The most trusted tools in the world
Home  |  Products  |  Accessories  |  Test Leads, Probes and Clips  |  

2600A-101 SPRT Probe

 2600A-101
Secondary Platinum Resistance Thermometer Probe

The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinium Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard. The 2600A-101 has stripped leads suitable for connection to any Fluke data acquisition product or other screw terminal connector. This probe comes with calibration certificates and data for -200 to 6600°C.

Probe hook up diagram:

The probe connects to channel 1(input) and 11(excitation) on the Hydra input connector.

Warning: Probe handle surfaces are conductive. Use caution near electric circuits. Voltage levels above 30V rms, 42V peak, or 60 Vdc should not be applied to the probe.

General Specifications
Use Temperature 200 °C to 660 °C
Stability Typical draft < 0.01°C after 100 times from 25 to 660 °C
Calibrated Accuracy
Temperature  Uncertainty
- 183°C  +/- 0.01°C
0°C  +/- 0.007°C
200°C  +/- 0.024°C
420°C  +/- 0.033°C
660 °C +/- 0.050 °C
Performance Specifications (nominal)
Ice point resistance:  100 Ω ± 0.1 Ω
Alpha value:  0.00385 ± 0.00001 Ω/Ω/C
Insulation resistance:  1000 Meg Ω at 100 Vdc at 20°C

 100 Meg Ohms at 100 Vdc at 400°C
Hysteresis:  0.01°C Max. using 0, 400°C as the end points
Self heating:  0.02°C / mw in well stirred bath
Thermoelectric voltage:  When tested in an ice bath, with immersion depth from 4 to 10 inches (10 to 25 cm), thermo-electric voltage will not exceed 2 µV
Time constant:  9 seconds typical for 63.2% response to step change
Long term stability:  Stability at 0°C after exposure to upper temperature limits
Dimensions
Sheath diameter (D):  .25 inches (6.3 mm)
Length (L):  12 inches (0.3 m)
Cable length (Y):  6 feet (1.83 m)
Wire:  22 AWG, Teflon jacketed, rated to 200 C
Wire terminations:  flying lead; stripped / tinned
Model NameDescriptionBuy it