You’re going to love what we’ve done with your calibrator.
The 753 is everything you needed and wished for in the 743B and more!
The 753 does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You'll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.
For documentation, the 753 automates calibration procedures and captures your data.
And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
The Fluke 753 Documenting Process Calibrator: Work smarter. Work faster.
|
Work smarter. Work faster.
The 753 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis.
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
- Power transmitters during test using loop supply with simultaneous mA measurement
- Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
- Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
- Holds up to a full week of downloaded procedures and calibration results.
- Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
- Easy to use
- Three year warranty
- Bright white dual display. Read both sourced and measured parameters simultaneously.
- Multi lingual interface
- Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
- Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
- Delivered with DPC/Track Sample software.
- Compatible with many Asset Management software packages.
|
Pressure Modules
Covers virtually any pressure application including gauge, differential, dual (compound), absolute, and vacuum.
- Display pressure readings in any of 10 different pressure units you specify in the calibrator setup.
- Rugged urethane molded cases protect the modules from rough handling and harsh conditions.
- Features internal temperature compensation from 0º to 50º C for full accuracy performance.
- Includes NIST-traceable calibration certificate.
- Modules can be calibrated locally, helping to control costs.
|
A family of 29 optional pressure modules provides pressure calibration and measurement capabilities. Twenty-eight modules are available, with basic accuracy specs to 0.05%. Ranges start at 0 — 1 in H20 (0 — .25kPa) and go to 0 — 10,000 psi (0 — 70,000 kPa). Additional information about pressure modules is available on the Pressure Modules Home Page.
Automated Procedures
Allow you to quickly set up powerful, automated calibration procedures for linear transmitters, DP flow transmitters, and one- and two-point limit switches. Simply select the appropriate measure and/or source functions and fill out the procedure template. The 750 Series does the rest. It quickly performs the test, calculates the errors, and displays the final results, highlighting out of tolerance points.
Custom Units
Enable you to map one unit to another, such as mV to ºC or ºF. Allows you to use the Fluke 750 Series with millivolt output accessories such as the Fluke 80T-IR Temperature Probe, and to document tests using non-supported units such as parts per million or revolutions per minute.
User-entered Values
Enables technicians to record calibration results that were sourced and/or measured by other devices such as panel meters or readout-only devices.
Limit Switch Calibration
Procedures perform fast, automated calibration of one and two-point limit switches for voltage, current, temperature, and pressure.
Differential Pressure Flow Instrument Calibration
Routines use a square root function to directly calibrate DP flow instruments.
Additional Features
Multifunctional
Calibrate temperature, pressure, voltage, current, resistance, and frequency. Since it both measures and sources, you can troubleshoot and calibrate all with one rugged tool.
Powerful, yet easy to use
The easy-to-follow menu-driven display guides you through any task. Get up to speed in minutes, not days. Programmable calibration routines enable you to create and run automated as-found/as-left procedures to ensure fast, consistent, calibrations.
Records and documents results
To support your ISO-9000 or regulatory standards the Fluke 753 captures your calibration results, eliminating the need to juggle a pen and pad in the field. The USB interface lets you transfer the results to a PC, thus saving the time of having to manually transcribe them when you return to the shop.
Truly hand-held
Small enough to fit easily into a tool bag and to use in tight spaces. Runs an entire shift on a rechargeable Li-ion battery pack.
Rugged and reliable
Count on Fluke's rugged design to deliver top accuracy and reliability in harsh environments. Overmolded urethane case stands up to rough handling in industrial environments.
Bright white display
Lets you read your results in any kind of light. Backlight has three (3) settings.
Soft keys
Provide one-touch access to enhanced functions such as task lists, automated procedures, scaling, min/max, stepping and ramping, and review memory.
Three operating modes
Measure, Source, or simultaneous Measure/Source, - enable technicians to troubleshoot, calibrate, or maintain instrumentation with just one tool.
Multi-lingual interface
Displays instructions in English, French, German, Spanish, and Italian.
Built-in algebraic calculator
With four functions-plus square root-stores, recalls, and performs calculations required for setting up instruments or evaluating data in the field. Use it to set the source function to a calculated value. There's no need to carry a pencil and paper or a separate calculator.
