The most trusted tools in the world
Home  |  Products  |  Process Calibration Tools  |  Multifunction Calibrators  |  

Fluke 754 Documenting Process Calibrator-HART

Fall in love. Again.

The Fluke 754 has features we know you’ll love: increased accuracy, enhanced reliability, and is easier to read in dark environments.
Contact Me
Request A Demo
Fluke 754
Fluke 754
Fluke 754
Fluke 754
Fluke 754
You’re going to love what we’ve done with your calibrator.

The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-on battery for longer life, USB port and new accessories to complete the package.

For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.

Supports popular instrumentation management software. The 753 and 754, 743 and 744 work with the Fluke DPC/TRACK2™ software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.

The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.

 

The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
Fluke 754

Work smarter. Work faster.

The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART® interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.

  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
  • Holds up to a full week of downloaded procedures and calibration results.
  • Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
  • Easy to use
  • Three year warranty
  • Bright white dual display. Read both sourced and measured parameters simultaneously.
  • Multi lingual interface
  • Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge.
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
  • Delivered with DPC/Track Sample software.
  • Compatible with many Asset Management software packages.
HART Capabilities
The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.
  • Requires no external box or second tool for everyday HART calibration and maintenance.
  • Offers fast HART communication.
  • Supports popular models of HART transmitters, with more devise-specific command support than any other HART field calibrator.
  • Works with multiple masters, burst mode, and multi-drop configurations.
  • Is easy to update as additional instruments are added and new HART versions are released.
  • Interrogate to determine device type, manufacturer, model, tag.
  • Reconfigure the sensor mapping of dual sensor temperature transmitters.
  • Read HART PV function and smart transmitter digital output while measuring analog mA output.
  • Read and write HART configuration functions to make field adjustments to PV range points, damping, and other top-level configuration settings.
  • Re-label smart transmitters by reading and writing to the HART tag field.


Versatile HART Protocol support

Manufacturer Pressure instruments Temperature instruments Coriolis instruments
ABB/Kent-Taylor 600T 658T1  
ABB/Hartmann & Braun Contrans P1,
AS 800 Series
   
Endress & Hauser CERABAR S,
CERABAR M,
DELTABAR S
TMT 1221,TMT 1821,
TMT 1621
 
Foxboro Eckardt   TI/RTT201  
Foxboro/Invensys I/A Pressure    
Fuji FCX
FCXAZ
FRC  
Honeywell ST3000 STT25T1,
STT25H1
 
Micro Motion     2000
2000 IS
9701
9712
9739
Moore Products   3441  
Rosemount 1151
2088
3001C
3051, 3051S
3044C
644
3144
3244, 3144P
 
Siemens SITRANS P DS
SITRANS P ES
   
SMAR LD301 TT3011  
Viatran I/A Pressure    
Wika UNITRANS T32H1  
Yokogawa EJA YTA 110, 310
and 320
 
1Sensor Trim not supported

The 754 supports the commands contained in HART protocol Version 5.7. With 2 MB of memory, the 754 supports a substantial set of HART instructions:

  • Universal commands provide functions that are implemented in all field devices, for example, read manufacturer and device type, read primary variable (PV), or read current output and percent of span.
  • Common practice commands provide functions that are common to many but not all field devices, for example read multiple variables, set damping time, or perform loop test.
  • Device-specific commands provide functions that are unique to a particular field device, for example sensor trim.


HART Operating Modes Supported
  • Point to Point operation , the most commonly used mode, connects the 754 to a single HART device in a 4-20 mA loop.
  • In Multi-Drop mode , several HART instruments can be bussed together. The 754 searches for each, identifies addresses in use, and allows you to select the instrument for calibration and related operations.
  • In Burst Mode, the HART instrument transmits bursts of data without waiting to be interrogated by a master unit. The 754 can take transmitters out of burst mode during test or calibration, then later restore them to burst mode.


Pressure Modules

Covers virtually any pressure application including gauge, differential, dual (compound), absolute, and vacuum.

  • Display pressure readings in any of 10 different pressure units you specify in the calibrator setup.
  • Rugged urethane molded cases protect the modules from rough handling and harsh conditions.
  • Features internal temperature compensation from 0º to 50º C for full accuracy performance.
  • Includes NIST-traceable calibration certificate.
  • Modules can be calibrated locally, helping to control costs.
A family of 29 optional pressure modules provides pressure calibration and measurement capabilities. Twenty-eight modules are available, with basic accuracy specs to 0.05%. Ranges start at 0 — 1 in H20 (0 — .25kPa) and go to 0 — 10,000 psi (0 — 70,000 kPa). Additional information about pressure modules is available on the Pressure Modules Home Page.

