You’re going to love what we’ve done with your calibrator.
The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.
The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.
For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2™ software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.
The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
|
Fluke 754 Documenting Process Calibrator-HART
4
5
5
5
Very limited HART functionality.
Can't update or load "HART Device Descriptors" which means very limited HART functionality. Do not make the mistake of thinking that you can substitute the Fluke 754 for a HART communicator. For many devices, you will need to bring the calibrator AND a real HART communicator to calibrate the instrument. There are modern times and I feel that Fluke really dropped the ball here.
May 1, 2013
Good but the HART needs work
I convinced my boss to buy this over an emmerson communicator. Come to find out it does not support all specific device DD's. The first instrument I used it on it communicated HART fine but could not display all the DD parameters. I called support and was told that when a firmware upgrade came out it might support more HART DD's but they didn't know if/when this would happen. Had to use an Emmerson communicator after all. As far as a calibrator its great. The pressure modules and hand pump are great. The DPCTrack2 software is good to keep track of everything. Dont plan on replacing a communicator with this one instrument though.
December 13, 2012
This does everything the 744 did; BETTER and LONGER!
For well over a dozen years I have used the 744 and when I heard news about the impending launch of the 754 I was salivating like Pavlov's Dogs in anticipation! We have hundreds of 744s throughout all of our divisions and we are already over a dozen 754s. The advances you have made in the legibility of the display out in varying light conditions and the battery life are worth the cost of admission alone. You have gone well beyond there with increased precision and reliability, and now the HART works with all the Rosemount Transmitters; I'd say you not only hit a homer, but hit a grand slam!
June 13, 2012
This is what i expected from fluke
great new color display! extra features that the 744 doesnt have are a plus......just what I needed to upgrade!
June 8, 2012
Very easy to use
I have been using this meter to calibrate everyday for several weeks now and it definitely delivers, it's a huge time saver, well designed, simple to operate, and well worth the investment.
June 7, 2012
|
Work smarter. Work faster.
The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART® interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
- Power transmitters during test using loop supply with simultaneous mA measurement
- Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
- Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
- Holds up to a full week of downloaded procedures and calibration results.
- Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
- Easy to use
- Three year warranty
- Bright white dual display. Read both sourced and measured parameters simultaneously.
- Multi lingual interface
- Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
- Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
- Delivered with DPC/Track Sample software.
- Compatible with many Asset Management software packages.
|
HART Capabilities
The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.
- Requires no external box or second tool for everyday HART calibration and maintenance.
- Offers fast HART communication.
- Supports popular models of HART transmitters, with more devise-specific command support than any other HART field calibrator.
- Works with multiple masters, burst mode, and multi-drop configurations.
- Is easy to update as additional instruments are added and new HART versions are released.
- Interrogate to determine device type, manufacturer, model, tag.
- Reconfigure the sensor mapping of dual sensor temperature transmitters.
- Read HART PV function and smart transmitter digital output while measuring analog mA output.
- Read and write HART configuration functions to make field adjustments to PV range points, damping, and other top-level configuration settings.
- Re-label smart transmitters by reading and writing to the HART tag field.
|
Versatile HART Protocol support
| Manufacturer |
Pressure instruments |
Temperature instruments |
Coriolis instruments |
| ABB/Kent-Taylor |
600T |
658T1 |
|
| ABB/Hartmann & Braun |
Contrans P1,
AS 800 Series |
|
|
| Endress & Hauser |
CERABAR S,
CERABAR M,
DELTABAR S |
TMT 1221,TMT 1821,
TMT 1621 |
|
| Foxboro Eckardt |
|
TI/RTT201 |
|
| Foxboro/Invensys |
I/A Pressure |
|
|
| Fuji |
FCX
FCXAZ |
FRC |
|
| Honeywell |
ST3000 |
STT25T1,
STT25H1 |
|
| Micro Motion |
|
|
2000
2000 IS
9701
9712
9739 |
| Moore Products |
|
3441 |
|
| Rosemount |
1151
2088
3001C
3051, 3051S |
3044C
644
3144
3244, 3144P |
|
| Siemens |
SITRANS P DS
SITRANS P ES |
|
|
| SMAR |
LD301 |
TT3011 |
|
| Viatran |
I/A Pressure |
|
|
| Wika |
UNITRANS |
T32H1 |
|
| Yokogawa |
EJA |
YTA 110, 310
and 320 |
|
1Sensor Trim not supported
The 754 supports the commands contained in HART protocol Version 5.7. With 2 MB of memory, the 754 supports a substantial set of HART instructions:
- Universal commands provide functions that are implemented in all field devices, for example, read manufacturer and device type, read primary variable (PV), or read current output and percent of span.