Programmable measurement delay
Inside automated procedures permits calibrating instruments that respond slowly.
|
|
| Measurement Accuracy |
| Voltage DC |
|
1 Year |
2 Years |
| 100.000 mV |
0.02%+0.005 mV |
0.03%+0.005 mV |
| 3.00000 V |
0.02%+0.00005 V |
0.03%+0.00005 V |
| 30.0000 V |
0.02%+0.0005 V |
0.03%+0.0005 V |
| 300.00 V |
0.05%+0.05 V |
0.07%+0.05 V |
|
| Voltage AC |
Range
40 to 500 Hz |
Resolution |
1 Year |
2 Years |
| 3.000 V |
0.001 V |
0.5%+0.002 V |
1.0%+0.004 V |
| 30.00 V |
0.01 V |
0.5%+0.02 V |
1.0%+0.04 V |
| 300.0 V |
0.1 V |
0.5%+0.2 V |
1.0%+0.2 V |
|
| Current DC |
| |
1 Year |
2 Years |
| 30.000 mA |
0.01% + 5 uA |
0.015%+7 uA |
| 110.00 mA |
0.01% + 20 uA |
0.015%+30 uA |
|
| Resistance |
| |
1 Year |
2 Years |
10.000 Ω
|
0.05% + 50 mΩ |
0.07%+70 mΩ |
| 100.00 Ω |
0.05% + 50 mΩ |
0.07%+70 mΩ |
| 1.0000 kΩ |
0.05% + 500 mΩ |
0.07%+0.5 Ω |
| 10.000 kΩ |
0.1% + 10 Ω |
0.15%+15 Ω |
|
| Frequency |
| |
Resolution |
Accuracy (2 Years) |
| 1.00 to 110.00 Hz |
0.01 Hz |
0.05 Hz |
| 110.1 to 1100.0 Hz |
0.1 Hz |
0.5 Hz |
| 1.101 to 11.000 kHz |
0.001 kHz |
0.005 kHz |
| 11.01 to 50.00 kHz |
0.01 kHz |
0.05 kHz |
|
| Source Accuracy |
| Voltage DC |
| |
1 Year |
2 Years |
100.000 mV
|
0.01%+0.005 mV |
0.015%+0.005 mV |
| 1.00000 V |
0.01%+0.00005 V |
0.015%+0.0005 V |
| 15.0000 V |
0.01%+0.0005 V |
0.015%+0.0005 V |
|
| Current DC |
| |
1 Year |
2 Years |
| 22.000 mA (Source) |
0.01%+ 0.003 mA |
0.02%+ 0.003 mA |
| Current Sink (Simulate) |
0.02% + 0.007 mA |
0.04% + 0.007 mA |
|
| Resistance |
| |
1 Year |
2 Years |
10.000 Ω
|
0.01% + 10 mΩ |
0.015% + 15 mΩ |
| 100.00 Ω |
0.01% + 20 mΩ |
0.015% + 30 mΩ |
| 1.0000 kΩ |
0.02% + 0.2 Ω |
0.03% + 0.3 Ω |
| 10.000 kΩ |
0.02% + 3 Ω |
0.03% + 5 Ω |
|
| Frequency |
| |
2 Years |
| 0.1 to 10.99 Hz |
0.01 Hz |
| 0.01 to 10.99 Hz |
0.01 Hz |
| 11.00 to 109.99 Hz |
0.1 Hz |
| 110.0 to 1099.9 Hz |
0.1 Hz |
| 1.100 to 21.999 kHz |
0.002 kHz |
22.000 to 50.000 kHz
|
0.005 kHz |
|
| Technical Data |
| Data Log Functions |
| Measure functions: |
Voltage, current, resistance, frequency, temperature, pressure |
| Reading rate: |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
| Maximum record length: |
8000 readings (7980 for 30 or 60 readings/minute) |
|
| Ramp Functions |
| Source functions: |
Voltage, current, resistance, frequency, temperature |
| Rate: |
4 steps/second |
| Trip detect: |
Continuity or voltage (continuity detection not available when sourcing current) |
|
| Loop Power Function |
| Voltage: |
Selectable, 26 V |
| Accuracy: |
10%, 18 V minimum at 22 mA |
| Maximum current: |
25 mA, short circuit protected |
