Get More Calibration Power
The Fluke 726 Precision Multifunction Process Calibrator is designed specifically for the Process industry with broad workload coverage, calibration power and unsurpassed accuracy in mind. The 726 measures and sources almost all process parameters and can calibrate almost anything. The 726 will also interpret results without the help of a calculator and store measurement data for later analysis.
|
|
- More precise measurement and calibration source performance, accuracies of 0.01%.
- Two separate channels; measure, source and view process signals simultaneously
- Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
- Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
- Measure or source pressure* using any of 29 Fluke 700Pxx Pressure Modules
- Source mA with simultaneous pressure measurement to conduct valve and I/P tests
- Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
- Memory storage for up to 8 calibration results, return stored calibration data from the field for later analysis
- Transmitter error% calculation, interpret calibration results without a calculator
- Perform fast linearity tests with auto step and auto ramp features
- Power transmitters during test using 24 V loop supply and simultaneous mA measurement
- Store frequently-used test setups for later use
- Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
- HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
- Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement.
- New voltage input protection design for improved reliability
- Three-year warranty
|
*Pressure pump required
|
|
| Measurement Accuracy |
| Voltage DC |
| 30.000 V |
0.01%+ 2 mV (upper display) |
| 20.000 V |
0.01%+ 2 mV (lower display) |
| 90.00 mV |
0.01%+ 10μV |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
| | Current DC | | | Resistance |
| 0.0 to 4000.0 Ω |
+/- 0.015% |
| | Frequency |
| 2.0 to 1000.0 CPM |
0.05 % + 0.1 cpm |
| 1.0 to 1000.0 Hz |
0.05 % + 0.1 Hz |
| 1.00 to 15.00 kHz |
0.05 % + 0.01 kHz |
| Sensitivity |
1 V peak-to-peak-minimum |
| | Pressure |
| Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
|
| Source Accuracy |
| Voltage DC |
| 100.00 mV |
0.01%+10 μV |
| 20.000 V |
0.01%+2 mV |
| -10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
| | Current DC |
| 24.000 mA (Source) |
0.01%+ 2 μA |
| 24.000 mA (Simulate) |
0.01%+ 2 μA |
| | Resistance |
| Range, Accuracy (Percent of reading + floor) and Excitation Current |
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA |
|
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA |
|
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA |
|
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
| | Frequency |
| 2.0 to 1000.0 CPM |
0.05% |
| 1.0 to 1000.0 Hz |
0.05% |
| 1.0 to 10.0 kHz |
0.25% |
| 10.0 to 15.0 kHz |
0.5% |
| Waveform: |
5 V p-p squarewave, -0.1 V offset |
|
| RTDs and Thermocouples |
| Measure and Source Accuracy: |
| Cu10 |
1.8 °C |
| NI-120 |
0.15 °C |
| PT-100 (385) |
0.45 °C |
| PT-100 (392) |
0.35 °C |
| PT-100 (391-JIS) |
0.35 °C |
| PT-200 (385) |
0.95 °C |
| PT-500 (385) |
0.55 °C |
| PT-1000 (385) |
0.35 °C |
| RTD Resolution: |
0.01 C (0.01 F)
|
| Cu10.01 C |
0.1 C (0.1 F) |
| J: |
0.4 °C |
| K: |
0.5 °C |
| T: |
0.4 °C |
| E: |
0.4 °C |
| R: |
1.4 °C |
| S: |
1.4 °C |
| B: |
1.4 °C |
| L: |
0.45 °C |
| U: |
0.45 °C |
| N: |
0.6 °C |
| XK: |
0.4 °C |
| BP: |
1.1 °C |
| TC Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
|
| Specifications |
| Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
| | Loop Power Function |
| Voltage: |
24 V |
| Accuracy: |
10% |
| Maximum current: |
22 mA, short circuit protected |
| | Step Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Steps |
25% of range, 100% of range |
|
| Environmental Specifications |
| Operating Temperature | | | Storage Temperature | | | Operating Altitude | |
| Safety Specifications |
| Agency Approvals |
|
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
|
| Mechanical & General Specifications |
| Size |
|
200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in) |
| | Weight | | | Batteries | | | Battery Replacement |
|
Separate battery compartment, accessible without breaking calibration seall |
| | Side port connections |
|
Pressure module connector, also used for remote real-time programming |
|
|
| Model Name | Description | Price List | Buy it | | Fluke 726 |
Precision Multifunction Process Calibrator including:
- One set of AC72 alligator clips
- One set of stackable test leads
- One set of TL 75 testleads
- User manuals
- NIST- traceable calibration report and data
|
|  | |
| CarePlans | Buy it | | G3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
| |  | | S3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
|  |  |
| Clips | Buy it | | AC220 | AC220 SureGrip™ Alligator Clips |  |  | | AC280 | AC280 SureGrip™ Hook Clips |  |  | | AC283 | AC283 SureGrip™ Pincer Clips |  |  |
| Differential Pressure Modules | Buy it | | 700P00 | 700P00 Differential Pressure Module |  | | | 700P01 | 700P01 Differential Pressure Module |  | | | 700P02 | 700P02 Differential Pressure Module |  | | | 700P03 | 700P03 Differential Pressure Module |  | | | 700P04 | 700P04 Differential Pressure Module |  | |
| Pumps and Accessories | Buy it | | 700PMP | 700 PMP Pressure Pump
|  |  | | 700HTH-1 | Fluke 700HTH-1 Hydraulic Test Hose |  |  | | 700PRV-1 | Fluke 700PRV-1 Pressure Relief Valve Kit |  |  | | 700PTP-1 | Fluke 700PTP-1 Pneumatic Test Pump |  |  | | 700HTP-2 | Fluke 700HTP-2 Hydraulic Test Pump |  |  |
|
|
|
|
|
|
| | |