Automated Procedures

Allow you to quickly set up powerful, automated calibration procedures for linear transmitters, DP flow transmitters, and one- and two-point limit switches. Simply select the appropriate measure and/or source functions and fill out the procedure template. The 750 Series does the rest. It quickly performs the test, calculates the errors, and displays the final results, highlighting out of tolerance points.

Custom Units

Enable you to map one unit to another, such as mV to ºC or ºF. Allows you to use the Fluke 750 Series with millivolt output accessories such as the Fluke 80T-IR Temperature Probe, and to document tests using non-supported units such as parts per million or revolutions per minute.

User-entered Values

Enables technicians to record calibration results that were sourced and/or measured by other devices such as panel meters or readout-only devices.

Limit Switch Calibration

Procedures perform fast, automated calibration of one and two-point limit switches for voltage, current, temperature, and pressure.

Differential Pressure Flow Instrument Calibration

Routines use a square root function to directly calibrate DP flow instruments.

Additional Features
Multifunctional

Calibrate temperature, pressure, voltage, current, resistance, and frequency. Since it both measures and sources, you can troubleshoot and calibrate all with one rugged tool.

Powerful, yet easy to use

The easy-to-follow menu-driven display guides you through any task. Get up to speed in minutes, not days. Programmable calibration routines enable you to create and run automated as-found/as-left procedures to ensure fast, consistent, calibrations.

Records and documents results

To support your ISO-9000 or regulatory standards the Fluke 754 captures your calibration results, eliminating the need to juggle a pen and pad in the field. The USB interface lets you transfer the results to a PC, thus saving the time of having to manually transcribe them when you return to the shop.

Truly hand-held

Small enough to fit easily into a tool bag and to use in tight spaces. Runs an entire shift on a rechargeable Li-ion battery pack.

Rugged and reliable

Count on Fluke's rugged design to deliver top accuracy and reliability in harsh environments. Overmolded urethane case stands up to rough handling in industrial environments.

Bright white display

Lets you read your results in any kind of light. Backlight has three (3) settings.

Soft keys

Provide one-touch access to enhanced functions such as task lists, automated procedures, scaling, min/max, stepping and ramping, and review memory.

Three operating modes

Measure, Source, or simultaneous Measure/Source, - enable technicians to troubleshoot, calibrate, or maintain instrumentation with just one tool.

Multi-lingual interface

Displays instructions in English, French, German, Spanish, and Italian.

Built-in algebraic calculator

With four functions-plus square root-stores, recalls, and performs calculations required for setting up instruments or evaluating data in the field. Use it to set the source function to a calculated value. There's no need to carry a pencil and paper or a separate calculator.

Programmable measurement delay

Inside automated procedures permits calibrating instruments that respond slowly.

 

Measurement Accuracy
Voltage DC
 1 Year  2 Years
100.000 mV        0.02%+0.005 mV  0.03%+0.005 mV
3.00000 V  0.02%+0.00005 V  0.03%+0.00005 V
30.0000 V  0.02%+0.0005 V  0.03%+0.0005 V
300.00 V  0.05%+0.05 V  0.07%+0.05 V
Voltage AC
Range
40 to 500 Hz     
 Resolution  1 Year 2 Years 
3.000 V  0.001 V  0.5%+0.002 V  1.0%+0.004 V
30.00 V  0.01 V  0.5%+0.02 V  1.0%+0.04 V
300.0 V  0.1 V  0.5%+0.2 V  1.0%+0.2 V
Current DC
   1 Year  2 Years
30.000 mA          0.01% + 5 uA   0.015%+7 uA
110.00 mA  0.01% + 20 uA  0.015%+30 uA
Resistance
   1 Year  2 Years
10.000 Ω          
 0.05% + 50 mΩ  0.07%+70 mΩ
100.00 Ω  0.05% + 50 mΩ  0.07%+70 mΩ
1.0000 kΩ   0.05% + 500 mΩ  0.07%+0.5 Ω
10.000 kΩ  0.1% + 10 Ω  0.15%+15 Ω
Frequency
   Resolution  Accuracy (2 Years)
1.00 to 110.00 Hz  0.01 Hz  0.05 Hz
110.1 to 1100.0 Hz  0.1 Hz  0.5 Hz
1.101 to 11.000 kHz   0.001 kHz  0.005 kHz
11.01 to 50.00 kHz  0.01 kHz  0.05 kHz

 