| Maximum input voltage: |
50 V DC |
|
| Step Functions |
| Source Functions |
Voltage, current, resistance, frequency, temperature |
| Manual Step |
Selectable step, change with arrow buttons |
| Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
|
| Environmental Specifications |
| Operating Temperature |
|
| Storage Temperature |
|
| Dust/water resistance |
|
| Operating Altitude |
|
3000 m above mean sea level (9842 ft) |
|
| Safety Specifications |
| Agency Approvals |
|
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
|
| Mechanical & General Specifications |
| Size |
|
136 x 245 x 63 mm (5.4“ x 9.6 x 2.5) |
|
| Weight |
|
| Batteries |
| Internal Battery Pack Li-ion: |
7.2V, 4400mAh, 30 Wh |
|
| Battery Life |
|
| Battery Replacement |
|
Via snap-shut door without opening calibrator; no tools required |
|
| Side port connections |
|
Pressure module connector |
|
USB Connector to interface to your PC |
|
Connection for optional battery charger/eliminator |
|
| Data storage capacity |
|
1 week of calibration procedures results |
|
| 90 day specifications |
|
The standard specification interval for the 750 Series are 1 and 2 years. |
|
Typical 90 day measurement and source accuracy can be estimated by |
|
dividing the one year "% of reading" or "%of output" specifications by 2. |
|
Floor specifications, expressed as "% of full scale" or |
|
"counts" or "ohms" remain constant. |
|
| Temperature, Resistance Temperature Detectors |
| Degrees or % of reading |
| Type (α) |
Range °C |
Measure °C1 |
Source current |
Source °C |
Allowable current2 |
| |
1 year |
2 year |
|
1 year |
2 year |
|
| 100 Ω Pt (385) |
-200 to 100
100 to 800 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 200 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
500 μA |
0.06 °C
0.017 % + 0.05 °C |
0.12 °C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 500 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04 % + 0.10 °C |
250 μA |
0.06 °C
0.017% + 0.05 °C |
0.12 °C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 1000 Ω Pt (385) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % + 0.05 °C |
0.14 °C
0.04% + 0.10 °C |
150 μA |
0.06 C
0.017 % + 0.05 °C |
0.12 C
0.034 % + 0.10 °C |
0.1 mA to 1 mA |
| 100 Ω Pt (3916) |
-200 to 100
100 to 630 |
0.07 °C
0.02 % +0.05 °C |
0.14 °C
0.04 % +0.10 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 100 Ω Pt (3926) |
-200 to 100
100 to 630 |
0.08 °C
0.02 % +0.06 °C |
0.16 °C
0.04 % +0.12 °C |
1 mA |
0.05 °C
0.0125 % + 0.04 °C |
0.10 °C
0.025 % + 0.08 °C |
0.1 mA to 10 mA |
| 10 Ω Cu (427) |
-100 to 260 |
0.2 °C |
0.4 °C |
3 mA |