- Common practice commands provide functions that are common to many but not all field devices, for example read multiple variables, set damping time, or perform loop test.
- Device-specific commands provide functions that are unique to a particular field device, for example sensor trim.
|
HART Operating Modes Supported
- Point to Point operation , the most commonly used mode, connects the 754 to a single HART device in a 4-20 mA loop.
- In Multi-Drop mode , several HART instruments can be bussed together. The 754 searches for each, identifies addresses in use, and allows you to select the instrument for calibration and related operations.
- In Burst Mode, the HART instrument transmits bursts of data without waiting to be interrogated by a master unit. The 754 can take transmitters out of burst mode during test or calibration, then later restore them to burst mode.
|
Pressure Modules
Covers virtually any pressure application including gauge, differential, dual (compound), absolute, and vacuum.
- Display pressure readings in any of 10 different pressure units you specify in the calibrator setup.
- Rugged urethane molded cases protect the modules from rough handling and harsh conditions.
- Features internal temperature compensation from 0º to 50º C for full accuracy performance.
- Includes NIST-traceable calibration certificate.
- Modules can be calibrated locally, helping to control costs.
|
A family of 29 optional pressure modules provides pressure calibration and measurement capabilities. Twenty-eight modules are available, with basic accuracy specs to 0.05%. Ranges start at 0 — 1 in H20 (0 — .25kPa) and go to 0 — 10,000 psi (0 — 70,000 kPa). Additional information about pressure modules is available on the Pressure Modules Home Page.
Automated Procedures
Allow you to quickly set up powerful, automated calibration procedures for linear transmitters, DP flow transmitters, and one- and two-point limit switches. Simply select the appropriate measure and/or source functions and fill out the procedure template. The 750 Series does the rest. It quickly performs the test, calculates the errors, and displays the final results, highlighting out of tolerance points.
Custom Units
Enable you to map one unit to another, such as mV to ºC or ºF. Allows you to use the Fluke 750 Series with millivolt output accessories such as the Fluke 80T-IR Temperature Probe, and to document tests using non-supported units such as parts per million or revolutions per minute.
User-entered Values
Enables technicians to record calibration results that were sourced and/or measured by other devices such as panel meters or readout-only devices.
Limit Switch Calibration
Procedures perform fast, automated calibration of one and two-point limit switches for voltage, current, temperature, and pressure.
Differential Pressure Flow Instrument Calibration
Routines use a square root function to directly calibrate DP flow instruments.
Additional Features
Multifunctional
Calibrate temperature, pressure, voltage, current, resistance, and frequency. Since it both measures and sources, you can troubleshoot and calibrate all with one rugged tool.
Powerful, yet easy to use
The easy-to-follow menu-driven display guides you through any task. Get up to speed in minutes, not days. Programmable calibration routines enable you to create and run automated as-found/as-left procedures to ensure fast, consistent, calibrations.
Records and documents results
To support your ISO-9000 or regulatory standards the Fluke 754 captures your calibration results, eliminating the need to juggle a pen and pad in the field. The USB interface lets you transfer the results to a PC, thus saving the time of having to manually transcribe them when you return to the shop.
Truly hand-held
Small enough to fit easily into a tool bag and to use in tight spaces. Runs an entire shift on a rechargeable Li-ion battery pack.
Rugged and reliable
Count on Fluke's rugged design to deliver top accuracy and reliability in harsh environments. Overmolded urethane case stands up to rough handling in industrial environments.
Bright white display
Lets you read your results in any kind of light. Backlight has three (3) settings.
Soft keys
Provide one-touch access to enhanced functions such as task lists, automated procedures, scaling, min/max, stepping and ramping, and review memory.
Three operating modes
Measure, Source, or simultaneous Measure/Source, - enable technicians to troubleshoot, calibrate, or maintain instrumentation with just one tool.
Multi-lingual interface
Displays instructions in English, French, German, Spanish, and Italian.
