ScopeMeter®  Firmware Upgrade


ScopeMeter® 190C Series Version 6 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 190C/2x5 Series is available, major changes are:
  • Several issues on CAN, FF and Profibus DP have been resolved (2x5C only)
  • F2 is disabled in eye-pattern button bar when bus health is off (2x5C only)
  • Fixed Dot-join status after leaving bushealth mode (2x5C only)
  • Fixed Trigger status indication after leaving bushealth mode (2x5C only)
  • No data banner no longer appears when wiring diagram is shown (2x5C only)
  • Fixed recall function, not correctly recalling screen + data (V8.04)
  • Fixed TrendPlot Save. UI only showed Screen + setting not Recording (V8.04)
  • Fixed Trend storage (V8.02)
  • Remote Control Fluke 190 Series improved (V7.06)
  • Zoom in 2s/div timebase setting improved and micro symbol added in readings symbol table (V7.04)
  • Improved screen update rate (V7.02)
  • Deeper waveform memory up to 3K spectrum analysis using Fast Fourier Transformation (V7.00)
  • Two new trigger modes (n-cycle and dual edge) (V7.00)
  • Cursor limited automatic measurement (rms Watt etc between cursors) (V7.00)

Upgrading Version V7.0x, V8.00 or V8.02 to Version V8.04 - Instruments with version V7.01 to V8.02 are field upgradable free of charge.
v8.04 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 190C/2x5C Series Color ScopeMeter, with firmware V7.0 1V 8 .0 2 . (See Firmware Level Check above);
  • A PC, running Windows XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 19xC/S model and with the SCC-190 kit) for data interconnection between the PC and the instrument

Upgrading earlier Versions - Instruments with firmware version V5.0x and V6.x cannot be upgraded to V8.04 in the field. These instruments require re-calibration after upgrading. Therefore these units must be returned to a Fluke Authorized Service Center. Contact your local service center and order a firmware upgrade and calibration according to PCN 4841. Note: For the firmware upgrade and calibration a nominal fee will be charged!

ScopeMeter® 190B Series Version 7 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 190B Series is available, major changes are:

  • Remote Control improved (V7.06)
  • Zoom in 2s/div timebase setting improved and micro symbol added in readings symbol table (V7.04)
  • Improved screen update rate (V7.02)
  • Deeper waveform memory up to 3K spectrum analysis using Fast Fourier Transformation (V7.00)
  • Two new trigger modes (n-cycle and dual edge) (V7.00)
  • Cursor limited automatic measurement (rms Watt etc between cursors) (V7.00)

Upgrading Version V7.00, V7.01, V7.02 or V7.04 to Version V7.06 - Instruments with version V7.00 to V7.04 are field upgradable free of charge.
v7.06 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 190 B Series ScopeMeter or 190B Series (Monochrome) ScopeMeter, with firmware V7.00 - V7.04. (See Firmware Level Check above);
  • A PC, running Windows 95, 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 19x B /S model and with the SCC-190 kit) for data interconnection between the PC and the instrument

Upgrading earlier Versions - Instruments with firmware version V5.0x and V6.x cannot be upgraded to V7.06 in the field. These instruments require re-calibration after upgrading. Therefore these units must be returned to a Fluke Authorized Service Center. Contact your local service center and order a firmware upgrade and calibration according to PCN 4841. Note: For the firmware upgrade and calibration a nominal fee will be charged!

ScopeMeter® 190C Series Version 6 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 190C Series is available, major changes are:

  • Improved TrendPlot scaling
  • Improved Voltage Readings stability
  • When performing a TrendPlot of an Input A measurement, the probe attenuation factor of Input B was shown.
  • If a meter TrendPlot is acquired in FlukeView 4.0, it can show a scaling error. This happens if a previously acquired TrendPlot had different range or probe attenuator settings.

Upgrading Version V5.03, V5.04 and V6.xx to Version V6.15 - Instruments with firmware version V5.03 – V6.14 are field upgradable free of charge.
v6.15 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 190C Series Color ScopeMeter, with firmware version V5.03 to V6.14. (See Firmware Level Check below);
  • A PC, running Windows 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 19xC/S model and with the SCC-190 kit) for data interconnection between the PC and the instrument

Firmware Level Check - Press USER and F3 [VERSION & CAL …] on your ScopeMeter to check its firmware version.

ScopeMeter® 190B Series 6 Version Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 190 B Series is available, major changes are:
  • Improved TrendPlot scaling
  • Improved Voltage Readings stability
  • When performing a TrendPlot of an Input A measurement, the probe attenuation factor of Input B was shown.
  • If a meter TrendPlot is acquired in FlukeView 4.0, it can show a scaling error. This happens if a previously acquired TrendPlot had different range or probe attenuator settings.