0.2 °C |
0.4 °C |
0.1 mA to 10 mA |
| 120 Ω Ni (672) |
-80 to 260 |
0.1 °C |
0.2 °C |
1 mA |
0.04 °C |
0.08 °C |
0.1 mA to 10 mA |
1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
| Temperature, Thermocouples |
| Type |
Source °C |
Measure °C |
Source °C |
|
|
1 year |
2 years |
1 year |
2 years |
| E |
-250 to -200 |
1.3 |
2.0 |
0.6 |
0.9 |
|
-200 to -100 |
0.5 |
0.8 |
0.3 |
0.4 |
|
-100 to 600 |
0.3 |
0.4 |
0.3 |
0.4 |
|
600 to 1000 |
0.4 |
0.6 |
0.2 |
0.3 |
| N |
-200 to -100 |
1.0 |
1.5 |
0.6 |
0.9 |
|
-100 to 900 |
0.5 |
0.8 |
0.5 |
0.8 |
|
900 to 1300 |
0.6 |
0.9 |
0.3 |
0.4 |
| J |
-210 to -100 |
0.6 |
0.9 |
0.3 |
0.4 |
|
-100 to 800 |
0.3 |
0.4 |
0.2 |
0.3 |
|
800 to 1200 |
0.5 |
0.8 |
0.3 |
0.3 |
| K |
-200 to -100 |
0.7 |
1.0 |
0.4 |
0.6 |
|
-100 to 400 |
0.3 |
0.4 |
0.3 |
0.4 |
|
400 to 1200 |
0.5 |
0.8 |
0.3 |
0.4 |
|
1200 to 1372 |
0.7 |
1.0 |
0.3 |
0.4 |
| T |
-250 to -200 |
1.7 |
2.5 |
0.9 |
1.4 |
|
-200 to 0 |
0.6 |
0.9 |
0.4 |
0.6 |
|
0 to 400 |
0.3 |
0.4 |
0.3 |
0.4 |
| B |
600 to 800 |
1.3 |
2.0 |
1.0 |
1.5 |
|
800 to 1000 |
1.0 |
1.5 |
0.8 |
1.2 |
|
1000 to 1820 |
0.9 |
1.3 |
0.8 |
1.2 |
| R |
-20 to 0 |
2.3 |
2.8 |
1.2 |
1.8 |
|
0 to 100 |
1.5 |
2.2 |
1.1 |
1.7 |
|
100 to 1767 |
1.0 |
1.5 |
0.9 |
1.4 |
| S |
-20 to 0 |
2.3 |
2.8 |
1.2 |
1.8 |
|
0 to 200 |
1.5 |
2.1 |
1.1 |
1.7 |
|
200 to 1400 |
0.9 |
1.4 |
0.9 |
1.4 |
|
1400 to 1767 |
1.1 |
1.7 |
1.0 |
1.5 |
| C |
0 to 800 |
0.6 |
0.9 |
0.6 |
0.9 |
|
800 to 1200 |
0.8 |
1.2 |
0.7 |
1.0 |
|
1200 to 1800 |
1.1 |
1.6 |
0.9 |
1.4 |
|
1800 to 2316 |
2.0 |
3.0 |
1.3 |
2.0 |
| L |
-200 to -100 |
0.6 |
0.9 |
0.3 |
0.4 |
|
-100 to 800 |
0.3 |
0.4 |
0.2 |
0.3 |
|
800 to 900 |
0.5 |
0.8 |
0.2 |
0.3 |
| U |
-200 to 0 |
0.6 |
0.9 |
0.4 |
0.6 |
| 0 to 600 |
0.3 |
0.4 |
0.3 |
0.4 |
| BP |
0 to 1000 |
1.0 |
1.5 |
0.4 |
0.6 |
|
1000 to 2000 |
1.6 |
2.4 |
0.6 |
0.9 |
|
2000 to 2500 |
2.0 |
3.0 |
0.8 |
1.2 |
| XK |
-200 to 300 |
0.2 |
0.3 |
0.2 |
0.5 |
| 300 to 800 |
0.4 |
0.6 |
0.3 |
0.6 |
|
| Model Name | Description | Price List | Buy it | | Fluke 753 | Documenting Process Calibrator including:
- BC7240 battery charger
- Li-on BP7240 battery pack
- DPCTrack 2™ Sample Software
- Instruction manual
- NIST-traceable calibration report and data
- Three sets of TP220 test probes with three sets of “extended tooth alligator clips
- Two sets AC280 hook clips
- C799 Soft Field Case
- USB communication cable
|
|  | |
| CarePlans | Buy it | | G3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
| |  | | S3P-Group4 |
- SILVER CarePlan.
- Models covered: 434-II, 434-II/BASIC, 435-II, 435-II/BASIC, 700BCW, 724, 725, 726, and 753.
