The i3000s Flex is an ac current clamp utilizing the Rogowski principle. It can be used to measure currents up to 3000 A when used in conjunction with oscilloscopes, recorders or data loggers. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas.
- Three ranges: 30 A, 300 A & 3000 A, 1, 10 or 100 mV/A output
- Ideal for current measurement on thick and hard to reach conductors without breaking the circuit
- Maximum conductor diameter 178 mm (7 in.)
- CAT III 600V safety rating
|
|
|
| Electrical specifications |
| Current ranges | | | Output sensitivity (AC-coupled) | | | Load impedance | | | Accuracy |
| at +25 °C |
± 1 % of range (45 Hz to 65 Hz) |
| | Linearity |
| 10 % to 100 % of range |
± 0.2 % of reading |
| | Noise |
|
8 mV ac rms (30 A) |
|
2 mV ac rms (300 A, 3000 A) |
| | Frequency range | | | Phase error |
|
< ± 1 º (45 Hz to 65 Hz) |
|
± 10 º (at 20 kHz) |
| | Position sensitivity |
| with cable > 25 mm from the coupling |
± 2 % of range |
| | External field |
| with cable > 200 mm from the head |
± 1 % of range |
| | Power supply |
|
2 x AA/MN 1500/LR6 Alkaline, , 400 hours, low battery indicator or dedicated external power supply |
| | Temperature Coefficient |
|
± 0.08% of reading per °C |
| | Working voltage |
| see Safety standards section |
600 V AC rms or DC |
|
| General Specifications |
| Head Circumference | | | Head cable diameter | | | Head bend radius | | | Cable length (head to electronics) | | | Output connection |
|
0.5 m cable terminated with safety BNC connector. Supplied with 4mm safety plug adapter |
| | Operating temperature range |
|
-20 °C to +90 °C (-4 °F to 190 °F) (head) |
|
-20 °C to +85 °C (-4 °F to 185 °F) (electronics) |
| | Storage temperature range |
|
-40 °C to +105 °C (-40 °F to 221 °F) (head) |
|
-20 °C to +85 °C (-4 °F to
185 °F) (electronics)
|
| | Operating humidity |
|
15 % to 85 % (non-condensing) |
| | Weight |
|
180 g (0.396 lb) (head) |
|
190 g (0.419 lb) (electronics) |
| | Safety standards |
|
BS EN 61010-1: 2001 |
|
BS EN 61010-2-032: 2002 |
|
BS EN 61010-031: 2002 |
|
600 Vrms, Category III, Pollution Degree 2
|
|
Use of the probe on uninsulated conductors is limited to 600 V AC rms or dc and frequencies below 1 kHz. |
| | EMC standards |
|
EN 61326: 1998 +A1, A2, & A3 |
| | Maximum conductor size |
|
610 mm or 914 mm, depending on model |
|
|
| Model Name | Description | Price List | | i3000 Flex-4PK | Fluke i3000 Flex AC Current Clamp, 4 Pack | On Request |
|
|
|
|