Forget the Hassle of Hand Pumping
Now you can calibrate and test pressure devices quickly and easily with one hand, saving valuable time. With the innovative, built-in electric pump, the Fluke 719 Electric Pressure Calibrator provides pressure calibration at your fingertips!
Use the programmable pump limit settings to eliminate over-pressurization, and to set ‘pump to’ pressure values. The unique cleanout port design reduces pump failures, and makes the 719 easy to clean without repair. The compact, lightweight form, combined with one-handed calibration, makes this pressure calibrator the ideal tool for busy process professionals.
Pressure calibration made easy, by a name you know and trust for all your process calibration needs.
|
|
Key Features:
- Electric pump for one handed pressure calibration
- Best in class pressure measurement uncertainty of 0.025%
- Measure and source mA with best in class 0.015% accuracy
- Precision pressure adjust vernier for accurate pressure calibration
- Variable release rate bleed valve for easy pressure adjustment
- Ideal performance for high accuracy transmitter calibration
- 30 PSI and 100 PSI pressure ranges
- Programmable pressure calibrator pump limit:
- eliminates over-pressurization
- allows the user to program target pressures for pumping
- Extended pressure measurement range with any of 29 700Pxx Pressure Modules
- Source mA with simultaneous pressure measurement to test valves and I/Ps
- Simulate mA signals to troubleshoot 4-20 mA loops
- Test pressure switches with built in switch test function
- Power transmitters during test using 24 V loop supply with simultaneous mA measurement
- Pressure Calibrator includes proven cleanout ports reduce pump failures
|
|
|
| General Specifications |
| Storage temperature | | | Operating temperature | | | Relative humidity |
|
95 % (-10 °C to 30 °C) |
|
75 % (30 °C to 40 °C) |
|
45 % (40 °C to 50 °C) |
|
35 % (50 °C to 55 °C) |
| | Shock | | | Safety |
|
CSA C22.2 No. 1010.1:1992 |
| | EMC |
|
EN50082-1:1992 and EN55022:1994 Class B |
| | Size (HxWxL) |
|
60 mm x 87 mm x 210 mm |
|
(2.19 in x 3.41 in x 8.28 in) |
| | Weight | | | Battery |
|
Two 9 V alkaline batteries |
| | Battery life |
|
12 hours with 12 mA into 500 Ω, 200
pump cycles to 100 psi, 500 pump cycles to 30 psi |
| | Warranty |
|
Three-years, one year for pump |
|
| Specification: |
| Fluke-719 30G |
| Range |
-12 to 36 psi,
-850 mbar to 2.4 bar |
| Resolution |
0.001 psi, 0.1 mbar |
| Accuracy |
0.025 % range for 6
months |
|
0.035 % of range for
one (1) year |
| | Fluke-719 100G |
| Range |
-12 to 120 psi,
-850 mbar to 8 bar |
| Resolution |
0.01 psi, 1 mbar |
| Accuracy |
0.025 % range for 6
months |
|
0.035 % of range for
one (1) year |
| | mA |
| Range |
0 to 24 mA |
| Resolution |
0.001 mA |
| Accuracy |
.015 % Rdg + 2 counts |
| Notes |
Max load, 1000 Ω,
750 Ω ohm in HART mode |
| | Loop supply |
| Range |
24 V dc |
| Resolution |
N/A |
| Accuracy |
± 10 % |
|
| Pump specifications |
| Maximum pressure generation |
|
120 PSI (8 BAR) (100 PSI model) |
| | Pump to |
|
- 30 PSI in less than 15 seconds* |
|
- 100 psi in less than 45 seconds* |
|
*(with 1 meter of 1/8 inch test hose) |
| | Pump cycles between rebuilds | | | Total pump life | |
|
| Model Name | Description | Price List | | Fluke 719 30G |
Electric Pressure Calibrator, 30 psi, 2 bar
All Models Include:
- Protective Yellow Holster
- Instructions Sheet (15 languages)
- Test Leads and Alligator Clip
- Two 9V alkaline batteries
- Hose Kit
- NIST - Traceable calibration report and data
|
| | Fluke 719 100G |
Electric Pressure Calibrator, 100 psi, 7 bar
|
| Differential Pressure Modules | | 700P00 | 700P00 Differential Pressure Module (2.5 mbar) | | 700P01 | 700P01 Differential Pressure Module | | 700P02 | 700P02 Differential Pressure Module (70 mbar) | | 700P03 | 700P03 Differential Pressure Module (340 mbar) | | 700P04 | 700P04 Differential Pressure Module (1000 mbar) |
| Differential Pressure Module Wet | | 700P22 | 700P22 Differential Pressure Module Wet (70 mbar) | | 700P23 | 700P23 Differential Pressure Module Wet (340 mbar) | | 700P24 | 700P24 Differential Pressure Module Wet |
| Absolute Pressure Modules | | 700PA3 | 700PA3 Absolute Pressure Module (340 mbar) | | 700PA4 | 700PA4 Absolute Pressure Module | | 700PA5 | 700PA5 Absolute Pressure Module (2 bar) | | 700PA6 | 700PA6 Absolute Pressure Module (7 bar) |
| Dual Pressure Modules | | 700PD2 | 700PD2 Dual Pressure Module (± 70 mbar) | | 700PD3 | 700PD3 Dual Pressure Module (± 340 mbar) | | 700PD4 | 700PD4 Dual Pressure Module (±1000 mbar) | | 700PD5 | 700PD5 Dual Pressure Module (-1000/+2000 mbar) | | 700PD6 | 700PD6 Dual Pressure Module (-1000 mbar/+6.9 bar) | | 700PD7 | 700PD7 Dual Pressure Module ( -1000 mbar/+13.8 bar) |
| Vacuum Pressure Modules | | 700PV3 | 700PV3 Vacuum Pressure Module (-340 mbar) | | 700PV4 | 700PV4 Vacuum Pressure Module( -1000 mbar) |
|
|
|
|
|
|
|
|
|