Built-in algebraic calculator
With four functions-plus square root-stores, recalls, and performs calculations required for setting up instruments or evaluating data in the field. Use it to set the source function to a calculated value. There's no need to carry a pencil and paper or a separate calculator.
Programmable measurement delay
Inside automated procedures permits calibrating instruments that respond slowly.
|
|
| Measurement Accuracy |
| Voltage DC |
|
1 Year |
2 Years |
| 100.000 mV |
0.02%+0.005 mV |
0.03%+0.005 mV |
| 3.00000 V |
0.02%+0.00005 V |
0.03%+0.00005 V |
| 30.0000 V |
0.02%+0.0005 V |
0.03%+0.0005 V |
| 300.00 V |
0.05%+0.05 V |
0.07%+0.05 V |
|
| Voltage AC |
Range
40 to 500 Hz |
Resolution |
1 Year |
2 Years |
| 3.000 V |
0.001 V |
0.5%+0.002 V |
1.0%+0.004 V |
| 30.00 V |
0.01 V |
0.5%+0.02 V |
1.0%+0.04 V |
| 300.0 V |
0.1 V |
0.5%+0.2 V |
1.0%+0.2 V |
|
| Current DC |
| |
1 Year |
2 Years |
| 30.000 mA |
0.01% + 5 uA |
0.015%+7 uA |
| 110.00 mA |
0.01% + 20 uA |
0.015%+30 uA |
|
| Resistance |
| |
1 Year |
2 Years |
10.000 Ω
|
0.05% + 50 mΩ |
0.07%+70 mΩ |
| 100.00 Ω |
0.05% + 50 mΩ |
0.07%+70 mΩ |
| 1.0000 kΩ |
0.05% + 500 mΩ |
0.07%+0.5 Ω |
| 10.000 kΩ |
0.1% + 10 Ω |
0.15%+15 Ω |
|
| Frequency |
| |
Resolution |
Accuracy (2 Years) |
| 1.00 to 110.00 Hz |
0.01 Hz |
0.05 Hz |
| 110.1 to 1100.0 Hz |
0.1 Hz |
0.5 Hz |
| 1.101 to 11.000 kHz |
0.001 kHz |
0.005 kHz |
| 11.01 to 50.00 kHz |
0.01 kHz |
0.05 kHz |
|
| Source Accuracy |
| Voltage DC |
| |
1 Year |
2 Years |
100.000 mV
|
0.01%+0.005 mV |
0.015%+0.005 mV |
| 1.00000 V |
0.01%+0.00005 V |
0.015%+0.0005 V |
| 15.0000 V |
0.01%+0.0005 V |
0.015%+0.0005 V |
|
| Current DC |
| |
1 Year |
2 Years |
| 22.000 mA (Source) |
0.01%+ 0.003 mA |
0.02%+ 0.003 mA |
| Current Sink (Simulate) |
0.02% + 0.007 mA |
0.04% + 0.007 mA |
|
| Resistance |
| |
1 Year |
2 Years |
10.000 Ω
|
0.01% + 10 mΩ |
0.015% + 15 mΩ |
| 100.00 Ω |
0.01% + 20 mΩ |
0.015% + 30 mΩ |
| 1.0000 kΩ |
0.02% + 0.2 Ω |
0.03% + 0.3 Ω |
| 10.000 kΩ |
0.02% + 3 Ω |
0.03% + 5 Ω |
|
| Frequency |
| |
2 Years |
| 0.1 to 10.99 Hz |
0.01 Hz |
| 0.01 to 10.99 Hz |
0.01 Hz |
| 11.00 to 109.99 Hz |
0.1 Hz |
| 110.0 to 1099.9 Hz |
0.1 Hz |
| 1.100 to 21.999 kHz |
0.002 kHz |
22.000 to 50.000 kHz
|
0.005 kHz |
|
| Technical Data |
| Data Log Functions |
| Measure functions: |
Voltage, current, resistance, frequency, temperature, pressure |
| Reading rate: |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
| Maximum record length: |
8000 readings (7980 for 30 or 60 readings/minute) |
|
| Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Rate: |
4 steps/second |
| Trip detect: |
Continuity or voltage (continuity detection not available when sourcing current) |
|
| Loop Power Function |
| Voltage: |
Selectable, 26 V |
| Accuracy: |
10%, 18 V minimum at 22 mA |
| Maximum current: |
25 mA, short circuit protected |
| Maximum input voltage: |
50 V DC |
