Pump Up the Pressure
The new Fluke 718Ex Intrinsically Safe Pressure Calibrator is a new, powerful intrinsically safe, self contained pressure calibrator for usage in Explosion endangered areas.
|
The Fluke 718Ex offers:
- ATEX II
1G EEx ia IIC T4 Compliant.
- Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
- 30, 100, and 300 psi ranges (2, 7, and 20 bar)
- Pressure measurement to 0.05% of full span, using an internal pressure sensor.
- Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
- Wide range of selectable measurement units for pressure.
- Current measurement with 0.02% accuracy and 0.001 mA resolution.
- Pressure switch test function.
- Min-Max Hold functions.
- Compact size and weight.
- Simple, push-button user interface.
- Rugged and reliable, ideal for field use.
|
|
| Measurement Accuracy |
| Pressure |
| Fluke 718Ex 30G |
|
| Range: |
-830 mbar to 2 bar (internal sensor) |
|
Over pressure: 3x FS |
| Resolution: |
0.0001 bar |
| Accuracy: |
0.05% FS |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
| Fluke 718Ex 100G |
|
| Range: |
-830 mbar to 7 bar (internal sensor) |
|
Over pressure: 2x FS |
| Resolution: |
0.001 bar |
| Accuracy: |
0.05% FS |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
| Fluke 718Ex 300G |
|
| Range: |
-830 mbar to 20 bar (internal sensor) |
|
Over pressure: 375 psi, 25 bar |
| Resolution: |
0.0001 bar |
| Accuracy: |
0.05% FS |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
| Fluke 718Ex series |
|
|
via pressure module connector |
| Range: |
29 pressure modules, 2.5 mbar to 700 bar |
|
Over pressure per pressure module specs |
| Resolution: |
Per pressure module specs |
| Accuracy: |
To 0.025% of full span, per pressure module specs |
| Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: media compatibility per pressure module specs |
| Source Pressure, built in pump |
|
| Range: |
-820 mbar to full scale |
| Note: |
Supported Pressure Units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C), |
|
cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg |
|
| Technical Data |
| Measure Accuracy: |
| Range: |
0-24 mA
|
| Resolution: |
0.001 mA |
| Accuracy: |
0.02% rdg ± 2 counts |
| | Display |
|
LCD, 5 digit pressure and current simultaneous |
|
| Environmental Specifications |
| Operating Temperature | | | Storage Temperature | | | Humidity (Without Condensation) |
|
95% (10ºC to 30ºC) |
|
75% (30ºC to 40ºC) |
|
45% (40ºC to 50ºC) |
|
35% (50ºC to 55ºC) |
| | Operating Altitude | |
| Safety Specifications |
| Safety |
|
ATEX II 1G EEx ia IIC T4 compliant |
|
CSA Certification I.S. Class I, |
|
Div 1 Groups A-D , T4 |
|
| Mechanical & General Specifications |
| Size |
|
216 mm L x 94 mm W x 66 mm H (with holster) |
| | Weight |
|
Approx. 992 g (with holster) |
| | Warranty |
|
Three years, one year for pump |
| | Shock | | | Vibration | | | Battery type |
|
One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V |
|
Battery life: 4 to 20 hours, depending on functions used |
|
|
| Model Name | Description | Price List | | Fluke 718EX 30G |
Pressure Calibrator 30 PSI, EX - I.S.
Each calibrator includes:
- Protective Red Holster
- TL75 Test Leads
- AC72 Test Clips
- CD Users Manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish,
Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
- Fluke 718Ex CCD content control drawing
- NIST- traceable calibration report and data
|
| | Fluke 718EX 100G |
Pressure Calibrator 100 PSI, EX - I.S
| | Fluke 718EX 300G |
Pressure Calibrator 300 PSI, EX - I.S.
|
| Absolute Pressure Modules | | 700PA4EX | 700PA4Ex Absolute Pressure Module |
| Test Leads | | TL238 | TL238 SureGrip™ Test Lead Set for High Energy Environments | | TL75 | TL75 Hard Point Test Lead Set |
|
|
|
|
|
|
|
|