Calibrated temperature systems for standards lab applications
The HP 34420/SPRT and HP 34420/PRT are precision thermometers based on Hewlett-Packard's HP 34420A NanoVolt/Micro-Ohm Meter. Each model comes with a precision temperature probe and provides uncertainties as low as ±0.003ºC for precise temperature calibrations.
The HP 34420 NanoVolt/Micro-Ohm meter is a high-sensitivity multimeter optimized for performing low-level measurements. It combines low-noise voltage measurements with resistance and temperature functions, setting a new standard in low-level flexibility and performance. It can be used as a replacement for analog null detectors such as the Fluke 845 Series.
The HP 34420/SPRT, HP 34420/PRT, and HP 34420 NanoVolt/Micro-Ohm meters are conveniently available from your local Fluke representative.
|
|
Take the uncertainty out of your low-level measurements
Low-noise input amplifiers and a highly tuned input protection scheme bring reading noise down to 8 nV pp. Combine this with 7 ½ digits of resolution, selectable analog and digital filtering, 2 ppm basic 24-hour dcV accuracy, and a shielded, copper pin connector and you've got accurate, repeatable measurements you can count on.
Two input channels
An integral two-channel programmable scanner simplifies voltage comparisons. Built-in ratio and difference functions enable automated two channel measurements without the need for an external nanoVolt scanner. Both channels share the same low noise specifications to ensure accurate comparisons.
Built-in resistance and temperature
The HP 34420A combines its low-noise nanoVolt input circuits with a high-stability current source to provide precise low-level resistance measurements - no more hassling with the cost and complexity of an external current source. Three resistance modes are included:
- Standard
- Low-power
- Voltage-limited for dry-circuit testing
|
Offset compensation is also provided to minimize thermal EMFs and associated errors.
SPRT measurements
Built-in ITS-90 conversion routines accept the calibration coefficients from your SPRT probe for direct temperature measurement and conversion. Thermocouples, Thermistors, and RTDs are also supported.
The HP 34420/SPRT comes with a Rosemount 162CE probe, the industry standard for accuracy. The HP 34420/PRT comes with a Hart Scientific 5628 probe, a more rugged solution for applications outside the lab. In both models, meter and probe are calibrated together as a system.
Unequaled versatility
The HP 34420A gives you the versatility to tackle your most challenging tasks, both on the benchtop and in your automated system. Standard features include: RS-232 and HP-IB interfaces, SCPI and Keithley 181 programming language, 1024-reading memory, scaling and statistics, and a chart recorded analog output.
Quality you can count on
The HP 34420A gives you the quality and reliability you expect from Hewlett-Packard. From the product's proven >150,000 hour Mean Time Between Failure, to its standard 3-year warranty, the HP 34420A brings you a new level of confidence in your low-level measurements.
|
| Technical Specifications |
| DC Voltage
(7 ½ digits resolution on all ranges) |
| 6 Measurement Ranges
|
mV to 100V |
| Best 1 Year Uncertainty |
±34 ppm of reading |
| Sensitivity |
0.1 nV |
| Input Resistance |
> 10 GΩ (up through 10V range) |
| Maximum p-p noise (2 minutes) |
8 nV |
| Input Channels
|
2, including DCV1/DCV2 ratio |
| Normal Mode Rejection |
100 dB |
| Input Protection
|
150V peak to Channel 1 LO |
| | Resistance
(7 ½ digits resolution on all ranges) |
| 7 Measurement Ranges |
1Ω to 1 MΩ |
| Best 1 Year Uncertainty |
±62 ppm of reading |
| Sensitivity |
0.1 mΩ |
| Test Current |
10 mZ to 5 mA |
| Measurement Modes |
2-wire and 4-wire |
| | Chart Out (Analog Out) | |
| Technical Specifications (Temperature Systems*) |
| Temperature Range |
| HP 34420A/SPRT
|
-200ºC to +661ºC |
| HP 34420/PRT |
-200ºC to +660ºC |
| | Probe Type |
| HP 34420A/SPRT |
Rosemount 162CE |
| Dia.: |
5.56 mm (0219 in.) |
| Length: |
457 mm (18 in.)
Pt 25.5 Ω |
| HP 34420A/PRT |
Hart Scientific 5628
|
| Dia.: |
6.34 mm (0.25 in.) |
| Length: |
304 mm (12 in.)
Pt 25.5 Ω |
| | System Uncertainty ** |
| HP 34420A/SPRT |
4 hour: 0003ºC* |
|
24 hour: 0.006ºC |
|
90 day: 0.013ºC |
|
1 year: 0.020ºC |
| HP 34420A/PRT
|
4 hour: 0.005ºC* |
|
24 hour: 0.007ºC |
|
90 day: 0.024ºC |
|
1 year: 0.053ºC |
| | Stability/Drift Rate |
| HP 34420A/SPRT: |
< 0.01 ºC/year |
| HP 34420A/PRT: |
< 0.05 ºC/year |
| | Lead Length |
| HP 34420A/SPRT: |
2.4M (8 ft.) |
| HP 34420A/PRT: |
2.1M (7 ft.) |
| | Sheath Material | | | System Resolution | |
*Probe and HP 34420A calibrated as a system using programmed constants with Rtpw check within last 4 hours
**Probe at 0ºC, HP 34420A @ 23ºC Typical specification -- absolute drift is highly dependent upon probe handling and usage
| General Specifications |
| Front Panel Connection |
|
Shielded, low thermal, 99% copper contacts |
| | Power Supply |
|
100V/120V/220V(230V)/240V ± 10% |
| | Power Line Frequency |
|
45 Hz to 66 Hz and 360 Hz to 440 Hz.
Automatically sensed at power-on. |
| | Power Consumption | | | Operating Environment |
|
Full accuracy for 0ºC to 55ºC. Full accuracy to 80% R.H. up to 30ºC. |
| | Storage Environment | | | Size |
|
254.4 mm W x 374.0 mm L x 103.6 mm H
(10.02 in W x 14.72 in L x 4.08 in H) |
| | Weight | | | Safety |
|
Designed to CSA, UL-1244, IEC-1010 |
| | RFI and ESD | | | Vibration and Shock |
|
MIL-T-28800E, Type III, Class 5 |
| | Warranty | |
|
| Model Name | Description | Price List | Buy it | | HP 34420/SPRT | NanoVolt/micro-Ohm meter with Rosemount 162CE SPRT calibration certificate
|  | | | HP 34420/PRT | NanoVolt/micro-Ohm meter with Hart Scientific 5628 PRT calibration certificate
|  | | | Rosemount 162CE | Probe, SPRT, 25.5 Ohm with Lemo connector
|  | |
|
|
|
|