Keep your world up and running.
Home  |  Products  |  Process Calibration Tools  |  Pressure Calibrators  |  

Fluke 718Ex Intrinsically Safe Pressure Calibrator

 Fluke 718EX
Pump Up the Pressure

The new Fluke 718Ex Intrinsically Safe Pressure Calibrator is a new, powerful intrinsically safe, self contained pressure calibrator for usage in Explosion endangered areas.


The Fluke 718Ex offers:
  • ATEX II 1G Ex ia IIC T4 Compliant.
  • Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
  • 30, 100, and 300 psi ranges (2, 7, and 20 bar)
  • Pressure measurement to 0.05% of full span, using an internal pressure sensor.
  • Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
  • Wide range of selectable measurement units for pressure.
  • Current measurement with 0.02% accuracy and 0.001 mA resolution.
  • Pressure switch test function.
  • Min-Max Hold functions.
  • Compact size and weight.
  • Simple, push-button user interface.
  • Rugged and reliable, ideal for field use.
Measurement Accuracy
Pressure
Fluke 718Ex 30G  
Range:  -830 mbar to 2 bar (internal sensor)
 Over pressure: 3x FS
Resolution:  0.0001 bar
Accuracy:  0.05% FS
Functions:  Zero, Min, Max, Hold, Damp
 Note: for use with non-corrosive gasses
Fluke 718Ex 100G  
Range:  -830 mbar to 7 bar (internal sensor)
 Over pressure: 2x FS
Resolution:  0.001 bar
Accuracy:  0.05% FS
Functions:  Zero, Min, Max, Hold, Damp
 Note: for use with non-corrosive gasses
Fluke 718Ex 300G  
Range:  -830 mbar to 20 bar (internal sensor)
 Over pressure: 375 psi, 25 bar
Resolution:  0.0001 bar
Accuracy:  0.05% FS
Functions:  Zero, Min, Max, Hold, Damp
 Note: for use with non-corrosive gasses
Fluke 718Ex series  
 via pressure module connector
Range:  29 pressure modules, 2.5 mbar to 700 bar
 Over pressure per pressure module specs
Resolution:  Per pressure module specs
Accuracy:  To 0.025% of full span, per pressure module specs
Functions:  Zero, Min, Max, Hold, Damp
 Note: media compatibility per pressure module specs
Source Pressure, built in pump  
Range:  -820 mbar to full scale

Note:  Supported Pressure Units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C),
 cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg

Technical Data
Measure Accuracy:
Range:  0-24 mA
Resolution:  0.001 mA
Accuracy:  0.02% rdg ± 2 counts
Display
 LCD, 5 digit pressure and current simultaneous

Environmental Specifications
Operating Temperature
 -10ºC to 55ºC
Storage Temperature
 -40ºC to 60ºC
Humidity (Without Condensation)
 95% (10ºC to 30ºC)
 75% (30ºC to 40ºC)
 45% (40ºC to 50ºC)
 35% (50ºC to 55ºC)
Operating Altitude
 3,000 m max

Safety Specifications
Safety
 ATEX II 1G Ex ia IIC T4 compliant
 CSA Certification I.S. Class I,
 Div 1 Groups A-D , T4

Mechanical & General Specifications
Size
 216 mm L x 94 mm W x 66 mm H (with holster)
Weight
 Approx. 992 g (with holster)
Warranty
 Three years, one year for pump
Shock
 1 meter drop test
Vibration
 Random, 2 G, 5-500 Hz
Battery type
 One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
 Battery life: 4 to 20 hours, depending on functions used

Model NameDescription
Fluke 718EX 30G Pressure Calibrator 30 PSI, EX - I.S.

Each calibrator includes:

  • Protective Red Holster
  • TL75 Test Leads
  • AC72 Test Clips
  • CD Users Manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and data
Fluke 718EX 100GPressure Calibrator 100 PSI, EX - I.S
Fluke 718EX 300GPressure Calibrator 300 PSI, EX - I.S.

Pumps and Accessories
700HTH-1Fluke 700HTH-1 Hydraulic Test Hose

Absolute Pressure Modules
700PA4EX700PA4Ex Absolute Pressure Module

Test Leads
TL238TL238 SureGrip™ Test Lead Set for High Energy Environments
TL75TL75 Hard Point™ Test Lead Set
        
Home  |   Site Map  |   Fluke Corporation  |  Safety Notices
Privacy Statement  |  Terms of Use   |  Terms and Conditions of Sale  |  Terms and Conditions of Service
© 1995 - 2017 Fluke Corporation