5627A Precision Industrial RTDs

Durable PRTs with temperature range to 420 °C and accuracy to 0.025 °C

  • Vibration and shock resistant
  • 19 mm (3/4-inch) bend radius for increased durability
  • NVLAP-accredited calibration included, lab code 200706-0

When buying a PRT, performance isn’t the only criterion you need to look at. The real issues are price-to-accuracy and price-to-durability ratios.

5627A probes have a temperature range up to 420 °C and an accuracy as good as ± 0.025 °C. They come in three different lengths. (Both six- and nine-inch models cover –200 °C to 300 °C.) Each instrument is shipped with its ITS-90 coefficients and a calibration table in 1 °C increments.

One of the best features of this sensor is that it conforms to the standard 385 curve, letting you use your DIN/IEC RTD meters fully. Why use a probe that’s less accurate than your meter?

The 5627A is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. (Bend, if any, should be specified at time of order.)

Six- and nine-inch 5627As are calibrated at –196 °C, –38 °C, 0 °C, 200 °C, and 300 °C. For 12-inch versions the point at 300 °C is replaced by a calibration point at 420 °C.

Each probe is individually calibrated and includes a NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.

This probe is an excellent value. It has the price-to-accuracy and price-to-durability ratios you should demand in every PRT you buy!

Specifications

Resistance 

Nominal 100W 

Temperature Coefficient 

0.00385 W/W/ °C nominal 

Temperature Range 

–200 °C to 420 °C
(5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C) 

Drift Rate (k=2) 

± 0.04 °C at 0 °C after 100 hours at 420 °C 

Sheath Material 

316 Stainless Steel 

Leads 

Teflon™-insulated, nickel-plated stranded copper, 22 AWG 

Termination 

Specify. See Ordering Information. 

Time Constant 

Four seconds maximum for 63.2 % response to step change in water moving at 3 fps. 

Bending Radius 

Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Hart lab requires 20 cm [8 in] of unbent sheath to re-calibrate.) 

Calibration 

Includes manufacturer’s NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from –196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included. 

Immersion 

At least 100 mm (4 in) recommended 

Calibrated Accuracy (k=2) 

± 0.026 °C at –196 °C
± 0.046 °C at 0 °C  
± 0.077 °C at 200 °C  
± 0.124 °C at 420 °C  

Size 

5627A-12: 305 mm x 6.35 mm (12 in x 1/4 in)
5627A-9:  229 mm x 4.7 mm (9 in x 3/16 in)
5627A-6: 152 mm x 4.7 mm (6 in x 3/16 in) 

Includes calibration uncertainty and 100 hr drift. 