Do more, carry less
The 743B does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. For documentation, the 743B automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations. The 743B has a PC interface that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. With its expanded memory, the 743B can hold a full week of calibrations and procedures.
|
Fluke 743B Documenting Process Calibrator with Data Logging
5
5
1
1
great
This is a must have for the proffesional as well as the student. I am a student at Lee College and I love how easy this Fluke 743b is to use when we do lab activitys that consist of troubleshooting conections, calibrating instruments or pushing a signal from the field in our pilot plant outside all the way to the other end of the campus where the control room is located. My hat is off to the Fluke company for thinking of the costumer first
February 13, 2013
|
The 743B is easy to use with flexible test strategies. And like all Fluke tools, it's rugged, reliable and tough as nails. It's like having several powerful tools in one - so you can work faster than ever:
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
- Power transmitters during test using loop supply with simultaneous mA measurement
- Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
- Measure and source with one compact, rugged, reliable tool
- Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
- Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
- Easy to use
- Three year warranty
- Bright white display. Read both sourced and measured parameters.
- Multi lingual interface
- Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1msec.
- Standard delivered with DPC/Track Sample software. Compatible with many Asset Management software packages.
|
|
| Measurement Accuracy |
| Voltage DC |
| 110.000 mV |
0.025%+0.015% * |
| 1.10000 V |
0.025%+0.005% * |
| 11.0000 V |
0.025%+0.005% * |
| 110.000 V |
0.05%+0.005% * |
| 300.00 V |
0.05%+0.005% * |
| | Voltage AC |
| 20 to 40 Hz |
2% + 10 |
| 40 to 500 Hz |
0.5% + 5 |
| 500 to 1 kHz |
2% + 10 |
| 1 kHz to 5 kHz |
10% + 20 |
| Ranges: |
1.1000, 11.000, 110.00, 300V |
| | Current DC |
| 30.000 mA |
0.01% + 0.015% * |
| 110.00 mA |
0.01% + 0.015% * |
| | Resistance |
| 11.000 Ω |
0.05% + 50 mΩ |
| 110.00 Ω |
0.05% + 50 mΩ |
| 1.1000 kΩ |
0.05% + 500 mΩ |
| 11.000 kΩ |
0.1% + 10 Ω |
| | Frequency |
| 1.00 to 109.99 Hz |
0.05 Hz |
| 110.0 to 1099.9 Hz |
0.5 Hz |
| 1.100 to 10.999 kHz |
5 Hz |
| 11.00 to 50.00 kHz |
50 Hz |
| | Pressure |
| Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
| | Note |
| * |
(% of reading + % of full scale) |
|
| Source Accuracy |
| Voltage DC |
| 110.000 mV |
0.01%+0.005% * |
| 1.10000 V |
0.01%+0.005% * |
| 15.0000 V |
0.01%+0.005% * |
| | Current DC |
| 22.000 mA (Source) |
0.01%+ 0.015% * |
| 22.000 mA (Simulate) |
0.02% + 0.03% * |
| | Resistance |
| 11.000 Ω |
0.01% + 20 mΩ |
| 110.00 Ω |
0.01% + 40 mΩ |
| 1.1000 kΩ |
0.02% + 500 mΩ |
| 11.000 kΩ |
0.03% + 5Ω |
| | Frequency |
| 0.00 to 10.99 Hz |
0.01 Hz |
| 11.00 to 109.99 Hz |
0.1 Hz |
| 110.0 to 1099.9 Hz |
0.1 Hz |
| 1100 to 21999 Hz |
2 Hz |
| 22.000 to 50.000 kHz |
5 Hz |
| | Note |
| * |
(% of reading + % of full scale) |
|
| RTDs and Thermocouples |
| Measure Accuracy: |
| 10 Ω Cu (427):
|
2 °C |
| 100 Ω Pt (3916): |
0.3 °C |
| 100 Ω Pt (3926): |
0.3 °C |
| 100 Ω Pt (385): |
0.3 °C |
| 200 Ω Pt (385): |
0.3 °C |
| 500 Ω Pt (385): |
0.3 °C |
| 1000 Ω Pt (385): |
0.3 °C |
| 120 Ω Ni (672): |
