Keeping your world up and running.
Home  |  Products  |  Process Calibration Tools  |  Pressure Calibrators  |  

Fluke 750P Series Pressure Modules

Fluke 750P Series Pressure Modules
Fluke 750P Series Pressure Modules
Fluke 750P Series Pressure Modules

A Complete Family of Pressure Modules

A family of 50 pressure modules covers pressure calibrations from 0 to 1 in H20 to 10,000 psi (2.5 mBar to 690 bar).

Gage pressure modules have one pressure fitting and measure the pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. Each module is clearly labeled for pressure range and media compatibility. All modules include NPT, metric (BSP) and M20 adapters. Compatible with all documenting, multifunction and pressure calibrators (including: Fluke 754, 752, 726, 725, 721, 719Pro, 719, 717 and legacy 741, 743, 744, 701 and 702 calibrators).

Quick and Easy Measurements

Fluke 750P Series Pressure Modules are easy to use. To measure pressure, connect the pressure module to a pressure source or hand pump and then connect the pressure module cable to the calibrator. Apply pressure from the pressure source and it is displayed digitally on the calibrator. At the touch of a button, the pressure may be displayed in up to 11 different engineering units. When used with the 750 Series Documenting Process Calibrators, pressure readings can be date/time stamped and stored electronically for later retrieval. This saves time, eliminates errors, and supports compliance with quality standards and regulations.

Choose from:

  • Differential modules
  • Gage modules
  • High pressure modules
  • Absolute modules
  • Vacuum modules
  • Dual range modules
  • Reference modules
  • Intrinsically safe modules

Precision Pressure Measurement for 75X and 720 Series Calibrators

The 750P Series Pressure Modules are the ideal pressure modules to enable gage, differential and absolute pressure measurement with Fluke 750 and 740 series DPCs and 725, 726 MPCs to measure pressure.

  • Up to 0.01 % reference uncertainty
  • 6-month and 1-year specifications
  • Temperature compensated 0 °C to 50 °C
  • Digital communication to calibrators, no analog losses or errors
  • Broad selection of ranges
  • Gage, differential, dual range, absolute and vacuum measurement models
  • Intrinsically safe models are Class I Div 1 Groups A-D T4, Ta = 0 °C to +50 °C, ATEX II 1G Ex ia IIC T4 compliant
  • Eight (8) Intrinsically Safe models that are certified to:
    • NEC-500: Class I Div 1, Groups A-D, Ga
    • ATEX: II 1 G Ex ia IIC T4 Ga
    • IECEx: Ex ia IIC T4 Ga *All certifications: Ta = -10 °C to +50 °C
Differential Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750P00 0 to 1 in H2O (0 to 2.5 mBar) 30X Dry Air Dry Air
750P3IN 0 to 3 in H2O (0 to 7.5 mBar) 10X Dry Air Dry Air
750P5IN 0 to 5 in H2O (0 to 12.5 mBar) 6X Dry Air Dry Air
750P017 0 to 10 in H2O (0 to 25 mBar) 3X Dry Air Dry Air
750P02 0 to 1 psi (0 to 70 mBar) 3X Dry Air Dry Air
750P22 0 to 1 psi (0 to 70 mBar) 3X 316 SS Dry Air
750P03 0 to 5 psi (0 to 350 mBar) 3X Dry Air Dry Air
750P23 0 to 5 psi (0 to 350 mBar) 4X 316 SS Dry Air
750P04 0 to 15 psi (0 to 1 bar) 3X Dry Air Dry Air
750P247 0 to 15 psi (0 to 1 bar) 4X 316 SS Dry Air
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750P00 ±0.15% ±0.3% ±0.35% ±0.25% ±0.30%
750P3IN ±0.15% ±0.3% ±0.35% ±0.25% ±0.35%
750P5IN ±0.15% ±0.3% ±0.35% ±0.25% ±0.35%
750P017 ±0.1% ±0.2% ±0.3% ±0.15% ±0.25%
750P02 ±0.050% ±0.1% ±0.15% ±0.075% ±0.125%
750P22 ±0.050% ±0.1% ±0.15% ±0.075% ±0.125%
750P03 ±0.02% ±0.04% ±0.05% ±0.035% ±0.04%
750P23 ±0.02% ±0.04% ±0.05% ±0.035% ±0.04%
750P04 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P247 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%