Source Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV         
 0.01%+0.005 mV  0.015%+0.005 mV
1.00000 V  0.01%+0.00005 V  0.015%+0.0005 V
15.0000 V  0.01%+0.0005 V  0.015%+0.0005 V
Current DC
   1 Year 2 Years 
22.000 mA (Source)  0.01%+ 0.003 mA  0.02%+ 0.003 mA
Current Sink (Simulate)  0.02% + 0.007 mA  0.04% + 0.007 mA
Resistance
   1 Year  2 Years
10.000 Ω             
 0.01% + 10 mΩ  0.015% + 15 mΩ
100.00 Ω  0.01% + 20 mΩ  0.015% + 30 mΩ
1.0000 kΩ  0.02% + 0.2 Ω  0.03% + 0.3 Ω
10.000 kΩ  0.02% + 3 Ω  0.03% + 5 Ω
Frequency
   2 Years
0.1 to 10.99 Hz  0.01 Hz
0.01 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1.100 to 21.999 kHz  0.002 kHz
22.000 to 50.000 kHz        
 0.005 kHz

Technical Data
Data Log Functions
Measure functions:  Voltage, current, resistance, frequency, temperature, pressure
Reading rate:  1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length:  8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Rate:  4 steps/second
Trip detect:  Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage:  Selectable, 26 V
Accuracy:  10%, 18 V minimum at 22 mA
Maximum current:  25 mA, short circuit protected
Maximum input voltage:  50 V DC
Step Functions
Source Functions  Voltage, current, resistance, frequency, temperature
Manual Step  Selectable step, change with arrow buttons
Autostep  Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
 -10 °C to +50 °C
Storage Temperature
 -20 °C to +60 °C
Dust/water resistance
 Meets IP52, IEC 529
Operating Altitude
 3000 m above mean sea level (9842 ft)

 

Safety Specifications
Agency Approvals
 CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
 136 x 245 x 63 mm (5.4“ x 9.6 x 2.5)
Weight
 1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion:  7.2V, 4400mAh, 30 Wh
Battery Life
 >8 hours typical
Battery Replacement
 Replace without opening calibrator; no tools required
Side port connections
 Pressure module connector
 USB Connector to interface to your PC
   Digital instrument (HART) connector
 Connection for optional battery charger/eliminator
Data storage capacity
 1 week of calibration procedures results
90 day specifications
 The standard specification interval for the 750 Series are 1 and 2 years.
 Typical 90 day measurement and source accuracy can be estimated by
 dividing the one year "% of reading" or "%of output" specifications by 2.
 Floor specifications, expressed as "% of full scale" or
 "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors
Degrees or % of reading
Type (α) Range °C Measure °C1 Source current Source °C Allowable current2
  1 year 2 year   1 year 2 year  
100 Ω Pt (385) -200 to 100
100 to 800
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
200 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
500 μA 0.06 °C
0.017 % + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
500 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
250 μA 0.06 °C
0.017% + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
1000 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04% + 0.10 °C
150 μA 0.06 C
0.017 % + 0.05 °C
0.12 C
0.034 % + 0.10 °C
0.1 mA to 1 mA
100 Ω Pt (3916) -200 to 100
100 to 630
0.07 °C
0.02 % +0.05 °C
0.14 °C
0.04 % +0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
100 Ω Pt (3926) -200 to 100
100 to 630
0.08 °C
0.02 % +0.06 °C
0.16 °C
0.04 % +0.12 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
10 Ω Cu (427) -100 to 260 0.2 °C 0.4 °C 3 mA 0.2 °C 0.4 °C 0.1 mA to 10 mA
120 Ω Ni (672) -80 to 260 0.1 °C 0.2 °C 1 mA 0.04 °C 0.08 °C 0.1 mA to 10 mA

1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms

Temperature, Thermocouples
Type Source °C Measure °C Source °C
1 year 2 years 1 year 2 years
E -250 to -200 1.3 2.0 0.6 0.9
-200 to -100 0.5 0.8 0.3 0.4
-100 to 600 0.3 0.4 0.3 0.4
600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
-100 to 900 0.5 0.8 0.5 0.8
900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1.0 0.4 0.6
-100 to 400 0.3 0.4 0.3 0.4
400 to 1200 0.5 0.8 0.3 0.4
1200 to 1372 0.7 1.0 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
-200 to 0 0.6 0.9 0.4 0.6
0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2.0 1.0 1.5
800 to 1000 1.0 1.5 0.8 1.2
1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
100 to 1767 1.0 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
200 to 1400 0.9 1.4 0.9 1.4
1400 to 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
800 to 1200 0.8 1.2 0.7 1.0
1200 to 1800 1.1 1.6 0.9 1.4
1800 to 2316 2.0 3.0 1.3 2.0
L -200 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
1000 to 2000 1.6 2.4 0.6 0.9
2000 to 2500 2.0 3.0 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6
Model NameDescriptionPrice List
Fluke 754Documenting Process Calibrator-HART including:
  • BC7240 battery charger
  • Li-on BP7240 battery pack
  • DPCTrack 2™ Sample Software
  • Instruction manual
  • NIST-traceable calibration report and data
  • Three sets of TP220 test probes with three sets of “extended tooth alligator clips
  • Two sets AC280 hook clips
  • C799 Soft Field Case
  • USB communication cable, Fluke 754HHC HART communications cable
On Request