0.3 °C |
| note: |
For 2-and 3-wire measurement, add 0.4 °C |
| E: |
0.3 °C |
| N: |
0.5 °C |
| J: |
0.3 °C |
| L: |
0.3 °C |
| K: |
0.3 °C |
| T: |
0.3 °C |
| U: |
0.3 °C |
| B: |
0.9 °C |
| R: |
1.0 °C |
| S: |
0.9 °C |
| C: |
0.6 °C |
| note: |
Accuracy with external cold junction, for internal junction add 0.2 °C |
| | Source Accuracy: |
| 10 Ω Cu (427): |
1 °C |
| 100 Ω Pt (3916): |
0.1°C |
| 100 Ω Pt (3926): |
0.1°C |
| 100 Ω Pt (385): |
0.1°C |
| 200 Ω Pt (385): |
0.1°C |
| 500 Ω Pt (385): |
0.1°C |
| 1000 Ω Pt (385): |
0.1°C |
| 120 Ω Ni (672): |
0.1°C |
| note: |
For 2-and 3-wire simulation, add 0.4 °C |
| E: |
0.2 °C |
| N: |
0.3 °C |
| J: |
0.2 °C |
| L: |
0.2 °C |
| K: |
0.3 °C |
| T: |
0.3 °C |
| U: |
0.3 °C |
| B: |
0.8 °C |
| R: |
0.9 °C |
| S: |
0.9 °C |
| C: |
0.6 °C |
| note: |
Accuracy with external cold junction, for internal junction add 0.2 °C |
|
| Specifications |
| Data Log Functions |
| Measure functions: |
Voltage, current, resistance, frequency, temperature, pressure |
| Reading rate: |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
| Maximum record length: |
8000 readings (7980 for 30 or 60 readings/minute) |
| | Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Rate: |
4 steps/second |
| Trip detect: |
Continuity or voltage (continuity detection not available when sourcing current) |
| | Loop Power Function |
| Voltage: |
Selectable, 24 V or 28 V |
| Accuracy: |
5% |
| Maximum current: |
22 mA, short circuit protected |
| Maximum input voltage: |
30 V DC |
| | Step Functions |
| Source Functions |
Voltage, current, resistance, frequency, temperature |
| Manual Step |
Selectable step, change with arrow buttons |
| Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
|
| Environmental Specifications |
| Operating Temperature |
|
-10°C to +50°C |
|
-20°C to +50°C * |
| | Storage Temperature | | | Dust/water resistance | | | Operating Altitude | | | Note |
| * |
Except frequency and AC |
|
| Safety Specifications |
| Agency Approvals |
|
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
|
| Mechanical & General Specifications |
| Size | | | Weight | | | Batteries | | | Battery Life | | | Battery Replacement |
|
Via snap-shut door without opening calibrator; no tools required |
| | Side port connections |
|
Pressure module connector |
|
RS-232 connector for PC interface cable |
|
Connection for optional battery eliminator
|
| | Data storage capacity |
|
1 week of calibration results |
| | 90 day specifications |
|
The standard specification intervals for the 743B is 1 and 2 years. |
|
Typical 90 day measurement and source accuracy can be estimated by |
|
dividing the one year "% of reading" or "%of output" specifications by 2. |
|
Floor specifications, expressed as "% of full scale" or |
|
"counts" or "ohms" remain constant. |
|
|
| Model Name | Description | Price List | Buy it | | Fluke 743B |
Documenting Process Calibrator including:
- BC7217 battery charger
- BP7217 battery pack
- DPC/TRACK Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- One set of TP220 test probes
- Serial port cable
- Two sets of AC220 test clips
- Two sets of TL224 industrial test leads
|
|  | |
| CarePlans | Buy it | | G3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
| |  | | S3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
|  |  |
| Clips | Buy it | | AC220 | AC220 SureGrip™ Alligator Clips |  |  | | AC280 | AC280 SureGrip™ Hook Clips |  |  | | AC283 | AC283 SureGrip™ Pincer Clips |  |  | | AC285 | AC285 SureGrip™ Alligator Clips |  |  |