Gage Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750P057 0 to 30 psi (0 to 2 bar) 4X 316 SS N/A
750P067 0 to 100 psi (0 to 7 bar) 4X 316 SS N/A
750P277 0 to 300 psi (0 to 20 bar) 4X 316 SS N/A
750P07 0 to 500 psi (0 to 35 bar) 4X 316 SS N/A
750P08 0 to 1000 psi (0 to 70 bar) 3X 316 SS N/A
750P097 0 to 1500 psi (0 to 100 bar) 3X 316 SS N/A
750P2000 0 to 2000 psi (0 to 140 bar) 3X 316 SS N/A
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750P057 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P067 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P277 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P07 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P08 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P097 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P2000 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%

High Pressure Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750P29 0 to 3000 psi (0 to 200 bar) 3X 316 SS N/A
750P30 0 to 5000 psi (0 to 340 bar) 3X 316 SS N/A
750P31 0 to 10000 psi (0 to 700 bar) 2X 316 SS N/A
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750P29 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P30 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750P31 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%

Absolute Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750PA3 0 to 5 psi (0 to 350 mBar) 4X 316 SS N/A
750PA47 0 to 15 psi (0 to 1 bar) 4X 316 SS N/A
750PA5 0 to 30 psi (0 to 2 bar) 4X 316 SS N/A
750PA6 0 to 100 psi (0 to 7 bar) 4X 316 SS N/A
750PA27 0 to 300 psi (0 to 20 bar) 4X 316 SS N/A
750PA7 0 to 500 psi (0 to 35 bar) 4X 316 SS N/A
750PA8 0 to 1000 psi (0 to 70 bar) 3X 316 SS N/A
750PA9 0 to 1500 psi (0 to 100 bar) 3X 316 SS N/A
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750PA3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA47 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA5 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA6 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA27 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA7 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA8 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PA9 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%

Vacuum Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750PV3 -5 psi (-350 mBar) 4X 316 SS Dry Air
750PV4 -15 psi (-1 bar) 4X 316 SS Dry Air
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750PV3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PV4 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%

Dual Range Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750PD2 -1 to 1 psi (-70 to 70 mBar) 4X 316 SS Dry Air
750PD3 -5 to 5 psi (-350 to 350 mBar) 4X 316 SS Dry Air
750PD10 -10 to 10 psi (-700 to 700 mBar) 4X 316 SS Dry Air
750PD4 -15 to 15 psi (-1 to 1 bar) 4X 316 SS Dry Air
750PD5 -15 to 30 psi (-1 to 2 bar) 4X 316 SS N/A
750PD50 -15 to 50 psi (-1 to 3.5 bar) 4X 316 SS N/A
750PD6 -15 to 100 psi (-1 to 7 bar) 4X 316 SS N/A
750PD7 -15 to 200 psi (-1 to 14 bar) 4X 316 SS N/A
750PD27 -15 to 300 psi (-1 to 20 bar) 4X 316 SS N/A
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750PD2 ±0.05% ±0.1% ±0.15% ±0.075% ±0.125%
750PD3 ±0.03% ±0.06% ±0.07% ±0.05% ±0.06%
750PD10 ±0.025% ±0.05% ±0.07% ±0.04% ±0.06%
750PD4 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750PD5 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750PD50 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750PD6 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750PD7 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%
750PD27 ±0.0175% ±0.035% ±0.045% ±0.03% ±0.04%