ClipsPrice List
AC220AC220 SureGrip™ Alligator Clip SetOn Request
AC280AC280 SureGrip™ Hook ClipsOn Request
AC285AC285 SureGrip™ Alligator Clip SetOn Request

Test ProbesPrice List
TP74TP74 Lantern Tip Test ProbesOn Request
TP920TP920 Test Probe Adapter KitOn Request

Differential Pressure ModulesPrice List
700P00700P00 Differential Pressure Module (2.5 mbar)On Request
700P01700P01 Differential Pressure ModuleOn Request
700P02700P02 Differential Pressure Module (70 mbar)On Request
700P03700P03 Differential Pressure Module (340 mbar)On Request
700P04700P04 Differential Pressure Module (1000 mbar)On Request

Differential Pressure Module WetPrice List
700P22700P22 Differential Pressure Module Wet (70 mbar)On Request
700P23700P23 Differential Pressure Module Wet (340 mbar)On Request
700P24700P24 Differential Pressure Module WetOn Request

Absolute Pressure ModulesPrice List
700PA3700PA3 Absolute Pressure Module (340 mbar)On Request
700PA4700PA4 Absolute Pressure ModuleOn Request
700PA5700PA5 Absolute Pressure Module (2 bar)On Request
700PA6700PA6 Absolute Pressure Module (7 bar)On Request

Batteries, Chargers and AdaptersPrice List
BC7240BC7240 Battery charger/Eliminator for 75xOn Request
BP7240BP7240 Li-Ion battery pack for 75xOn Request

Cases and HolstersPrice List
C700C700 Hard Case (700 Series)On Request
C781C781 Meter CaseOn Request

Dual Pressure ModulesPrice List
700PD2700PD2 Dual Pressure Module (± 70 mbar)On Request
700PD3700PD3 Dual Pressure Module (± 340 mbar)On Request
700PD4700PD4 Dual Pressure Module (±1000 mbar)On Request
700PD5700PD5 Dual Pressure Module (-1000/+2000 mbar)On Request
700PD6700PD6 Dual Pressure Module (-1000 mbar/+6.9 bar)On Request
700PD7700PD7 Dual Pressure Module ( -1000 mbar/+13.8 bar) On Request

SoftwarePrice List
750SWFluke 750 SW DPC/TRACK2 SoftwareOn Request

Vacuum Pressure ModulesPrice List
700PV3700PV3 Vacuum Pressure Module (-340 mbar)On Request
700PV4700PV4 Vacuum Pressure Module( -1000 mbar)On Request

TemperaturePrice List
80CJ-M80CJ-M Male Mini Connectors (Type J)On Request
80CK-M80CK-M Male Mini Connectors (Type K)On Request
80PJ-180PJ-1 Bead ProbeOn Request
80PJ-EXT80PJ-EXT Extension Wire Kit (Type J)On Request
80PK-180PK-1 Bead Probe (Type K) On Request
80PK-2480PK-24 Air Probe On Request
80PK-2580PK-25 Piercing Probe On Request
80PK-2680PK-26 General Purpose ProbeOn Request
80PK-980PK-9 General Purpose Probe (Type K)On Request
80PK-EXT80PK-EXT Extension Wire Kit (Type K)On Request
80PT-2580PT-25 SureGrip Piercing Temperature ProbeOn Request
80PK-2280PK-22 SureGrip™ Immersion Temperature ProbeOn Request
80PK-2780PK-27 SureGrip™ Industrial Surface Temperature ProbeOn Request
80PK-880PK-8 Pipe Clamp Temperature ProbeOn Request

Test LeadsPrice List
TL940TL940 Mini-Hook Test Lead SetOn Request
TL950TL950 Mini-Pincer Test Lead SetOn Request

Pressure Calibrator Resources

Automated field calibration for the oil and gas industry (.pdf) »
Home   |   Site Map   |   Fluke Corporation   |   Safety Notices
Privacy Statement   |   Terms of Use  |   Terms and Conditions of Sale