-Type Wireless Temperature Module
TL175 Test Leads
AC175 Alligator Clips
80PK-1 K-Type Bead Thermocouple
Magnetic Hanging Strap
Soft Carrying Case
|  | |
| Clips | Buy it | | AC285 | AC285 SureGrip™ Alligator Clips |  |  |
| Current Probes | Buy it | | i2500-10 | i2500-10 iFlex™ Flexible Current Probes | | |
| Cases and Holsters | Buy it | | CNX C3000 | CNX C3000 Premium Modular Tool Bag
Webbed bag for customizable carrying system, when used with other CNX modular cases. | | | | CNX C3001 | CNX Small Modular Case C3001
Compatible with wireless modules Fluke CNX | | | | CNX C3002 | CNX C3002 Modular DMM 2-Compartment Soft Case
Compatible with Fluke CNX Wireless Multimeter
Inner pocket and carrying handle | | | | CNX C3003 | CNX C3003 Modular 3-Compartment Soft Case
Compatible with Fluke CNX wireless modules and meters
Three adjustable compartments, inner pocket, and carrying handle | | |
| PC Adapter and Software | Buy it | | CNX pc3000 | Fluke CNX pc3000 PC Adapter Fluke CNX sw3000 software | | |
| Test Leads | Buy it | | TL175 | TL175 TwistGuard™ Test Leads
|  |  | | TL224 | TL224 SureGrip™ Insulated Test Leads |  |  |
|
|
|
|
|
|
|
|
|