Reference Module Specifications
Model Parameter / Range Burst Rating6 Hi Side Media2 Lo Side Media2
750R045 0 to 15 psi (0 to 1 bar) 3X Dry Air Dry Air
750R065 0 to 100 psi (0 to 7 bar) 4X 316 SS N/A
750R27 0 to 300 psi (0 to 20 bar) 4X 316 SS N/A
750R07 0 to 500 psi (0 to 35 bar) 4X 316 SS N/A
750R085 0 to 1000 psi (0 to 70 bar) 3X 316 SS N/A
750R29 0 to 3000 psi (0 to 200 bar) 3X 316 SS N/A
750R30 0 to 5000 psi (0 to 340 bar) 3X 316 SS N/A
750R315 0 to 10000 psi (0 to 700 bar) 2X 316 SS N/A
750RD5 -15 to 30 psi (-1 to 2 bar) 4X Dry Air N/A
750RD65 -12 to 100 psi (-0.8 to 7 bar) 4X 316 SS N/A
750RD27 -12 to 300 psi (-0.8 to 20 bar) 4X 316 SS N/A
Model Reference Uncertainty4 Total Uncertainty
1-year
(15-35°C)
Total Uncertainty
1-year1
Total Uncertainty
6-month
(15-35°C)
Total Uncertainty
6-month1
750R045 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R065 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R27 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R07 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R085 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R29 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R30 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750R315 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750RD5 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750RD65 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
750RD27 ±0.01% of FS ±0.02% of FS ±0.04% of FS ±0.015% of FS ±0.035% of FS
1. Total uncertainty, % of full span for temperature range 0°C to +50°C, one year interval. Total uncertainty, 1.0% of full span for temperature range -10°C to 0°C, one year interval. No 6 month specification available for range -10°C to 0°C.
2. "Non-corrosive gasses" indicates dry air or non-corrosive gas as compatible media. "Stainless Steel 316-SS" indicates media compatible with Type 316 Stainless Steel.
3. Specifications % of Full Span unless otherwise noted.
4. Reference Uncertainty is the specification for as left data for 24 hours.
5. When reference class modules are used with fixed resolution products (717, 718, 719 series, 725 and 726) calibrators add ±1 count to the overall accuracy specification.
6. Burst rating specification refers to the multiplier times full scale of the module for the rated burst pressure.
7. Intrinsically Safe pressure module available in this pressure range.
Model NameDescription
750P5INPressure Module, 0 to 5 in H2O (0 to 12.5 mBar), (0 to 1.25 kPa)
750P00Pressure Module, 0 to 1 in H2O (0 to 2.5 mBar), (0 to 0.25 kPa)
750P01Pressure Module, 0 to 10 in H2O (0 to 25 mBar), (0 to 2.5 kPa)
750P22Pressure Module, 0 psi to 1 psi (0 to 70 mBar), (0 to 7 kPa)
750P23Pressure Module, 0 psi to 5 psi (0 to 350 mBar), (0 to 35 kPa)
750P04Pressure Module, 0 psi to 15 psi (0 to 1 bar), (0 to 100 kPa)
750P24Pressure Module, 0 psi to 15 psi (0 to 1 bar), (0 to 100 kPa)
750P05Pressure Module, 0 psi to 30 psi (0 to 2 bar), (0 to 200 kPa)
750P06Pressure Module, 0 psi to 100 psi (0 to 7 bar), (0 to 700 kPa)
750P27Pressure Module, 0 psi to 300 psi (0 to 20 bar), (0 to 2000 kPa)
750P07Pressure Module, 0 psi to 500 psi (0 to 35 bar), (0 to 3500 kPa)
750P08Pressure Module, 0 psi to 1000 psi (0 to 70 bar), (0 to 7000 kPa)