|
| Step Functions |
| Source Functions |
Voltage, current, resistance, frequency, temperature |
| Manual Step |
Selectable step, change with arrow buttons |
| Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
|
| Environmental Specifications |
| Operating Temperature |
|
| Storage Temperature |
|
| Dust/water resistance |
|
| Operating Altitude |
|
3000 m above mean sea level (9842 ft) |
|
| Safety Specifications |
| Agency Approvals |
|
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
|
| Mechanical & General Specifications |
| Size |
|
136 x 245 x 63 mm (5.4“ x 9.6 x 2.5) |
|
| Weight |
|
| Batteries |
| Internal Battery Pack Li-ion: |
7.2V, 4400mAh, 30 Wh |
|
| Battery Life |
|
| Battery Replacement |
|
Replace without opening calibrator; no tools required |
|
| Side port connections |
|
Pressure module connector |
|
USB Connector to interface to your PC |
| |
Digital instrument (HART) connector |
|
Connection for optional battery charger/eliminator |
|
| Data storage capacity |
|
1 week of calibration procedures results |
|
| 90 day specifications |
|
The standard specification interval for the 750 Series are 1 and 2 years. |
|
Typical 90 day measurement and source accuracy can be estimated by |
|
dividing the one year "% of reading" or "%of output" specifications by 2. |
|
Floor specifications, expressed as "% of full scale" or |
|
"counts" or "ohms" remain constant. |
|
| Temperature, Resistance Temperature Detectors |
| Degrees or % of reading |
| Type (α) |
Range °C |
Measure °C1 |
Source current |
Source °C |
Allowable current2 |
| |
1 year |
2 year |
|
1 year |
2 year |
|
| 100 Ω Pt (385) |
-200 to 100
100 to 800 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 200 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
500 μA |
0.06 °C
0.017 % + 0.05 °C |
0.12 °C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 500 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
250 μA |
0.06 °C
0.017% + 0.05 °C |
0.12 °C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 1000 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04% + 0.10 °C |
150 μA |
0.06 C
0.017 % + 0.05 °C |
0.12 C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 100 Ω Pt (3916) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % +0.05 °C |
0.14 °C
0.04 % +0.10 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 100 Ω Pt (3926) |
-200 to 100
100 to 630 |
0.08 °C
0.02 % +0.06 °C |
0.16 °C
0.04 % +0.12 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 10 Ω Cu (427) |
-100 to 260 |
0.2 °C |
0.4 °C |
3 mA |
0.2 °C |
0.4 °C |
0.1 mA to 10 mA |
| 120 Ω Ni (672) |
-80 to 260 |
0.1 °C |
0.2 °C |
1 mA |
0.04 °C |
0.08 °C |
0.1 mA to 10 mA |
1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
| Temperature, Thermocouples |
| Type |
Source °C |
Measure °C |
Source °C |
|
|
1 year |
2 years |
1 year |
2 years |
| E |
-250 to -200 |
1.3 |
2.0 |
0.6 |
0.9 |
|
-200 to -100 |
0.5 |