- Extends warranty from three to five years.
|  |  |
| Clips | Buy it | | AC220 | AC220 SureGrip™ Alligator Clips |  |  | | AC280 | AC280 SureGrip™ Hook Clips |  |  | | AC285 | AC285 SureGrip™ Alligator Clips |  |  |
| Test Probes | Buy it | | TP74 | TP74 Lantern Tip Test Probes |  |  | | TP920 | TP920 Test Probe Adapter Set |  |  |
| Differential Pressure Modules | Buy it | | 700P00 | 700P00 Differential Pressure Module |  | | | 700P01 | 700P01 Differential Pressure Module |  | | | 700P02 | 700P02 Differential Pressure Module |  | | | 700P03 | 700P03 Differential Pressure Module |  | | | 700P04 | 700P04 Differential Pressure Module |  | |
| Pressure Calibration Kit | Buy it | | 700PCK | Fluke 700PCK
Pressure Calibration Kit
|  | |
| Pumps and Accessories | Buy it | | 700PMP | 700 PMP Pressure Pump
|  |  | | 700HTH-1 | Fluke 700HTH-1 Hydraulic Test Hose |  |  | | 700PRV-1 | Fluke 700PRV-1 Pressure Relief Valve Kit |  |  | | 700PTP-1 | Fluke 700PTP-1 Pneumatic Test Pump |  |  | | 700HTP-2 | Fluke 700HTP-2 Hydraulic Test Pump |  |  |
| Differential Pressure Module Wet | Buy it | | 700P22 | 700P22 Differential Pressure Module Wet |  | | | 700P23 | 700P23 Differential Pressure Module Wet |  | | | 700P24 | 700P24 Differential Pressure Module Wet |  | |
| Absolute Pressure Modules | Buy it | | 700PA3 | 700PA3 Absolute Pressure Module |  | | | 700PA4 | 700PA4 Absolute Pressure Module |  | | | 700PA5 | 700PA5 Absolute Pressure Module |  | | | 700PA6 | 700PA6 Absolute Pressure Module |  | |
| Batteries, Chargers and Adapters | Buy it | | BC7240 | BC7240 Battery charger/Eliminator for 75x | | | | BP7240 | BP7240 Li-Ion Battery Pack for 75x |  | |
| Cases and Holsters | Buy it | | C700 | C700 Hard Carrying Case
|  |  | | C781 | C781 Meter Case |  |  | | C799 | Soft Field Case |  |  |
| Dual Pressure Modules | Buy it | | 700PD2 | 700PD2 Dual Pressure Module |  | | | 700PD3 | 700PD3 Dual Pressure Module |  | | | 700PD4 | 700PD4 Dual Pressure Module |  | | | 700PD5 | 700PD5 Dual Pressure Module |  | | | 700PD6 | 700PD6 Dual Pressure Module |  | | | 700PD7 | 700PD7 Dual Pressure Module |  | |
| Software | Buy it | | 750SW | Fluke 750 SW DPC/TRACK2 Software |  |  |
| Test Leads, Probes and Clips | Buy it | | 700TLK | Fluke 700TLK Process Test lead kit |  | |
| Vacuum Pressure Modules | Buy it | | 700PV3 | 700PV3 Vacuum Pressure Module |  | | | 700PV4 | 700PV4 Vacuum Pressure Module |  | |
| Temperature | Buy it | | 80CJ-M | 80CJ-M type J Male Mini-Connectors
|  |  | | 80CK-M | 80CK-M type K Male Mini-Connectors
|  |  | | 80PJ-EXT | 80PJ-EXT Extension Wire Kit |  |  | | 80PK-1 | 80PK-1 Bead Probe |  |  | | 80PK-22 | 80PK-22 SureGrip™ Immersion Temperature Probe |  |  | | 80PK-24 | 80PK-24 SureGrip™ Air Temperature Probe |  |  | | 80PK-25 | 80PK-25 SureGrip™ Piercing Temperature Probe |  |  | | 80PK-26 | 80PK-26 SureGrip™ Tapered Temperature Probe |  |  | | 80PK-27 | 80PK-27 SureGrip™ Industrial Surface Temperature Probe |  |  | | 80PK-8 | 80PK-8 Pipe Clamp Temperature Probe |  |  | | 80PK-9 | 80PK-9 General Purpose Probe |  |  | | 80PK-EXT | 80PK-EXT Extension Wire Kit |  |  | | 80PT-25 | 80PT-25 SureGrip Piercing Temperature Probe |  |  |
| Test Leads | Buy it | | TL940 | TL940 Mini-Hook Test Leads |  |  | | TL950 | TL950 Mini-Pincer Test Leads |  |  |
|
|
|
|
|
|
|
|
|
|
|
|
| | |