| Test Probes | Buy it | | TP920 | TP920 Test Probe Adapter Set |  |  |
| Differential Pressure Modules | Buy it | | 700P00 | 700P00 Differential Pressure Module |  | | | 700P01 | 700P01 Differential Pressure Module |  | | | 700P02 | 700P02 Differential Pressure Module |  | | | 700P03 | 700P03 Differential Pressure Module |  | | | 700P04 | 700P04 Differential Pressure Module |  | |
| Pressure Calibration Kit | Buy it | | 700PCK | Fluke 700PCK
Pressure Calibration Kit
|  | |
| Pumps and Accessories | Buy it | | 700PMP | 700 PMP Pressure Pump
|  |  | | 700HTH-1 | Fluke 700HTH-1 Hydraulic Test Hose |  |  | | 700PRV-1 | Fluke 700PRV-1 Pressure Relief Valve Kit |  |  | | 700PTP-1 | Fluke 700PTP-1 Pneumatic Test Pump |  |  | | 700HTP-2 | Fluke 700HTP-2 Hydraulic Test Pump |  |  |
| Differential Pressure Module Wet | Buy it | | 700P22 | 700P22 Differential Pressure Module Wet |  | | | 700P23 | 700P23 Differential Pressure Module Wet |  | | | 700P24 | 700P24 Differential Pressure Module Wet |  | |
| Absolute Pressure Modules | Buy it | | 700PA3 | 700PA3 Absolute Pressure Module |  | | | 700PA4 | 700PA4 Absolute Pressure Module |  | | | 700PA5 | 700PA5 Absolute Pressure Module |  | | | 700PA6 | 700PA6 Absolute Pressure Module |  | |
| Batteries, Chargers and Adapters | Buy it | | BC7217 | BC7217 Battery Charger |  |  | | BE9005 120 | BE9005 120 Battery Eliminator |  |  | | BP7235 | BP7235 NiMH Battery Pack |  |  |
| Cases and Holsters | Buy it | | C550 | C550 Tool Bag |  |  | | C700 | C700 Hard Carrying Case
|  |  | | C75 | C75 Accessory Case |  |  | | C781 | C781 Meter Case |  |  | | C789 | C789 Meter &
Accessory Case
|  |  |
| Connectors | Buy it | | 700BCA | Fluke 700BCA Bar Code Adapter |  | | | 700BCW | Fluke 700BCW Bar Code Wand
|  | | | Fluke 787CAL | Fluke 787CAL Calibration Module |  | |
| Current Clamps | Buy it | | i200 | i200 AC Current Clamp |  |  | | i200s | i200s AC Current Clamp |  |  |
| Dual Pressure Modules | Buy it | | 700PD2 | 700PD2 Dual Pressure Module |  | | | 700PD3 | 700PD3 Dual Pressure Module |  | | | 700PD4 | 700PD4 Dual Pressure Module |  | | | 700PD5 | 700PD5 Dual Pressure Module |  | | | 700PD6 | 700PD6 Dual Pressure Module |  | | | 700PD7 | 700PD7 Dual Pressure Module |  | |
| Software | Buy it | | 750SW | Fluke 750 SW DPC/TRACK2 Software |  |  |
| Vacuum Pressure Modules | Buy it | | 700PV3 | 700PV3 Vacuum Pressure Module |  | | | 700PV4 | 700PV4 Vacuum Pressure Module |  | |
| Temperature | Buy it | | 700TC2 | 700TC2 Thermocouple Plug Kit
700TC2
|  |  | | 80CJ-M | 80CJ-M type J Male Mini-Connectors
|  |  | | 80CK-M | 80CK-M type K Male Mini-Connectors
|  |  | | 80PJ-EXT | 80PJ-EXT Extension Wire Kit |  |  | | 80PK-1 | 80PK-1 Bead Probe |  |  | | 80PK-22 | 80PK-22 SureGrip™ Immersion Temperature Probe |  |  | | 80PK-24 | 80PK-24 SureGrip™ Air Temperature Probe |  |  | | 80PK-25 | 80PK-25 SureGrip™ Piercing Temperature Probe |  |  | | 80PK-26 | 80PK-26 SureGrip™ Tapered Temperature Probe |  |  | | 80PK-27 | 80PK-27 SureGrip™ Industrial Surface Temperature Probe |  |  | | 80PK-8 | 80PK-8 Pipe Clamp Temperature Probe |  |  | | 80PK-9 | 80PK-9 General Purpose Probe |  |  | | 80PK-EXT | 80PK-EXT Extension Wire Kit |  |  | | 80PT-25 | 80PT-25 SureGrip Piercing Temperature Probe |  |  |
| Test Leads | Buy it | | TL940 | TL940 Mini-Hook Test Leads |  |  | | TL950 | TL950 Mini-Pincer Test Leads |  |  |
|
|
|
|
|
|
|
|
|
|
|
|
| | |