0.8 |
0.3 |
0.4 |
|
-100 to 600 |
0.3 |
0.4 |
0.3 |
0.4 |
|
600 to 1000 |
0.4 |
0.6 |
0.2 |
0.3 |
| N |
-200 to -100 |
1.0 |
1.5 |
0.6 |
0.9 |
|
-100 to 900 |
0.5 |
0.8 |
0.5 |
0.8 |
|
900 to 1300 |
0.6 |
0.9 |
0.3 |
0.4 |
| J |
-210 to -100 |
0.6 |
0.9 |
0.3 |
0.4 |
|
-100 to 800 |
0.3 |
0.4 |
0.2 |
0.3 |
|
800 to 1200 |
0.5 |
0.8 |
0.3 |
0.3 |
| K |
-200 to -100 |
0.7 |
1.0 |
0.4 |
0.6 |
|
-100 to 400 |
0.3 |
0.4 |
0.3 |
0.4 |
|
400 to 1200 |
0.5 |
0.8 |
0.3 |
0.4 |
|
1200 to 1372 |
0.7 |
1.0 |
0.3 |
0.4 |
| T |
-250 to -200 |
1.7 |
2.5 |
0.9 |
1.4 |
|
-200 to 0 |
0.6 |
0.9 |
0.4 |
0.6 |
|
0 to 400 |
0.3 |
0.4 |
0.3 |
0.4 |
| B |
600 to 800 |
1.3 |
2.0 |
1.0 |
1.5 |
|
800 to 1000 |
1.0 |
1.5 |
0.8 |
1.2 |
|
1000 to 1820 |
0.9 |
1.3 |
0.8 |
1.2 |
| R |
-20 to 0 |
2.3 |
2.8 |
1.2 |
1.8 |
|
0 to 100 |
1.5 |
2.2 |
1.1 |
1.7 |
|
100 to 1767 |
1.0 |
1.5 |
0.9 |
1.4 |
| S |
-20 to 0 |
2.3 |
2.8 |
1.2 |
1.8 |
|
0 to 200 |
1.5 |
2.1 |
1.1 |
1.7 |
|
200 to 1400 |
0.9 |
1.4 |
0.9 |
1.4 |
|
1400 to 1767 |
1.1 |
1.7 |
1.0 |
1.5 |
| C |
0 to 800 |
0.6 |
0.9 |
0.6 |
0.9 |
|
800 to 1200 |
0.8 |
1.2 |
0.7 |
1.0 |
|
1200 to 1800 |
1.1 |
1.6 |
0.9 |
1.4 |
|
1800 to 2316 |
2.0 |
3.0 |
1.3 |
2.0 |
| L |
-200 to -100 |
0.6 |
0.9 |
0.3 |
0.4 |
|
-100 to 800 |
0.3 |
0.4 |
0.2 |
0.3 |
|
800 to 900 |
0.5 |
0.8 |
0.2 |
0.3 |
| U |
-200 to 0 |
0.6 |
0.9 |
0.4 |
0.6 |
| 0 to 600 |
0.3 |
0.4 |
0.3 |
0.4 |
| BP |
0 to 1000 |
1.0 |
1.5 |
0.4 |
0.6 |
|
1000 to 2000 |
1.6 |
2.4 |
0.6 |
0.9 |
|
2000 to 2500 |
2.0 |
3.0 |
0.8 |
1.2 |
| XK |
-200 to 300 |
0.2 |
0.3 |
0.2 |
0.5 |
| 300 to 800 |
0.4 |
0.6 |
0.3 |
0.6 |
|
| Model Name | Description | Price List | Buy it | | Fluke 754 | Documenting Process Calibrator-HART including:
- BC7240 battery charger
- Li-on BP7240 battery pack
- DPCTrack 2™ Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- Three sets of TP220 test probes with three sets of “extended tooth alligator clips
- Two sets AC280 hook clips
- C799 Soft Field Case
- USB communication cable, Fluke 754HHC HART communications cable
|
|  | |
| CarePlans | Buy it | | G3P-Group2 |
- GOLD CarePlan.
- Models covered: 50 Series II (51 II, 52 II, 53 II, 54 II), 190 Series II (190-062/AM, 190-062/AM/S, 190-102/AM, 190-102/AM/S, 190-104/AM, 190-104/AM/S, 190-202/AM, 190-202/AM/S, 190-204/AM, 190-204/AM/S, 190-502/AM, 190-502/AM/S), 381, 705, 707, 709, 709H, 712, 714, 715, 717, 718, 719, 754, 787, 789, 810, 1550C, 1577, and 1587.
- Extends warranty from three to five years and includes five annual NIST traceable calibrations with data.
| |  | | S3P-Group2 |
- SILVER CarePlan.
- Models covered: 50 Series II (51 II, 52 II, 53 II, 54 II), 190 Series II (190-062/AM, 190-062/AM/S, 190-102/AM, 190-102/AM/S, 190-104/AM, 190-104/AM/S, 190-202/AM, 190-202/AM/S, 190-204/AM, 190-204/AM/S, 190-502/AM, 190-502/AM/S), 381, 705, 707, 709, 709H, 712, 714, 715, 717, 718, 719, 754, 787, 789, 810, 1550C, 1577, and 1587.