750P09Pressure Module, 0 psi to 1500 psi (0 to 100 bar), (0 to 10 MPa)
750P2000Pressure Module, 0 psi to 2000 psi (0 to 140 bar), (0 to 14 MPa)
750P29Pressure Module, 0 psi to 3000 psi (0 to 200 bar), (0 to 20 MPa)
750PA27Pressure Module, 0 psi to 300 psi (0 to 20 bar), (0 to 2000 kPa)
750P30Pressure Module, 0 psi to 5000 psi (0 to 340 bar), (0 to 34 MPa)
750P31Pressure Module, 0 psi to 10000 psi (0 to 700 bar), (0 to 70 MPa)
750PA3Pressure Module, 0 psi to 5 psi (0 to 350 mBar), (0 to 35 kPa)
750PA4Pressure Module, 0 psi to 15 psi (0 to 1 bar), (0 to 100 kPa)
750PA5Pressure Module, 0 psi to 30 psi (0 to 2 bar), (0 to 200 kPa)
750PA6Pressure Module, 0 psi to 100 psi (0 to 7 bar), (0 to 700 kPa)
750PA7Pressure Module, 0 psi to 500 psi (0 to 35 bar), (0 to 3500 kPa)
750PA8Pressure Module, 0 psi to 1000 psi (0 to 70 bar), (0 to 7000 kPa)
750PA9Pressure Module, 0 psi to 1500 psi (0 to 100 bar), (0 to 10 MPa)
750PV3Pressure Module, -5 psi (-350 mBar), (-35 kPa)
750PV4Pressure Module, -15 psi (-1 bar), (-100 kPa)
750PD2Pressure Module, -1 psi to 1 psi (-70 to 70 mBar), (-7 to 7 kPa)
750PD3Pressure Module, -5 psi to 5 psi (-350 to 350 mBar), (-35 to 35 kPa)
750PD10Pressure Module, -10 psi to 10 psi (-0.7 to 0.7 bar), (-70 to 70 kPa)
750PD4Pressure Module, -15 psi to 15 psi (-1 to 1 bar), (-100 to 100 kPa)
750PD5Pressure Module, -15 psi to 30 psi (-1 to 2 bar), (-100 to 200 kPa)
750PD50Pressure Module, -15 psi to 50 psi (-1 to 3.5 bar), (-100 to 350 kPa)
750PD6Pressure Module, -15 psi to 100 psi (-1 to 7 bar), (-100 to 700 kPa)
750PD7Pressure Module, -15 psi to 200 psi (-1 to 14 bar), (-100 to 1400 kPa)
750PD27Pressure Module, -15 psi to 300 psi (-1 to 20 bar), (-100 to 2000 kPa)
750R04Pressure Module, 0 psi to 15 psi (0 to 1 bar),  (0 to 100 kPa)
750R06Pressure Module, 0 psi to 100 psi (0 to 7 bar), (0 to 700 kPa)
750R27Pressure Module, 0 psi to 300 psi (0 to 20 bar), (0 to 2000 kPa)
750R07Pressure Module, 0 psi to 500 psi (0 to 35 bar), (0 to 3500 kPa)
750R08Pressure Module, 0 psi to 1000 psi (0 to 70 bar), (0 to 7000 kPa)
750R29Pressure Module, 0 psi to 3000 psi (0 to 200 bar), (0 to 20 MPa)
750R30Pressure Module, 0 psi to 5000 psi (0 to 340 bar), (0 to 34 MPa)
750R31Pressure Module, 0 psi to 10000 psi (0 to 700 bar), (0 to 70 MPa)
750RD5Pressure Module, -15 psi to 30 psi (-1 to 2 bar), (-100 to 200 kPa)
750RD6Pressure Module, -12 psi to 100 psi (-0.8 to 7 bar), (-80 to 700 kPa)
750RD27Pressure Module, -12 psi to 300 psi (-0.8 to 20 bar), (-80 to 2000 kPa)
750P03Pressure Module, 0 psi to 5 psi (0 to 350 mBar), (0 to 35 kPa)
750P02Pressure Module, 0 psi to 1 psi (0 to 70 mBar), (0 to 7 kPa)

Pumps and Accessories
700PMP700 PMP Pressure Pump
700HTH-1Fluke 700HTH-1 Hydraulic Test Hose
700LTP-1Fluke 700LTP-1 Low-pressure Test Pump
700PRV-1Fluke 700PRV-1 Pressure Relief Valve Kit
700PRV-1Fluke 700PRV-1 Pressure Relief Valve Kit
700PTP-1Fluke 700PTP-1 Pneumatic Test Pump
71X Hose KitFluke-71X, Hose Kit Accessory
700HTP-2Fluke 700HTP-2 Hydraulic Test Pump
        
Home  |   Site Map  |   Fluke Corporation  |  Safety Notices
Privacy Statement  |  Terms of Use   |  Terms and Conditions of Sale  |  Terms and Conditions of Service
© 1995 - 2018 Fluke Corporation