Upgrading Version V6.00, V6.10, V6.11 and V6.12 to V6.13 - Instruments with firmware version V6.00 – V6.12 are field upgradable free of charge. 
v6.13 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 190 B Series ScopeMeter, with firmware version V 6.00 to V6.12 (See Firmware Level Check below);
  • A PC, running Windows 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 19x/S model and with the SCC-190 kit) for data interconnection between the PC and the instrument

Firmware Level Check - Press USER and F3 [VERSION & CAL …] on your ScopeMeter to check its firmware version.

ScopeMeter ® 125 Series Version 3.04 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 125 Series is available,
  • Measurement of total power (W) in Harmonics corrected (V3.03)
  • Foundation Fieldbus limits H1 according IEC 61158 (V3.02)
  • Improved task scheduler for Bushealth (V3.01)

Upgrading Version V3.00, V3.01,V3.02 or V3.03 to Version V3.04 - Instruments with firmware version V3.00, V3.01,V3.02 or V3.03 are field upgradeable free of charge.
v3.04 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 125 Series ScopeMeter, with firmware version V3.00, V3.01 or V3.02 (See Firmware Level Check below);
  • A PC, running Windows 95, 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 125/S model and with the SCC-120 kit) for data interconnection between the PC and the instrument

Firmware Level Check - Press USER OPTIONS and F3 [VERSION & CAL …] on your ScopeMeter to check its firmware version.

ScopeMeter ® 123/124 Series Version 2 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 123/124  is available, which gives you the following additional functions:
  • Battery refresh time changed from 12 to 19 hours to reflect the real situation (V2.02)
  • Postscript printing improved. (V2.01)
  • A screen refresh is done after a 'send setup' action (V2.01)

Upgrading Version V2.00, V2.01 or V2.02  to Version V2.03 - Instruments with firmware version V2.0x  are field upgradable free of charge.
v2.03 FlashTool Upgrade »

  • Fluke 123 Series ScopeMeter, with firmware version V2.00 to V2.02 (See Firmware Level Check below);
  • A PC, running Windows 95, 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 12x/S model and with the SCC-120 kit) for data interconnection between the PC and the instrument

Firmware Level Check - Press USER OPTIONS and F3 [VERSION & CAL …] on your ScopeMeter to check its firmware version.

Upgrading earlier Versions - Instruments with firmware version V1.0x cannot be upgraded to V2.03

ScopeMeter ® 123 Series Version 1 Firmware Upgrade

A firmware upgrade for ScopeMeter test tool 120 Series is available, which gives you the following additional functions:
  • The new firmware provides probe settings 20:1 and 200:1 for the DP120 differential probe. The 1000:1 probe setting has been canceled.
  • In the previous firmware the print function only supports A4 paper format
  • Probe settings 1000:1, 10 mV/mA, and 0.1 mV/A have been added.
  • When a temperature or diode measurement is trendplotted, and the Fluke123 is switched off and on again, the vertical scaling information is shown.
  • The trigger icon in VIDEO mode is automatically set to negative edge ( ) at VIDEO POLARITY POSITIVE, and to positive edge ( ) at VIDEO POLARITY NEGATIVE. This reflects the actual trigger edge.
  • The following bug has been solved:
    Set Input B OFF, select TRENDPLOT, press HOLD, press V Hz A (INPUT B MEASUREMENTS menu), press right arrow key (MEASURE on B), press F4 (ENTER), press F4 (YES, restart TRENDPLOT); TRENDPLOT is restarted now, but the display still shows HOLD in V01.03 and lower.

Upgrading Version V1.00, V1.02, V1.03 or V1.04  to Version V1.08 - Instruments with firmware version V1.0x  are field upgradable free of charge.
v1.08 FlashTool Upgrade »

FlashTool Software requirements:

  • Fluke 123 Series ScopeMeter, with firmware version V1.00, V1.02, V1.03 or V1.04 (See Firmware Level Check below);
  • A PC, running Windows 95, 98, Me, 2000, NT 4.0, XP or Vista, with a free RS-232 or USB port to perform the software installation;
  • PM9080 or OC4USB (RevII) optically isolated serial interface cable (as included with the ScopeMeter 123/S model and with the SCC-120 kit) for data interconnection between the PC and the instrument

Firmware Level Check - Press USER OPTIONS and F3 [VERSION & CAL …] on your ScopeMeter to check its firmware version.