- Extends warranty from three to five years.
|  |  |
| Clips | Buy it | | AC220 | AC220 SureGrip™ Alligator Clips |  |  | | AC280 | AC280 SureGrip™ Hook Clips |  |  | | AC285 | AC285 SureGrip™ Alligator Clips |  |  |
| Test Probes | Buy it | | TP74 | TP74 Lantern Tip Test Probes |  |  | | TP920 | TP920 Test Probe Adapter Set |  |  |
| Differential Pressure Modules | Buy it | | 700P00 | 700P00 Differential Pressure Module |  | | | 700P01 | 700P01 Differential Pressure Module |  | | | 700P02 | 700P02 Differential Pressure Module |  | | | 700P03 | 700P03 Differential Pressure Module |  | | | 700P04 | 700P04 Differential Pressure Module |  | |
| Pressure Calibration Kit | Buy it | | 700PCK | Fluke 700PCK
Pressure Calibration Kit
|  | |
| Pumps and Accessories | Buy it | | 700PMP | 700 PMP Pressure Pump
|  |  | | 700HTH-1 | Fluke 700HTH-1 Hydraulic Test Hose |  |  | | 700PRV-1 | Fluke 700PRV-1 Pressure Relief Valve Kit |  |  | | 700PTP-1 | Fluke 700PTP-1 Pneumatic Test Pump |  |  | | 700HTP-2 | Fluke 700HTP-2 Hydraulic Test Pump |  |  |
| Differential Pressure Module Wet | Buy it | | 700P22 | 700P22 Differential Pressure Module Wet |  | | | 700P23 | 700P23 Differential Pressure Module Wet |  | | | 700P24 | 700P24 Differential Pressure Module Wet |  | |
| Absolute Pressure Modules | Buy it | | 700PA3 | 700PA3 Absolute Pressure Module |  | | | 700PA4 | 700PA4 Absolute Pressure Module |  | | | 700PA5 | 700PA5 Absolute Pressure Module |  | | | 700PA6 | 700PA6 Absolute Pressure Module |  | |
| Batteries, Chargers and Adapters | Buy it | | BC7240 | BC7240 Battery charger/Eliminator for 75x | | | | BP7240 | BP7240 Li-Ion Battery Pack for 75x |  | |
| Cases and Holsters | Buy it | | C700 | C700 Hard Carrying Case
|  |  | | C781 | C781 Meter Case |  |  | | C799 | Soft Field Case |  |  |
| Dual Pressure Modules | Buy it | | 700PD2 | 700PD2 Dual Pressure Module |  | | | 700PD3 | 700PD3 Dual Pressure Module |  | | | 700PD4 | 700PD4 Dual Pressure Module |  | | | 700PD5 | 700PD5 Dual Pressure Module |  | | | 700PD6 | 700PD6 Dual Pressure Module |  | | | 700PD7 | 700PD7 Dual Pressure Module |  | |
| Software | Buy it | | 750SW | Fluke 750 SW DPC/TRACK2 Software |  |  |
| Test Leads, Probes and Clips | Buy it | | 700TLK | Fluke 700TLK Process Test lead kit |  | |
| Vacuum Pressure Modules | Buy it | | 700PV3 | 700PV3 Vacuum Pressure Module |  | | | 700PV4 | 700PV4 Vacuum Pressure Module |  | |
| Temperature | Buy it | | 80CJ-M | 80CJ-M type J Male Mini-Connectors
|  |  | | 80CK-M | 80CK-M type K Male Mini-Connectors
|  |  | | 80PJ-EXT | 80PJ-EXT Extension Wire Kit |  |  | | 80PK-1 | 80PK-1 Bead Probe |  |  | | 80PK-22 | 80PK-22 SureGrip™ Immersion Temperature Probe |  |  | | 80PK-24 | 80PK-24 SureGrip™ Air Temperature Probe |  |  | | 80PK-25 | 80PK-25 SureGrip™ Piercing Temperature Probe |  |  | | 80PK-26 | 80PK-26 SureGrip™ Tapered Temperature Probe |  |  | | 80PK-27 | 80PK-27 SureGrip™ Industrial Surface Temperature Probe |  |  | | 80PK-8 | 80PK-8 Pipe Clamp Temperature Probe |  |  | | 80PK-9 | 80PK-9 General Purpose Probe |  |  | | 80PK-EXT | 80PK-EXT Extension Wire Kit |  |  | | 80PT-25 | 80PT-25 SureGrip Piercing Temperature Probe |  |  |
| Test Leads | Buy it | | TL940 | TL940 Mini-Hook Test Leads |  |  | | TL950 | TL950 Mini-Pincer Test Leads |  |  |
|
|
|
|
|
|
|
|
|
|
|
|
| | |